In the lab we have a total of 16 wet benches. They are divided into the following 8 types:

General Information

  • Abandoned glassware will be collected and no returned
  • Stuff MUST be labeled
  • Another warning about safety gear

Solvent Cleaning Benches

Solvent Cleaning Benches
SolventBench.jpg
Tool Type Wet Processing
Location Bays 5, 6 & 7
Supervisor Aidan Hopkins
Supervisor Phone (805) 893-2343
Supervisor E-Mail hopkins@ece.ucsb.edu
Description Custom Solvent Wet Bench
Manufacturer Pure Aire Corporation


About

The facility contains 4 stainless steel solvent benches for general processing using organic solvents. All solvent waste (except for certain chemicals that are collected) are poured into cups and centrally collected. The benches consist of 2 embedded variable power Crest-ultrasonic units, one kept at 70°C for heating resist strippers and the other kept at room temperature for general solvent work. Laminar flow fume hoods with hepa filtration. Nitrogen guns are also on both sides of the bench for drying samples. Digital stirring hotplates are included for heating and stirring of resist stripers only (No solvent heating). POLOS spin dryers are included in the benches. All processes are done in user-supplied glassware. Solvents are stored in large steel flammable materials cabinets. Two people may use the bench at a time and protective gear such as proper gloves must be used. Solvents supplied by the laboratory are: Acetone, Methanol, Isopropanol, MIBK, MEK, Toluene, EBR100. PR strippers supplied by the lab are: 1165 (NMP), AZ300T (NMP + TMAH), PRX-127. Other chemicals should be OK'd by laboratory staff prior to use in the laboratory.

Detailed Specifications

  • 2 7" x 10" Crest Ultrasonic Tanks with heating / variable power
  • 2 Nitrogen guns for sample drying
  • Solvent based processing; NO SOLVENT HEATING
  • Organic material removal up to 80°C
  • Cyanide-based plating and etching (NO ACIDS at bench)
  • Digital hot plate stirrers
  • POLOS rinse dryers


Develop Benches

Develop Benches
DevelopBench.jpg
Tool Type Wet Processing
Location Bays 6 & 7
Supervisor Aidan Hopkins
Supervisor Phone (805) 893-2343
Supervisor E-Mail hopkins@ece.ucsb.edu
Description Custom Developer Bench
Manufacturer Pure Aire Corporation


About

The facility contains 3 polypropylene Developing Benches for photolithographic development using TMAH / TEAH / KOH based developers. All is actively neutralized. The benches consist of two sinks with DI water hook ups and industrial water plenum flushes. Laminar flow fume hoods with hepa filtration. Nitrogen guns are also on both sides of the bench for drying samples. Digital hotplates are included for resist post exposure bakes or post develop hard bakes. POLOS spray puddle develop etch systems are included in the benches. All processes are done in user-supplied glassware. Two people may use the bench at a time and protective gear such as proper gloves must be used. Developers supplied by the laboratory include: AZ300MIF (0.26N TMAH developer), MF701 (0.24N TMAH developer), AZ400K (KOH based developer), pre-diluted AZ400K:DI 1:4. Other developers should be OK'd by laboratory staff prior to use in the laboratory.

Detailed Specifications

  • 2 Sinks, DI water
  • 2 Nitrogen guns for sample drying
  • NO SOLVENTS or STRIPPERS in bench
  • Class 100
  • Hot plates for post bakes
  • POLOS Spray-Puddle-Develop-Rinse-Dry systems


Spin Coat Benches

Spin Coat Benches
SpinBench.jpg
Tool Type Wet Processing
Location Bays 6 & 7
Supervisor Aidan Hopkins
Supervisor Phone (805) 893-2343
Supervisor E-Mail hopkins@ece.ucsb.edu
Description Custom Photoresist Spin Coat Bench
Manufacturer Pure Aire Corporation


About

The facility contains 3 stainless steel solvent benches for photoresist spin coating. The benches consist each of 2 integrated Headway PWM32 series photoresist spinners. Overhead and foot controls are provided. 8 preprogrammed recipes and 2 user-programmable recipes are offered. Automatic wafer lifting and centering stations are offered for spinning large wafers up to 8” in diameter. 6” or 8” Cee ultra-flat hotplates with 0.1°C temperature stability are preset for standard resist bake temperatures (90, 95, 100, 105, 110, 115°C). Other user-changeable hot plates are also provided. Nitrogen guns are also on both sides of the bench. Large resist bottles are stored in a refrigerator amd small user bottles are stored by group in a ventilated steel cabinet. Waste pipets and resist soaked wipes are collected for disposal. A list of stocked resists appears on the lithography processing web page. Other chemicals should be OK'd by laboratory staff prior to use in the laboratory.

Detailed Specifications

  • 2 Headway PWM32 spinners per bench, multi-step programming
  • 8 preset spin programs, 2 user defineable
  • Variety of wafer chucks for 5 mm x 5 mm pieces to 6” wafers
  • Lifters for large wafer centering
  • Preset 0.1°C stable, ultra-flat hotplates (90, 95, 1100, 105, 110, 115°C)
  • NO WAX on hotplates
  • Other hotplates for user defined temperatures
  • Hepa filtered laminar flow for Class 100
  • Manual dispense of resist; particle filtering available


Toxic Corrosive Benches

Toxic Corrosive Bench
ToxicBench.jpg
Tool Type Wet Processing
Location Bay 5
Supervisor Aidan Hopkins
Supervisor Phone (805) 893-2343
Supervisor E-Mail hopkins@ece.ucsb.edu
Description Custom Acid Wet Bench
Manufacturer Pure Aire Corporation


About

The facility contains 2 polypropylene toxic corrosive benches for general wet processing using Acids and Bases. All waste (except for certain chemicals that are collected) are actively neutralized. The benches consist of two sinks with DI water hook ups and industrial water plenum flushes. Laminar flow fume hoods with hepa filtration. Nitrogen guns are also on both sides of the bench for drying samples. Digital stirring hotplates and/or small benchtop ultrasonic units are included for agitation and heating during the process on select benches. POLOS spin rinse dryers or spray puddle etch systems are included in the benches. Custom-built spray etchers (with bench supplied nitrogen) can also be used. Commonly used acids and bases such as H2O2, NH4OH, HCl, H2SO4, and H3PO4 can be dispensed into glassware using pipets connected to small bottles in the process wells at the back of the bench. All processes are done in user-supplied glassware. Acids and Bases are stored in separate cabinets and collection/disposal policies are posted. Two people may use the bench at a time and protective gear such as proper gloves must be used. Acids supplied by the laboratory are: H2SO4, HNO3, H3PO4, 1.0 M Citric Acid, Acetic Acid, HCl, Chrome mask etchant, Aluminum Etchants A and D, Gold Etchant (TFA), and Nickel Etchant (TFB). Bases supplied by the lab are: NH4OH, NaOH, KOH, and H2O2. Other chemicals should be OK'd by laboratory staff prior to use in the laboratory.

Detailed Specifications

  • 2 Sinks, DI water
  • 2 Nitrogen guns for sample drying
  • Digital hot-plate stirrers
  • Ultrasonic Baths
  • NO SOLVENTS
  • HF & TMAH NOT permitted at these benches
  • POLOS Spray-Spin-Rinse systems
  • Hepa Filtered Laminar Flow Hoods


HF/TMAH Processing Bench

HF/TMAH Processing Bench
HFBench.jpg
Tool Type Wet Processing
Location Bay 5
Supervisor Aidan Hopkins
Supervisor Phone (805) 893-2343
Supervisor E-Mail hopkins@ece.ucsb.edu
Description ?
Manufacturer Pure Aire Corporation


NOTE: Proper safety gear must be worn at all times when using the Acid Benches. For the HF benches this includes an apron and face shield. Failure to abide by this rule will result in immediate suspension from the lab. NO WARNINGS. PERIOD.

About

Detailed Specifications

  • 2 Sinks, DI water
  • 2 Nitrogen guns for sample drying
  • Digital hot-plate stirrers
  • Ultrasonic Baths
  • NO SOLVENTS
  • HF Processing limited to these two benches
  • POLOS Spray-Spin-Rinse systems
  • Hepa Filtered Laminar Flow Hoods


Plating Bench

Plating Bench
PlatingBench.jpg
Tool Type Wet Processing
Location Bay 5
Supervisor Aidan Hopkins
Supervisor Phone (805) 893-2343
Supervisor E-Mail hopkins@ece.ucsb.edu
Description ?
Manufacturer Pure Aire Corporation


About

Detailed Specifications