Line 22: Line 22:


*1000 W ICP source, 500 W RF Sample Bias Source in etching chamber
*1000 W ICP source, 500 W RF Sample Bias Source in etching chamber
*Room Temp. 80°C sample temperature for etching. Default 15°C Chuck temperature.
*Room Temp. to 80°C sample temperature for etching. Default 15°C Chuck temperature.
*Optimal Emission Monitoring
*Etch pressure from 0.1 Pa to 5 Pa (0.75 mT - 37.5 mT)
*Etch pressure from 0.1 Pa to 5 Pa (0.75 mT - 37.5 mT)
*Cl<sub>2</sub>, BCl<sub>3</sub>, (Ar or CHF<sub>3</sub>), (CF<sub>4</sub> or SF<sub>6</sub>), N<sub>2</sub>, and O<sub>2</sub> in etch chamber
*Cl<sub>2</sub>, BCl<sub>3</sub>, (Ar or CHF<sub>3</sub>), (CF<sub>4</sub> or SF<sub>6</sub>), N<sub>2</sub>, and O<sub>2</sub> in etch chamber
*O<sub>2</sub>, N<sub>2</sub>, CF<sub>4</sub>, H<sub>2</sub>O Vapor for ashing chamber
*Single 6” diameter wafer capable system
*Single 6” diameter wafer capable system
*Pieces possible by mounting to 6” wafer
*Pieces possible by mounting to 6” wafer
Line 33: Line 31:
==Documentation==
==Documentation==


*[https://wiki.nanofab.ucsb.edu/w/images/f/f3/ICP-_2_SOP_Rev_A.pdf ICP #2 Operating Instructions]
*{{file|ICP-Etch-2-Operating-Manual.pdf|Operating Instruction Manual}}
*{{file|Panasonic2.pdf|Training Notes}}
*{{file|Panasonic2.pdf|Training Notes}}
*[https://wiki.nanofab.ucsb.edu/w/images/2/2b/ICP_-2_Gas_Change_SOP_Rev_B.pdf ICP #2 Gas Change Procedure]
*{{file|Gas-Change.pdf|Gas Change Instructions}}
*{{file|manualwafertransfer.pdf|Manual Wafer Transfer Instructions}}
*[https://wiki.nanofab.ucsb.edu/w/images/2/2e/Manual_Wafer_Transfer.pdf ICP #2 Manual Wafer Transfer Instructions]
*[[Laser Etch Monitoring|Laser Etch Monitor procedures]]
*[[Laser Etch Monitoring|Laser Etch Monitor procedures]]


Line 48: Line 46:
*'''Recipes > Dry Etch >''' [https://wiki.nanotech.ucsb.edu/w/index.php?title=ICP_Etching_Recipes#ICP_Etch_2_.28Panasonic_E640.29 '''<u>ICP2 Etching Recipes</u>''']
*'''Recipes > Dry Etch >''' [https://wiki.nanotech.ucsb.edu/w/index.php?title=ICP_Etching_Recipes#ICP_Etch_2_.28Panasonic_E640.29 '''<u>ICP2 Etching Recipes</u>''']
**Starting point recipes for ICP2 specifically.
**Starting point recipes for ICP2 specifically.
*[[ICP Etching Recipes#Process Control Data .28Panasonic 2.29|Process Control Data]]
**Historical Data records "calibration" etches to test tool performance.


*'''Recipes > [https://wiki.nanotech.ucsb.edu/w/index.php?title=Dry_Etching_Recipes <u>Dry Etching Recipes</u>]'''
*'''Recipes > [https://wiki.nanotech.ucsb.edu/w/index.php?title=Dry_Etching_Recipes <u>Dry Etching Recipes</u>]'''
**Table of all dry etching recipes, showing '''etched materials vs. tool''' etc.
**Table of all dry etching recipes, showing '''etched materials vs. tool''' etc.
**

== [[Process Group - Process Control Data#Panasonic ICP#2 - Process Control|Process Control Data]] ==

* Data showing "calibration" etches over time to test tool performance.

:[[File:ICP2 Process Control Data Example.jpg|alt=example ICP2 process control chart|none|thumb|250x250px|[https://docs.google.com/spreadsheets/d/1m0l_UK2lDxlgww4f6nfXe4aQedNeDZsLs46jQ5wR4zw/edit#gid=1804752281 Click for Process Control Charts]|link=https://docs.google.com/spreadsheets/d/1m0l_UK2lDxlgww4f6nfXe4aQedNeDZsLs46jQ5wR4zw/edit#gid=1804752281]]

Latest revision as of 16:19, 31 March 2026

ICP Etch 2 (Panasonic E626I)
Location Bay 2
Tool Type Dry Etch
Manufacturer Panasonic Factory Solutions
Model E626I
Description ICP Etch

Primary Supervisor Lee Sawyer
(805) 893-2123
lee_sawyer@ucsb.edu

Secondary Supervisor

Tony Bosch


Recipes Dry Etch Recipes

SignupMonkey: Sign up for this tool


About

This is a single-chamber tool for etching of a variety of materials. The chamber is configured as an ICP etching tool with 1000 W ICP power, 500 W RF substrate power, and RT - 80°C operation with back-side He cooling and an electrostatic chuck to maintain controlled surface temperatures during etching. This chamber has Cl2, BCl3, CF4, CHF3, SF6, Ar, N2, and O2 for gas sources and can be used to etch a variety of materials from SiO2 to metals to compound semiconductors. The chamber is evacuated with a 2000 lpm Osaka Vacuum magnetically levitated turbo pump, allowing for fast pump down.

The system is also equipped with a red laser monitoring system from Intellemetrics for more precise etch stop control.

Detailed Specifications

  • 1000 W ICP source, 500 W RF Sample Bias Source in etching chamber
  • Room Temp. to 80°C sample temperature for etching. Default 15°C Chuck temperature.
  • Etch pressure from 0.1 Pa to 5 Pa (0.75 mT - 37.5 mT)
  • Cl2, BCl3, (Ar or CHF3), (CF4 or SF6), N2, and O2 in etch chamber
  • Single 6” diameter wafer capable system
  • Pieces possible by mounting to 6” wafer
  • 670nm laser endpoint detector with camera and simulation software: Intellemetrics LEP 500

Documentation

Online Training Video

  • Panasonic ICP#2 Training Video
  • Important: This video is for reference only, and does not give you authorization to use the tool. You must be officially authorized by the supervisor before using this machine.

Recipes

  • Recipes > Dry Etching Recipes
    • Table of all dry etching recipes, showing etched materials vs. tool etc.

Process Control Data

  • Data showing "calibration" etches over time to test tool performance.
example ICP2 process control chart
Click for Process Control Charts