DUV Flood Expose: Difference between revisions
Jump to navigation
Jump to search
Content deleted Content added
No edit summary |
No edit summary |
||
| Line 1: | Line 1: | ||
<nowiki><nowiki>Insert non-formatted text here</nowiki><nowiki><nowiki>Insert non-formatted text here</nowiki></nowiki></nowiki>{{tool|{{PAGENAME}} |
|||
{{tool|{{PAGENAME}} |
|||
|picture=DUV.jpg |
|picture=DUV.jpg |
||
|type = Lithography |
|type = Lithography |
||
Revision as of 16:22, 12 March 2014
<nowiki>Insert non-formatted text here<nowiki>Insert non-formatted text here</nowiki></nowiki>
|
About
This unit consists of a collimated deep ultraviolet (DUV) light source (mercury gas discharge short arc lamp) and power supply. The substrate is placed on a rotating chuck (not presently working) and is exposed by opening a timer-controlled shutter.
Materials that are exposed are primarily spun-on thin films such as PMMA, PMGI, etc.
Detailed Specifications
- DUV wavelengths are 200-260 nm; the lamp power is limited to either 1250 or 2100 watts and can operate in either constant intensity or constant power mode
- The full spectrum of Hg emission wavelengths from 200-450 nm is present on the sample
- Exposure can be performed on a 4" wafer
- Lamps are nominally rated for 400 hours
- A reset timer was added in Feb. 1995 to limit "on" time (ie. provide for auto shut-off) for bulb life conservation
Spectra Images
Aligner with No Filter

Aligner with Filter for i-line Exposure

DUV Exposure System Spectra