Difference between revisions of "DUV Flood Expose"
Jump to navigation
Jump to search
Line 2: | Line 2: | ||
|picture=DUV.jpg |
|picture=DUV.jpg |
||
|type = Lithography |
|type = Lithography |
||
− | |super= |
+ | |super= Lee Sawyer |
− | |phone=(805) |
+ | |phone=(805)893-2123 |
|location=Bay 6 |
|location=Bay 6 |
||
− | |email= |
+ | |email=lee_sawyer@ucsb.edu |
|description = Deep UV Flood Exposer / Light source and HA-1KC2 power supply |
|description = Deep UV Flood Exposer / Light source and HA-1KC2 power supply |
||
|manufacturer = AB Manufacturing |
|manufacturer = AB Manufacturing |
Revision as of 14:59, 1 April 2019
<nowiki>Insert non-formatted text here<nowiki>Insert non-formatted text here</nowiki></nowiki>
|
About
This unit consists of a collimated deep ultraviolet (DUV) light source (mercury gas discharge short arc lamp) and power supply. The substrate is placed on a rotating chuck (not presently working) and is exposed by opening a timer-controlled shutter.
Materials that are exposed are primarily spun-on thin films such as PMMA, PMGI, etc.
Detailed Specifications
- DUV wavelengths are 200-260 nm; the lamp power is limited to 1000 watts and can operate in either constant intensity or constant power mode
- The full spectrum of Hg emission wavelengths from 200-450 nm is present on the sample
- Exposure can be performed on a 4" wafer
- Lamps are nominally rated for 400 hours
- A reset timer was added in Feb. 1995 to limit "on" time (ie. provide for auto shut-off) for bulb life conservation