Automated Wafer Cleaver (Loomis LSD-155LT): Difference between revisions
Jump to navigation
Jump to search
Content deleted Content added
| Line 29: | Line 29: | ||
==Recipes== |
==Recipes== |
||
*Recipes > Packaging > '''Wafer Cleaver Recipes (LSD-155LT)''' |
*Recipes > Packaging > '''[[Wafer Cleaver Recipes (LSD-155LT)]]''' |
||
*Wafer thinning is critical, see the Wafer Lapping (Allied XYZ) tool as well. |
*Wafer thinning is critical, see the Wafer Lapping (Allied XYZ) tool as well. |
||
Revision as of 21:48, 15 August 2023
| ||||||||||||||||||||||||||
About
The Loomis LSD-155Lt is a production scribe and break system that can be used for processing large grids, arrays of laser bars, cleaving high quality mirror facets, and dicing wafers.
Detailed Specifications
- Maximum Wafer Size: 4"
- Parts mounted to low tack tape and used in conjunction with 6" plastic rings.
- Automated cut maps at multiple angles (0° and 90° typical)
- ~few micron alignment to on-wafer features.
- Multiple scribe and cleave options including 'scribe and break', 'notch and cleave', and 'peck and cleave'.
- Users should try and follow the 3 to 1 ratio (distance between cleaves should be 3X the thickness of the substrate) although some material like InP and GaAs can occasionally be cleaved closer to 2 to 1.
Operating Procedures
Recipes
- Recipes > Packaging > Wafer Cleaver Recipes (LSD-155LT)
- Wafer thinning is critical, see the Wafer Lapping (Allied XYZ) tool as well.