Data are collected by our [https://nanofab.ucsb.edu/workforce#internships Process Group Interns] and reviewed by [[Staff List#Process Group|Process Group Staff]]. |
Data are collected by our [https://nanofab.ucsb.edu/workforce#internships Process Group Interns] and reviewed by [[Staff List#Process Group|Process Group Staff]]. |
||
=Deposition, Dielectric (Process Control Data)= |
=Deposition, Dielectric (Process Control Data)= |
||
[[File:PECVD SPC Chart Example.png|alt=SPC chart example|thumb|228x228px|Example Process Control Charts (SPC) for thin-film DepCals.]] |
[[File:PECVD SPC Chart Example.png|alt=SPC chart example|thumb|228x228px|Example Process Control Charts (SPC) for thin-film DepCals.|link=]] |
||
[[File:Surfscan 230113A7G2 after low particles.jpg|alt=screenshot of surfscan particle count|thumb|205x205px|Example particle counts taken on each tool+film.]] |
[[File:Surfscan 230113A7G2 after low particles.jpg|alt=screenshot of surfscan particle count|thumb|205x205px|Example particle counts taken on each tool+film.|link=]] |
||
[[File:PECVD1 SiO2 F50 WaferMap example.jpg|alt=Screenshot of Filmetrics F50 wafermap of typical DepCals film|thumb|215x215px|Example of DepCals Thickness/Refractive Index uniformity measurement (4% shown here).]] |
[[File:PECVD1 SiO2 F50 WaferMap example.jpg|alt=Screenshot of Filmetrics F50 wafermap of typical DepCals film|thumb|215x215px|Example of DepCals Thickness/Refractive Index uniformity measurement (4% shown here).|link=]] |
||
''Process Control data for various deposition tools in the lab.'' |
''Process Control data for various deposition tools in the lab.'' |
||
| Line 58: | Line 58: | ||
[[File:Process Control - Sputter 5 Titanium Sheet Resistance plot.png|alt=Screenshot of Sputter 5 Titanium Sheet Resistance process control chart|thumb|Sputter 5 Titanium - Sheet Resistance - Process Control Chart - '''Example''' data only. |
[[File:Process Control - Sputter 5 Titanium Sheet Resistance plot.png|alt=Screenshot of Sputter 5 Titanium Sheet Resistance process control chart|thumb|Sputter 5 Titanium - Sheet Resistance - Process Control Chart - '''Example''' data only. |
||
← click a link for live data.]] |
← click a link for live data.|link=]] |
||
=== Sputter 5: Ti === |
=== Sputter 5: Ti === |
||
| Line 75: | Line 75: | ||
[[File:Process Control - E-Beam 4 Titanium Stress plot.png|alt=Screenshot of E-Beam 4 Titanium Stress - process control chart|thumb|E-Beam 4 Titanium Stress - Process Control Chart. '''Example''' data only - |
[[File:Process Control - E-Beam 4 Titanium Stress plot.png|alt=Screenshot of E-Beam 4 Titanium Stress - process control chart|thumb|E-Beam 4 Titanium Stress - Process Control Chart. '''Example''' data only - |
||
← click a link for live data.]] |
← click a link for live data.|link=]] |
||
=== E-Beam 4: Ti === |
=== E-Beam 4: Ti === |
||
| Line 215: | Line 215: | ||
|[[File:ASML CD Cals - Example Plot.jpg|alt=ASML CD Calibration Data - Screenshot of SPC Plot|none|thumb|''Example SPC Chart - Measured Critical Dimension "CD" versus Date. [https://docs.google.com/spreadsheets/d/1xW1TFH_QjPMWl9T1jiKzwmYe4B2wg7KY-nqOKUoXttI/edit#gid=1804752281 Click for charts.]''|link=https://docs.google.com/spreadsheets/d/1xW1TFH_QjPMWl9T1jiKzwmYe4B2wg7KY-nqOKUoXttI/edit#gid=1804752281]] |
|[[File:ASML CD Cals - Example Plot.jpg|alt=ASML CD Calibration Data - Screenshot of SPC Plot|none|thumb|''Example SPC Chart - Measured Critical Dimension "CD" versus Date. [https://docs.google.com/spreadsheets/d/1xW1TFH_QjPMWl9T1jiKzwmYe4B2wg7KY-nqOKUoXttI/edit#gid=1804752281 Click for charts.]''|link=https://docs.google.com/spreadsheets/d/1xW1TFH_QjPMWl9T1jiKzwmYe4B2wg7KY-nqOKUoXttI/edit#gid=1804752281]] |
||
|} |
|} |
||
== Maskless Aligner (Heidelberg MLA150) - Process Control == |
|||
* ''The Process Group measures the smallest "clean" line-space pitch (by SEM) at different locations on the wafer, and calls this the "best resolution" for that region of the wafer.'' ''Measured with 0.25µm steps.'' |
|||
* [https://docs.google.com/spreadsheets/d/1DbvFFNYeKeS1hG0Hnd-vCQs8CfqnMj2ABdjUv3lyssg/edit?usp=drive_link '''Data sheet'''], listing measured values, includes links to travelers & SEM's etc. |
|||
* [https://docs.google.com/spreadsheets/d/1DbvFFNYeKeS1hG0Hnd-vCQs8CfqnMj2ABdjUv3lyssg/edit?gid=167054213#gid=167054213 '''Plots''' of "best" resolution across wafer] - chart of resolution vs. time |
|||
[[File:MLA LithoCal Chart screenshot.png|alt=MLA LithoCal Chart screenshot|none|thumb|Example SPC Charts for MLA LithoCal "best" resolution. ]] |
|||
: |
|||
Latest revision as of 17:48, 11 August 2026
Process Control Data are standardized processes, run by the NanoFab, allowing for day-to-day or year-by-year comparisons of a tool's performance at the process level. This is similar Statistical Process Control (SPC).
These are the same links are found on individual tool pages, in the <<tool page>> > Process Control section.
Data are collected by our Process Group Interns and reviewed by Process Group Staff.
Deposition, Dielectric (Process Control Data)



Process Control data for various deposition tools in the lab.
PECVD #1 (PlasmaTherm 790) - Process Control
PECVD #2 (Advanced Vacuum) - Process Control
ICP-PECVD (Unaxis VLR Dep) - Process Control
- ICP-PECVD: Plots of SiO2 Films
- ICP-PECVD: Plots of Si3N4 Films
- ICP-PECVD: SiO2 Low-Dep Rate (LDR)
- ICP-PECVD: SiO2 High-Dep Rate (HDR)
- ICP-PECVD: Si3N4
- ICP-PECVD: Si3N4 Low-Stress
Ion Beam Sputter Deposition (Veeco Nexus) - Process Control
Old Data (Pre 2022)
Old data in a different format can be found below:
Deposition, Metal - Process Control Data
Process Control data for various deposition tools in the lab.
Sputter 5 (AJA ATC 2200-V) - Process Control

Sputter 5: Ti
Sputter 5: Al
Sputter 5: Cr
E-Beam 4 (CHA) - Process Control

E-Beam 4: Ti
E-Beam 4: Au
E-Beam 4: Cr
E-Beam 4: Ni
E-Beam 1 (Sharon) - Process Control
E-Beam 1: Ti
E-Beam 1: Au
E-Beam 1: Cr
E-Beam 1: Ni
Etching (Process Control Data)
Process Control data for various dry etching tools in the lab.
PlasmaTherm SLR Fluorine Etcher - Process Control
SiO2 Etching (FL-ICP Process Control)
We have found that SiO2 etching is fairly insensitive to chamber condition.
- SiO2 Etching with CHF3/CF4 - Etch Data
- SiO2 Etching with CHF3/CF4 - Plots

Click for Process Control Charts
Old SiO2 Process Control Data
- SiO2 Etching with CHF3/CF4 (FL-ICP) - No data prior to 2023-01-20
Si Etching (FL-ICP Process Control)
This Si etch is much more sensitive to chamber condition, allowing us to detect chamber contamination faster.
- Recipe: SiVertHFv2 - C4F8/SF6/CF4 etch of 100mm Silicon Wafer with ~50% open area and resist mask
- Si Etching with C4F8/SF6/CF4 - Etch Data
- Si Etching with C4F8/SF6/CF4 - Plots

Click for Process Control Charts
Panasonic ICP #1 - Process Control
Old Process Control Data
- SiO2 Etch with CHF3/CF4 (Panasonic 1) - No data prior to 2023-01-20
Panasonic ICP#2 - Process Control
GaAs Etch with N2/Cl2 - Process Control Data (Panasonic 2)
SiO2 Etch with CHF3/CF4 - Process Control Data (Panasonic 2)
Old Process Control Data
- SiO2 Etching with CHF3/CF4 - ICP2 - No data prior to 2023-01-20
Unaxis VLR Etch - Process Control
Oxford PlasmaPro Cobra Etcher - Process Control
Std InP Ridge Etch Cl2/H2/Ar/200°C
Calibration / Process testing data taken using the "InP Ridge Etch" process: Cl2/H2/Ar @ 200°C, 1cm piece with ~50% SiO2 hardmask.
- "Std InP Ridge Etch" Cl2/H2/Ar 200°C - Etch Data Tables
- "Std InP Ridge Etch" Cl2/H2/Ar 200°C - Plots

Std InP Ridge Etch: Cl2/CH4/H2/60°C
Calibration / Process testing data taken using the "InP Ridge Etch" process: Cl2/CH4/H2 @ 60°C, 1cm piece with ~50% SiO2 hardmask. No longer calibrating this recipe.
- "Std InP Ridge Etch" Cl2/CH4/H2/60°C - Etch Data Tables
- "Std InP Ridge Etch" Cl2/CH4/H2/60°C - Plots

Click for Process Control Charts
Old InP Ridge Etch Data
- InP Ridge Etch with Cl2/CH4/H2 @ 60°C - No data prior to 2023-01-20
GaN Etch (Cl2/BCl3/Ar/200°C)
Recipe: Std GaN Etch - BCl3/Cl2/Ar - 200C (Public), on 1cm x 1cm ~1.2µm deep GaN etch with Cl2/BCl3/Ar at 200°C. Sapphire substrate with SiO2 mask for GaN.
- GaN Etching with Cl2/BCl3/Ar at 200°C - Etch Data
- GaN Etching with Cl2/BCl3/Ar at 200°C - Plots

Click for Process Control Charts
PlasmaTherm DSEIII Deep Silicon Etcher - Process Control
Si Bosch Etching C4F8/SF6/Ar
- Recipe: STD_Bosch_Si (⭐️Production), on 100mm Si Wafer with ~50% open area, photoresist mask, ~40µm deep
- Si Etching with C4F8/SF6/Ar - Etch Data
- Si Etching with C4F8/SF6/Ar - Plots

Click for Process Control Charts
Lithography (Process Control Data)
Process Control Data for Nanofab Lithography/patterning tools.
Stepper #3 (ASML DUV) - Process Control
- The Process Group regularly measures data on lithography Critical Dimension ("CD") and Wafer-stage Particulate Contamination for this tool, using a sensitive lithography process that will reveal small changes in Dose repeatability and wafer flatness.
- Plots of CD Repeatability
- Data for CD Uniformity and Particulate Contamination

Example of Data Table with SEM's of 320nm features. Click for full data table. 
Example SPC Chart - Measured Critical Dimension "CD" versus Date. Click for charts.
Maskless Aligner (Heidelberg MLA150) - Process Control
- The Process Group measures the smallest "clean" line-space pitch (by SEM) at different locations on the wafer, and calls this the "best resolution" for that region of the wafer. Measured with 0.25µm steps.
- Data sheet, listing measured values, includes links to travelers & SEM's etc.
- Plots of "best" resolution across wafer - chart of resolution vs. time