*[[Chemical List|'''The Chemical Lists''']] show stocked chemicals, photolithography chemicals, and how to bring new chemicals.
*[[Chemical List|'''The Chemical Lists''']] show stocked chemicals, photolithography chemicals, and how to bring new chemicals.


==Table of Wet Etching Recipes==
=='''Table of Wet Etching Recipes'''==
''Use the ↑ ↓ Arrows in the header row to sort the entire table by material, selectivity, etchant etc.''
''Use the ↑ ↓ Arrows in the header row to sort the entire table by material, selectivity, etchant etc.''
{| class="wikitable sortable"
{| class="wikitable sortable"
Line 11: Line 11:
!Material!!Etchant!!Rate (nm/min)!!Anisotropy!!Selective to!!Selectivity!!Ref.!!Notes!!Confirmed By/Date
!Material!!Etchant!!Rate (nm/min)!!Anisotropy!!Selective to!!Selectivity!!Ref.!!Notes!!Confirmed By/Date
|-
|-
|Photoresist, polymers/organics
|InP||HCl:H3PO4 (1:3)||~1000||Highly||InGaAs||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]|| ||Jon Doe
|H2SO4:H2O2 = 3:1
[[Wet Etching Recipes#Organic%20removal|Piranha Solution]]
|typ. 5-10min etch for polymer residue
|
|Cr, W, Au, Pt, Si, SiO2, SiN
|
|
|'''''Dangerous''''' boiling hazard - see '''''[[Wet Etching Recipes#Piranha%20Solution|Piranha Solution]]'''''
section below. Etches Ti, Ni, Hf
|[[Demis D. John]], 2017
|-
|-
|InP
|InP
|HCl:H3PO4 (1:3)
|H3PO4:HCl = 3:1
|~1000
|~1000
|Highly
|Highly
Line 21: Line 31:
|[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]
|[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]
|
|
|
|Jon Doe
|-
|-
|InP||H3PO4:HCl (3:1)||~1000||Highly||InGaAs||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]|| ||Jon Doe
|InP||H3PO4:HCl = 3:1||~1000||Highly||InGaAs||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]||
|
|-
|-
|InP
|InP
|H3PO4:HCl (3:1)
|HCl:H2O = 3:1
|~1000
|~5000
|
|Highly
|InGaAs or InGaAsP
~200nm stop-etch
|High
|
|Bubbles while etching
|
|-
|InGaAs
|H2SO4:H2O2:H2O = 1:1:10
|~600
|
|InP
|High
|
|Exothermic, may reduce selectivity if hot
|
|-
|InGaAsP
|InGaAsP
|H2SO4:H2O2:H2O = 1:1:10
|~600
|
|InP
|High
|High
|[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]
|
|
|Exothermic, may reduce selectivity if hot
|Jon Doe
(same as InGaAs stop-etch removal)
|
|-
|GaAs
|NH4OH:H2O2 = 1:30
|
| -
|AlGaAs,
Al > 80%

~200nm stop-etch
|High
|
|
|Garrett Cole
|-
|GaSb
|[https://dtrinkle.matse.illinois.edu/MatSE584/kap_6/advanced/t6_1_2.html Sirtl etch]:
CrO3 powder mixed with HF
|TBD
|TBD
|InAs
|Infinite
|PhD. Dissertation of Shamsul Arafin, T.U. Munich.
|Do not store mixed - will corrode and leak out of the bottle! Mix fresh every time. Read MSDS's!
|
|-
|AlGaAs,
Al ≥80%
|HF:H2O = 1:20
|
| -
|GaAs
|High
|
|
|Garrett Cole
|-
|Oxide of InP
|NH4OH:H2O = 1:10
|1min to remove
|
|InP
|unknown
|
|
|[[Ning Cao]]
|-
|Oxide of GaAs
|HCl:H2O = 1:10
|1min to remove
|
|GaAs
|unknown
|
|
|[[Demis D. John]]
|-
|-
|Al2O3 ''(ALD Plasma 300C)''||Developer: 300MIF||~1.6||None
|Al2O3 ''(ALD Plasma 300C)''||Developer: 300MIF||~1.6||None
Line 84: Line 172:
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiO2_deposition_.28PECVD_.231.29 SiO2 ''(PECVD #1)'']
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiO2_deposition_.28PECVD_.231.29 SiO2 ''(PECVD #1)'']
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|~500
|~550
|None
|None
|Photoresist
|Photoresist
Line 91: Line 179:
|May need to increase adhesion with 100°C baked HMDS.
|May need to increase adhesion with 100°C baked HMDS.
|Biljana Stamenic
|Biljana Stamenic
2023
2017
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiO2_deposition_.28PECVD_.232.29 SiO2 ''(PECVD #2)'']
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiO2_deposition_.28PECVD_.232.29 SiO2 ''(PECVD #2)'']
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|~500
|~680
|None
|None
|Photoresist
|Photoresist
Line 102: Line 190:
|May need to increase adhesion with 100°C baked HMDS.
|May need to increase adhesion with 100°C baked HMDS.
|Biljana Stamenic
|Biljana Stamenic
2023
2017
|-
|SiO2 ''(ALD -BDEAS 300C)''
|HF ("Buffered")
Diluted with DI
BHF:H2O = 1:100
|~7.46
|
|
|
|Measured in-house
|
|Biljana Stamenic
2024
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#SiO2_deposition_.28IBD.29 SiO2 (IBD)]
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#SiO2_deposition_.28IBD.29 SiO2 (IBD)]
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|~350
|~260
|None
|Photoresist
|High
|Measured in-house
|
|Biljana Stamenic
2023
|-
|[https://wiki.nanofab.ucsb.edu/wiki/PECVD_Recipes#ICP-PECVD_.28Unaxis_VLR.29 SiO2 LDR ''(Unaxis VLR)'']
|HF ("Buffered HF Improved", Transene)
|~170
|None
|Photoresist
|High
|Measured in-house
|
|Biljana Stamenic
2023
|-
|[https://wiki.nanofab.ucsb.edu/wiki/PECVD_Recipes#ICP-PECVD_.28Unaxis_VLR.29 SiO2 HDR ''(Unaxis VLR)'']
|HF ("Buffered HF Improved", Transene)
|~230
|None
|None
|Photoresist
|Photoresist
Line 113: Line 236:
|
|
|Biljana Stamenic
|Biljana Stamenic
2023
2016
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiN_deposition_.28PECVD_.231.29 Si3N4 (PECVD#1)]
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiN_deposition_.28PECVD_.231.29 Si3N4 (PECVD#1)]
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|~120
|85
|None
|None
|Photoresist
|Photoresist
Line 124: Line 247:
|
|
|Biljana Stamenic
|Biljana Stamenic
2023
2018-04
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiN_deposition_.28PECVD_.232.29 Si3N4 (PECVD#2)]
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiN_deposition_.28PECVD_.232.29 Si3N4 (PECVD#2)]
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|~35
|35–45
|None
|None
|Photoresist
|Photoresist
Line 135: Line 258:
|
|
|Biljana Stamenic
|Biljana Stamenic
2023
2018-05
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#Low-Stress_SiN_deposition_.28PECVD_.232.29 Si3N4 Low-Stress (PECVD#2)]
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#Low-Stress_SiN_deposition_.28PECVD_.232.29 Si3N4 Low-Stress (PECVD#2)]
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|~30
|35–50
|None
|None
|Photoresist
|Photoresist
Line 146: Line 269:
|
|
|Biljana Stamenic
|Biljana Stamenic
2023
2018-05
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Si3N4_deposition_.28IBD.29 Si3N4 (IBD)]
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Si3N4_deposition_.28IBD.29 Si3N4 (IBD)]
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|~5
|5–15
|None
|None
|Photoresist
|Photoresist
Line 157: Line 280:
|
|
|Biljana Stamenic
|Biljana Stamenic
2023
2014
|-
|[https://wiki.nanofab.ucsb.edu/wiki/PECVD_Recipes#ICP-PECVD_.28Unaxis_VLR.29 SiN ''(Unaxis VLR)'']
|HF ("Buffered HF Improved", Transene)
|~10
|None
|Photoresist
|High
|Measured in-house
|
|Biljana Stamenic
2023
|-
|[https://wiki.nanofab.ucsb.edu/wiki/PECVD_Recipes#ICP-PECVD_.28Unaxis_VLR.29 SiN Low Stress ''(Unaxis VLR)'']
|HF ("Buffered HF Improved", Transene)
|~135
|None
|Photoresist
|High
|Measured in-house
|
|Biljana Stamenic
2023
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Ta2O5_deposition_.28IBD.29 Ta2O5 (IBD)]
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Ta2O5_deposition_.28IBD.29 Ta2O5 (IBD)]
|HF ("Buffered HF Improved", Transene)
|HF ("Buffered HF Improved", Transene)
|0.4
|0.07
|None
|None
|Photoresist
|Photoresist
Line 168: Line 313:
|
|
|Biljana Stamenic
|Biljana Stamenic
2023
2016-12
|-
|-
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#TiO2_deposition_.28IBD.29 TiO2 (IBD)]
|[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#TiO2_deposition_.28IBD.29 TiO2 (IBD)]
Line 196: Line 341:
|-
|-
!Material!!Etchant!!Rate (nm/min)!!Anisotropy!!Selective to!!Selectivity!!Ref.!!Notes!!Confirmed By/Date
!Material!!Etchant!!Rate (nm/min)!!Anisotropy!!Selective to!!Selectivity!!Ref.!!Notes!!Confirmed By/Date
|}

===Adding a new entry to the Table===
When entering a new etch into the table:

*Make a row for every etchant used in the solution such that the information can be sorted by etchant. For example, the InP etch HCl:H3PO4(1:3) and H3PO4:HCl(3:1).
*Likewise, if etch is known to be selective to multiple materials the etch should have a row for each material. For example HCl:H3PO4(1:3) is selective to both InGaAs and InGaAsP.

This multiple entry method may seem laborious for the person entering a new etch, however the power of sorting by selective materials and chemicals in a table with all materials is great.

{| class="wikitable sortable"
|-
!Material!!Etchant!!Rate (nm/min)!!Anisotropy!!Selective to!!Selectivity!!Ref.!!Notes!!Confirmed by!!Extra column
|-
|InP||HCl:H3PO4(1:3)||~1000||Highly||InGaAs||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]||Example||Jon Doe||Example
|-
|InP||HCl:H3PO4(1:3)||~1000||Highly||InGaAsP||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]||Example||Jon Doe||Example
|-
|InP||H3PO4:HCl(3:1)||~1000||Highly||InGaAs||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]||Example||Jon Doe||Example
|-
|InP||H3PO4:HCl(3:1)||~1000||Highly||InGaAsP||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]||Example||Jon Doe||Example
|}
|}


Line 228: Line 352:


===Compound Semiconductor Etching===
===Compound Semiconductor Etching===
[http://www.sciencedirect.com/science/article/pii/S0927796X00000279 Guide to references on III±V semiconductor chemical etching]
[http://www.sciencedirect.com/science/article/pii/S0927796X00000279 Guide to references on III±V semiconductor chemical etching (A.R. Clawson, 2001)]


* Impressively vast list of various III-V wet etches, organized by various applications (eg. "selective GaAs against AlGaAs" or "non-selective InP/InGaAsP" etc.)
Please add any confirmed etches from this reference to the {{HLink|Wet Etching Recipes|The Master Table of Wet Etching (Include All Materials)}}.
* Please add any confirmed etches from this reference to the {{HLink|Wet Etching Recipes|The Master Table of Wet Etching (Include All Materials)}}.


===Metal Etching===
===Metal Etching===
Line 246: Line 371:
==Organic removal==
==Organic removal==


===Piranha Solution===
*Piranha Solution. '''Careful!''' Read about how to prepare and handle this safely:

*'''Careful!''' Read about how to prepare and handle this safely:
**[http://web.mit.edu/cortiz/www/PiranhaSafety.doc MIT's Piranha Solution safety document]
**[http://web.mit.edu/cortiz/www/PiranhaSafety.doc MIT's Piranha Solution safety document]
**[https://www.ehs.harvard.edu/sites/default/files/lab_safety_guideline_piranha_etch.pdf Harvard EHS's Handling Document]
**[https://www.ehs.harvard.edu/sites/default/files/lab_safety_guideline_piranha_etch.pdf Harvard EHS's Handling Document]
**Can be used for etching away Photoresist residues after dry etching
**Used for etching away Photoresist residues after dry etching.

*PureStrip @ 70°C (Transene)
===PureStrip (Transene)===

*Heat to @ 70°C
**[[Wet Benches#Wafer Toxic Corrosive Benches|Vertical heated bath on Wafer Toxic-Corrosive bench in Bay 4]]
**[[Wet Benches#Wafer Toxic Corrosive Benches|Vertical heated bath on Wafer Toxic-Corrosive bench in Bay 4]]
**After a few days heated, it loses potency - must drain + replenish with fresh solution.
**After a few days heated, it loses potency - must drain + replenish with fresh solution.
**~30-90min will remove stubborn, microscopic PR residues from dry etching.
**~30-90min will remove stubborn, microscopic PR residues from dry etching.


==Gold Plating==
==[[Gold Plating Bench|Gold Plating Bench (Technic SEMCON 1000)]]==
Standard plating recipes are described and taught during the equipment training for this tool.
''To Be Added''

Electroplating first requires a Gold seed layer to be present on all surfaces to be plated. Common ways to produce this are to:

*[[Tool List#Sputter Deposition|Sputter-coat]] a thin (~100nm) Au "seed layer" on all surfaces of the wafer, being careful to consider shadowing effects during the dep (eg. in high-aspect ratio trenches).
*Perform photolithography to protect (and prevent plating) in desired areas - areas with no photoresist will be plated.
**Leave ''Open'' the 4 contact points on the edge of the wafer, or some other contact point, for an electrode to make contact to the Seed layer.
**Optionally use EBR100 to swab away photoresist from the contact points.
*Perform the electroplating on the [[Gold Plating Bench|Technic SemCon]], contacting the seed layer with electrodes and executing the program for the desired current/time to achieve the plating thickness (typically microns).
*Strip the photoresist using standard solvents.
*Use the [[CAIBE (Oxford Ion Mill)|Oxford Ion Mill]] to blanket etch the seed layer all over the wafer (the 100nm removed from the plated regions will be negligible compared to the ~1µm plated).


==[[Chemical-Mechanical Polisher (Logitech)|Chemi-Mechanical Polishing (CMP)]]==
==[[Chemical-Mechanical Polisher (Logitech)|Chemi-Mechanical Polishing (CMP)]]==

Latest revision as of 03:25, 5 September 2026

Chemicals Available

  • The Chemical Lists show stocked chemicals, photolithography chemicals, and how to bring new chemicals.

Table of Wet Etching Recipes

Use the ↑ ↓ Arrows in the header row to sort the entire table by material, selectivity, etchant etc.

Material Etchant Rate (nm/min) Anisotropy Selective to Selectivity Ref. Notes Confirmed By/Date
Photoresist, polymers/organics H2SO4:H2O2 = 3:1

Piranha Solution

typ. 5-10min etch for polymer residue Cr, W, Au, Pt, Si, SiO2, SiN Dangerous boiling hazard - see Piranha Solution

section below. Etches Ti, Ni, Hf

Demis D. John, 2017
InP H3PO4:HCl = 3:1 ~1000 Highly InGaAsP High Lamponi (p.102)
InP H3PO4:HCl = 3:1 ~1000 Highly InGaAs High Lamponi (p.102)
InP HCl:H2O = 3:1 ~5000 InGaAs or InGaAsP

~200nm stop-etch

High Bubbles while etching
InGaAs H2SO4:H2O2:H2O = 1:1:10 ~600 InP High Exothermic, may reduce selectivity if hot
InGaAsP H2SO4:H2O2:H2O = 1:1:10 ~600 InP High Exothermic, may reduce selectivity if hot

(same as InGaAs stop-etch removal)

GaAs NH4OH:H2O2 = 1:30 - AlGaAs,

Al > 80%

~200nm stop-etch

High Garrett Cole
GaSb Sirtl etch:

CrO3 powder mixed with HF

TBD TBD InAs Infinite PhD. Dissertation of Shamsul Arafin, T.U. Munich. Do not store mixed - will corrode and leak out of the bottle! Mix fresh every time. Read MSDS's!
AlGaAs,

Al ≥80%

HF:H2O = 1:20 - GaAs High Garrett Cole
Oxide of InP NH4OH:H2O = 1:10 1min to remove InP unknown Ning Cao
Oxide of GaAs HCl:H2O = 1:10 1min to remove GaAs unknown Demis D. John
Al2O3 (ALD Plasma 300C) Developer: 300MIF ~1.6 None Most non-Al Materials. High Measured in-house Rate slows with time. JTB
Al2O3 (ALD Plasma 300C) Developer: 400K ~2.2 None Most non-Al Materials. High Measured in-house Rate slows with time. JTB
Al2O3 (ALD Plasma 300C) Developer: 400K (1:4) ~1.6 None Most non-Al Materials. High Measured in-house Rate slows with time. JTB
Al2O3 (ALD Plasma 300C) NH4OH:H2O2:H2O (1:2:50) ~<0.5 Measured in-house Rate slows with time JTB
Al2O3 (IBD) HF ("Buffered HF Improved", Transene) ~170 None Photoresist High Measured in-house May need to increase adhesion with thin SiO2 layer, and 100°C baked HMDS. Biljana Stamenic,

2017-12

Al2O3 (IBD) Developer: 726 MiF 3.5 None Most non-Al Materials. High Measured in-house Demis D. John,

2017-11

Al2O3 (AJA#4) Developer: 300 MiF 4.30 None Most non-Al Materials. High Measured in-house Demis D. John

2018-02

SiO2 (PECVD #1) HF ("Buffered HF Improved", Transene) ~550 None Photoresist High Measured in-house May need to increase adhesion with 100°C baked HMDS. Biljana Stamenic

2023

SiO2 (PECVD #2) HF ("Buffered HF Improved", Transene) ~680 None Photoresist High Measured in-house May need to increase adhesion with 100°C baked HMDS. Biljana Stamenic

2023

SiO2 (ALD -BDEAS 300C) HF ("Buffered")

Diluted with DI BHF:H2O = 1:100

~7.46 Measured in-house Biljana Stamenic

2024

SiO2 (IBD) HF ("Buffered HF Improved", Transene) ~260 None Photoresist High Measured in-house Biljana Stamenic

2023

SiO2 LDR (Unaxis VLR) HF ("Buffered HF Improved", Transene) ~170 None Photoresist High Measured in-house Biljana Stamenic

2023

SiO2 HDR (Unaxis VLR) HF ("Buffered HF Improved", Transene) ~230 None Photoresist High Measured in-house Biljana Stamenic

2023

Si3N4 (PECVD#1) HF ("Buffered HF Improved", Transene) ~120 None Photoresist High Measured in-house Biljana Stamenic

2023

Si3N4 (PECVD#2) HF ("Buffered HF Improved", Transene) ~35 None Photoresist High Measured in-house Biljana Stamenic

2023

Si3N4 Low-Stress (PECVD#2) HF ("Buffered HF Improved", Transene) ~30 None Photoresist High Measured in-house Biljana Stamenic

2023

Si3N4 (IBD) HF ("Buffered HF Improved", Transene) ~5 None Photoresist High Measured in-house Biljana Stamenic

2023

SiN (Unaxis VLR) HF ("Buffered HF Improved", Transene) ~10 None Photoresist High Measured in-house Biljana Stamenic

2023

SiN Low Stress (Unaxis VLR) HF ("Buffered HF Improved", Transene) ~135 None Photoresist High Measured in-house Biljana Stamenic

2023

Ta2O5 (IBD) HF ("Buffered HF Improved", Transene) 0.07 None Photoresist High Measured in-house Biljana Stamenic

2023

TiO2 (IBD) HF ("Buffered HF Improved", Transene) 1.0–2.0 None Photoresist High Measured in-house Biljana Stamenic

2014-12

Si (<100> crystalline) KOH (45%) @ 87°C ~730 High, Crystallographic, ~55° Low-Stress Si3N4 - either PECVD #2 or Commercial LPCVD Si3N4

Other Si3N4 also OK.

LS-SiN: High

PR etches quickly, SiO2 etches slowly.

Measured In-House

- Search online.

Use the Covered, Heated vertical bath (Dedi cated bath in Bay 4). Slight Bubbler. Brian Thibeault

2017

Material Etchant Rate (nm/min) Anisotropy Selective to Selectivity Ref. Notes Confirmed By/Date

Wet Etching References

  1. Etch rates for Micromachining Processing (IEEE Jnl. MEMS, 1996) - includes tables of etch rates of numerous metals vs. various wet and dry etchants.
  2. Etch rates for micromachining-Part II (IEEE Jnl. MEMS, 2003) - expanded tables containing resists, dielectrics, metals and semiconductors vs. many wet etch chemicals.
  3. Guide to references on III±V semiconductor chemical etching - exhaustive list of wet etchants for etching various semiconductors, including selective etches.
  4. Transene's Chemical Compatibility Chart provides a useful quick-reference for which Transene etchants attack which materials.
    1. As a side-note, Transene provides many pre-mixed solutions that you can order, saving you the time and uncertainty of measuring/mixing such chemicals yourself. Make sure you check with us before ordering so we know how to handle the chemical before it arrives.

Compound Semiconductor Etching

Guide to references on III±V semiconductor chemical etching (A.R. Clawson, 2001)

  • Impressively vast list of various III-V wet etches, organized by various applications (eg. "selective GaAs against AlGaAs" or "non-selective InP/InGaAsP" etc.)
  • Please add any confirmed etches from this reference to the The Master Table of Wet Etching (Include All Materials).

Metal Etching

Silicon etching

Etch rates for micromachining processing

Etch rates for micromachining processing-part II

Please add any confirmed etches from this reference to the The Master Table of Wet Etching (Include All Materials).

Organic removal

Piranha Solution

PureStrip (Transene)

Gold Plating Bench (Technic SEMCON 1000)

Standard plating recipes are described and taught during the equipment training for this tool.

Electroplating first requires a Gold seed layer to be present on all surfaces to be plated. Common ways to produce this are to:

  • Sputter-coat a thin (~100nm) Au "seed layer" on all surfaces of the wafer, being careful to consider shadowing effects during the dep (eg. in high-aspect ratio trenches).
  • Perform photolithography to protect (and prevent plating) in desired areas - areas with no photoresist will be plated.
    • Leave Open the 4 contact points on the edge of the wafer, or some other contact point, for an electrode to make contact to the Seed layer.
    • Optionally use EBR100 to swab away photoresist from the contact points.
  • Perform the electroplating on the Technic SemCon, contacting the seed layer with electrodes and executing the program for the desired current/time to achieve the plating thickness (typically microns).
  • Strip the photoresist using standard solvents.
  • Use the Oxford Ion Mill to blanket etch the seed layer all over the wafer (the 100nm removed from the plated regions will be negligible compared to the ~1µm plated).

Chemi-Mechanical Polishing (CMP)

To Be Added

Mechanical Polishing (Allied)

To Be Added