Photolithography - Improving Adhesion Photoresist Adhesion: Difference between revisions
Jump to navigation
Jump to search
Content deleted Content added
m mentioned lift-off delam |
Updated underlayer process and tips |
||
| (One intermediate revision by the same user not shown) | |||
| Line 1: | Line 1: | ||
| ⚫ | |||
==Underlayers for improving PR adhesion== |
==Underlayers for improving PR adhesion== |
||
High-temperature (>150°C) underlayers, such as LOL1000/2000, BARC layers tend to improve adhesion of the imaging resist. |
High-temperature (>150°C) underlayers, such as LOL1000/2000, BARC layers tend to improve adhesion of the imaging resist. |
||
| Line 7: | Line 5: | ||
*Apply high-temp underlayer, eg. LOL1000/2000, PMGI, DSK101 (DUV), DSK42P (DUV) |
*Apply high-temp underlayer, eg. LOL1000/2000, PMGI, DSK101 (DUV), DSK42P (DUV) |
||
**Beware of underlayers that have a develop-rate - undercutting the imaging PR fully can also cause delamination) |
**Beware of underlayers that have a develop-rate - undercutting the imaging PR fully can also cause delamination) |
||
*Bake at recommended temperature for the underlayer, typically >> 150°C. |
|||
*Apply your Imaging photoresist as normal |
*Apply & bake your Imaging photoresist as normal for that resist, |
||
**Bake temp should be lower than the underlayer bake temp. |
|||
*If your underlayer dissolves during the same develop step as the imaging resist, you will need to increase your develop time to go through both layers. |
|||
| ⚫ | |||
==HMDS Process for PR Improving Adhesion== |
==HMDS Process for PR Improving Adhesion== |
||
This process tends to improve adhesion between the substrate and subsequent photoresist layers. Use this procedure if you are finding poor adhesion (resist delaminating during develop etc.), or for chemicals (like BHF) that attack the PR adhesion interface strongly. |
This process tends to improve adhesion between the substrate and subsequent photoresist layers. Use this procedure if you are finding poor adhesion (resist delaminating during develop etc.), or for chemicals (like BHF) that attack the PR adhesion interface strongly. |
||
*150°C bake ~2min (Dehydration step) |
|||
*O2 plasma ([[Ashers (Technics PEII)|PEii Technics]]), 10sec |
|||
**Optionally could instead do Dehydration of: O2 plasma ([[Ashers (Technics PEII)|PEii Technics]]), 10sec; however this could create more surface SiO2 on silicon. |
|||
**This takes the place of the dehydration bake. |
|||
*''No bake'' |
|||
*HMDS application & soak 40s |
*HMDS application & soak 40s |
||
**Do this with sample loaded on the PR spinner. |
**Do this with sample loaded on the PR spinner. |
||
Latest revision as of 19:33, 10 September 2026
Underlayers for improving PR adhesion
High-temperature (>150°C) underlayers, such as LOL1000/2000, BARC layers tend to improve adhesion of the imaging resist.
- Substrate prep: O2 plasma (PEii Technics), 10sec
- Apply high-temp underlayer, eg. LOL1000/2000, PMGI, DSK101 (DUV), DSK42P (DUV)
- Beware of underlayers that have a develop-rate - undercutting the imaging PR fully can also cause delamination)
- Bake at recommended temperature for the underlayer, typically >> 150°C.
- Apply & bake your Imaging photoresist as normal for that resist,
- Bake temp should be lower than the underlayer bake temp.
- If your underlayer dissolves during the same develop step as the imaging resist, you will need to increase your develop time to go through both layers.
- Don't forget that, for lift-off processes, the Imaging (top) Photoresist will lift-off as well if you undercut it too much during develop!
HMDS Process for PR Improving Adhesion
This process tends to improve adhesion between the substrate and subsequent photoresist layers. Use this procedure if you are finding poor adhesion (resist delaminating during develop etc.), or for chemicals (like BHF) that attack the PR adhesion interface strongly.
- 150°C bake ~2min (Dehydration step)
- Optionally could instead do Dehydration of: O2 plasma (PEii Technics), 10sec; however this could create more surface SiO2 on silicon.
- HMDS application & soak 40s
- Do this with sample loaded on the PR spinner.
- Spin-dry - any spin speed, ~15-30sec
- HMDS bake at 100-110°C for 90s.
- Then your normal PR process on top.