ICP-PECVD (Unaxis VLR): Difference between revisions
Jump to navigation
Jump to search
(→About) |
No edit summary |
||
Line 21: | Line 21: | ||
*Multiple 4” diameter wafer capable system |
*Multiple 4” diameter wafer capable system |
||
*Pieces possible by mounting or placing on 4 ” wafer |
*Pieces possible by mounting or placing on 4 ” wafer |
||
=Documentation= |
|||
*[[media:PM1_Wiki_Operational_Procedure_3-14-14.pdf|Operating Instructions]] |
Revision as of 20:19, 18 March 2014
|
About
This system is configured as an ICP PECVD deposition tool with 1000 W ICP power, 600 W RF substrate power, and RT-350°C operation. This chamber has 100% SiH4, N2, O2, and Ar for gas sources. The high density PECVD produces a more dense, higher quality SiO2 and Si3N4, as compared with conventional PECVD. With the high density plasma, deposition of high quality films can be deposited as low as 50°C for processes requiring lower temperatures. Stress compensation for silicon nitride is characterized.
Detailed Specifications
- 1000W ICP source, 600W RF Sample Bias Power Supply
- 50 - 350°C sample temperature
- 100% SiH4, Ar, N2, O2
- Multiple 4” diameter wafer capable system
- Pieces possible by mounting or placing on 4 ” wafer