Atomic Force Microscope (Bruker ICON): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
Zwarburg (talk | contribs)
No edit summary
Zwarburg (talk | contribs)
No edit summary
Line 1: Line 1:
{{tool|{{PAGENAME}}
{{tool|{{PAGENAME}}
|picture=AFMNanoman.jpg
|picture=AFMNanoman.jpg
|type = Inspection, Test, and Characterization
|type = Inspection, Test and Characterization
|super= Bill Mitchell
|super= Bill Mitchell
|phone= (805)893-4974
|phone= (805)893-4974

Revision as of 16:40, 10 July 2012

Atomic Force Microscope (Bruker ICON)
File:AFMNanoman.jpg
Tool Type Inspection, Test and Characterization
Location Bay 7
Supervisor Bill Mitchell
Supervisor Phone (805) 893-4974
Supervisor E-Mail mitchell@ece.ucsb.edu
Description Dimension 3100 Nanoman
Manufacturer Veeco Metrology, LLC


About

The Dimension 3100 Nanoman AFM from Veeco provides a variety of high resolution surface imaging techniques and the ability to manipulate or create nanoscale structures directly. Techniques available for imaging include contact mode AFM, tapping mode AFM, Scanning Tunnelling AFM, Conductive AFM, and Scanning Capacitance Microscopy. With the Nanoman option, the X and Y deflecting piezo-elements are independently controlled, allowing for precise and direct placement of the tip anywhere within the field. Direct manipulation of particles on the surface is then possible by dragging the tip in the desired direction. This feedback control, coupled with direct control of conductive tip and substrate voltages, allows for direct-write oxidation on a variety of surfaces to create or modify nanostructures through local anodization.

Detailed Specifications

  • Conductive AFM modules for nA-microAmp current measurements
  • Tunneling AFM modulefor pA-nA current measurements
  • Scanning Capacitance Microscopy
  • Resolution: Sub-nm height-measurement capability; X-Y resolution tip dependent
  • Registration tolerance to a known mark: Field size dependent
  • Minimum substrate size: small pieces
  • Largest substrate size: 100 mm wafer
  • Oxidation Line widths: Call for info.

See Also