E-Beam Evaporation Recipes: Difference between revisions
Jump to navigation
Jump to search
Content deleted Content added
No edit summary |
|||
| Line 1: | Line 1: | ||
{{recipes|Vacuum Deposition}} |
{{recipes|Vacuum Deposition}} |
||
=[[E-Beam 1 (Sharon)]]= |
=[[E-Beam 1 (Sharon)]]= |
||
==Materials Table== |
|||
{| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center; font-size: 95%" class="collapsible wikitable" |
|||
|- |
|||
! colspan=8 width=1100 height=35 bgcolor="#D0E7FF" align="center"|<div style="font-size: 150%;">Materials Table</div> |
|||
|- bgcolor="#D0E7FF" |
|||
! width="75" bgcolor="#D0E7FF" align="center" | '''Material''' |
|||
! width="75" bgcolor="#D0E7FF" align="center" | '''Position''' |
|||
! width="75" bgcolor="#D0E7FF" align="center" | '''Hearth / Crucible''' |
|||
! width="85" bgcolor="#D0E7FF" align="center" | '''Film Number''' |
|||
! width="75" bgcolor="#D0E7FF" align="center" | '''Density''' |
|||
! width="75" bgcolor="#D0E7FF" align="center" | '''Z Ratio''' |
|||
! width="75" bgcolor="#D0E7FF" align="center" | '''Tooling''' |
|||
! width="500" bgcolor="#D0E7FF" align="center" | '''Comments''' |
|||
|- |
|||
| Ag |
|||
| 4 |
|||
| C |
|||
| 5 |
|||
| 10.5 |
|||
| 0.524 |
|||
| 140 |
|||
| Brian, could you please check this table? Z ratio should be 0.529 |
|||
|- |
|||
| Al |
|||
| 2 |
|||
| C |
|||
| 6 |
|||
| 2.7 |
|||
| 1.080 |
|||
| 118 |
|||
| |
|||
|- |
|||
| Al<sub>2</sub>O<sub>3</sub> |
|||
| 1 |
|||
| C |
|||
| 6 |
|||
| 3.97 |
|||
| 0.50 |
|||
| 169 |
|||
| Z-ratio should be 0.336 |
|||
|- |
|||
| Au |
|||
| 4 |
|||
| C |
|||
| 4 |
|||
| 19.3 |
|||
| 0.381 |
|||
| 138 |
|||
| Bazookas can be used at 20-30Å/sec. |
|||
|- |
|||
| AuGe |
|||
| 3 |
|||
| C |
|||
| 5 |
|||
| 17.63 |
|||
| 0.397 |
|||
| 151 |
|||
| Composition unpredictable unless you practically empty the crucible. |
|||
|- |
|||
| Cr |
|||
| 3 |
|||
| H |
|||
| 6 |
|||
| 7.2 |
|||
| 0.305 |
|||
| 140 |
|||
| Do not evaporate more than 200Å of Cr in the E-Beam evaporator. |
|||
|- |
|||
| Fe |
|||
| |
|||
| |
|||
| |
|||
| 7.86 |
|||
| 0.349 |
|||
| |
|||
| |
|||
|- |
|||
| Ge |
|||
| 3 |
|||
| C |
|||
| 6 |
|||
| 5.35 |
|||
| 0.516 |
|||
| 130 |
|||
| |
|||
|- |
|||
| MgO |
|||
| 1 |
|||
| |
|||
| 6 |
|||
| 3.58 |
|||
| 0.411 |
|||
| |
|||
| |
|||
|- |
|||
| Mo |
|||
| |
|||
| |
|||
| |
|||
| 10.2 |
|||
| 0.257 |
|||
| |
|||
| |
|||
|- |
|||
| Ni |
|||
| 1 |
|||
| H |
|||
| 1 |
|||
| 8.91 |
|||
| 0.331 |
|||
| 140 |
|||
| Prone to spitting. Cool down for 15 minutes before venting. |
|||
|- |
|||
| NiCr |
|||
| 1 |
|||
| H |
|||
| 6 |
|||
| 8.23 |
|||
| 0.321 |
|||
| |
|||
| Density 8.52, z ratio 0.3258 |
|||
|- |
|||
| Nb |
|||
| 4 |
|||
| C |
|||
| 6 |
|||
| 8.57 |
|||
| 0.516 ( should be 0.492) |
|||
| |
|||
| Cool down for at least 35 minutes before venting. |
|||
|- |
|||
| Pd |
|||
| 1 |
|||
| H |
|||
| 9 |
|||
| 12.0 |
|||
| 0.357 |
|||
| 140 |
|||
| |
|||
|- |
|||
| Pt |
|||
| 1 |
|||
| C |
|||
| 8 |
|||
| 21.40 |
|||
| 0.245 |
|||
| 140 |
|||
| Prone to spitting. Evaporate at 1.5Å/sec or less. |
|||
|- |
|||
| Si |
|||
| 2 |
|||
| H |
|||
| 2 |
|||
| 2.32 |
|||
| 0.712 |
|||
| 150 |
|||
| Cool down very slowly after evaporating lest you crack the source. |
|||
|- |
|||
| SiO |
|||
| |
|||
| C |
|||
| 6 |
|||
| 2.13 |
|||
| 0.87 |
|||
| 132 |
|||
| |
|||
|- |
|||
| SiO<sub>2</sub> |
|||
| 1 |
|||
| C |
|||
| 6 |
|||
| 2.2 ( should be 2.648) |
|||
| 1.07 ( should be 1.000) |
|||
| 140 |
|||
| Please change the crystal and the upper mirror after evaporating oxide. |
|||
|- |
|||
| SrF<sub>2</sub> |
|||
| 1 |
|||
| C |
|||
| 6 |
|||
| 4.28 |
|||
| 0.727 |
|||
| 140 |
|||
| |
|||
|- |
|||
| Ta |
|||
| 1 |
|||
| H |
|||
| 6 |
|||
| 16.6 |
|||
| 0.262 |
|||
| |
|||
| Requires extremely high current. Minimum 35 minute cool down. Hearth #3 may be used. Call me before you try Ta. |
|||
|- |
|||
| W |
|||
| 1 |
|||
| C |
|||
| 6 |
|||
| 19.3 |
|||
| 0.163 |
|||
| 138 |
|||
| |
|||
|- |
|||
| Ti |
|||
| 3 |
|||
| H |
|||
| 3 |
|||
| 4.50 |
|||
| 0.628 |
|||
| 139 |
|||
| |
|||
|} |
|||
=[[E-Beam 2 (Custom)]]= |
=[[E-Beam 2 (Custom)]]= |
||
=[[E-Beam 3 (Temescal)]]= |
=[[E-Beam 3 (Temescal)]]= |
||
Revision as of 17:58, 7 September 2012
Back to Vacuum Deposition Recipes.
E-Beam 1 (Sharon)
Materials Table
Materials Table
| |||||||
|---|---|---|---|---|---|---|---|
| Material | Position | Hearth / Crucible | Film Number | Density | Z Ratio | Tooling | Comments |
| Ag | 4 | C | 5 | 10.5 | 0.524 | 140 | Brian, could you please check this table? Z ratio should be 0.529 |
| Al | 2 | C | 6 | 2.7 | 1.080 | 118 | |
| Al2O3 | 1 | C | 6 | 3.97 | 0.50 | 169 | Z-ratio should be 0.336 |
| Au | 4 | C | 4 | 19.3 | 0.381 | 138 | Bazookas can be used at 20-30Å/sec. |
| AuGe | 3 | C | 5 | 17.63 | 0.397 | 151 | Composition unpredictable unless you practically empty the crucible. |
| Cr | 3 | H | 6 | 7.2 | 0.305 | 140 | Do not evaporate more than 200Å of Cr in the E-Beam evaporator. |
| Fe | 7.86 | 0.349 | |||||
| Ge | 3 | C | 6 | 5.35 | 0.516 | 130 | |
| MgO | 1 | 6 | 3.58 | 0.411 | |||
| Mo | 10.2 | 0.257 | |||||
| Ni | 1 | H | 1 | 8.91 | 0.331 | 140 | Prone to spitting. Cool down for 15 minutes before venting. |
| NiCr | 1 | H | 6 | 8.23 | 0.321 | Density 8.52, z ratio 0.3258 | |
| Nb | 4 | C | 6 | 8.57 | 0.516 ( should be 0.492) | Cool down for at least 35 minutes before venting. | |
| Pd | 1 | H | 9 | 12.0 | 0.357 | 140 | |
| Pt | 1 | C | 8 | 21.40 | 0.245 | 140 | Prone to spitting. Evaporate at 1.5Å/sec or less. |
| Si | 2 | H | 2 | 2.32 | 0.712 | 150 | Cool down very slowly after evaporating lest you crack the source. |
| SiO | C | 6 | 2.13 | 0.87 | 132 | ||
| SiO2 | 1 | C | 6 | 2.2 ( should be 2.648) | 1.07 ( should be 1.000) | 140 | Please change the crystal and the upper mirror after evaporating oxide. |
| SrF2 | 1 | C | 6 | 4.28 | 0.727 | 140 | |
| Ta | 1 | H | 6 | 16.6 | 0.262 | Requires extremely high current. Minimum 35 minute cool down. Hearth #3 may be used. Call me before you try Ta. | |
| W | 1 | C | 6 | 19.3 | 0.163 | 138 | |
| Ti | 3 | H | 3 | 4.50 | 0.628 | 139 | |