==Materials Table (E-Beam #3)== |
|||
{| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center; font-size: 95%" class="collapsible wikitable" |
|||
|- |
|||
! colspan=9 width=1300 height=35 bgcolor="#D0E7FF" align="center"|<div style="font-size: 150%;">Materials Table</div> |
|||
|- bgcolor="#D0E7FF" |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Material''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Gun''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Hearth /Crucible''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Process Gain, A/sec/%pwr''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Film Number''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Density, g/cm3''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Z Ratio''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Tooling, %''' |
|||
! width="100" bgcolor="#D0E7FF" align="center" | '''Comments''' |
|||
|- |
|||
|Ag |
|||
| Rear |
|||
| C |
|||
| 10.0 |
|||
| 2 |
|||
|10.50 |
|||
| 0.529 |
|||
| 67 |
|||
| |
|||
|- |
|||
| Al |
|||
| Rear |
|||
| C |
|||
| 10.0 |
|||
| 1 |
|||
|2.70 |
|||
| 1.080 |
|||
| 53 |
|||
| |
|||
|- |
|||
| Au |
|||
| Front |
|||
| C |
|||
| 2.0 |
|||
| 3 |
|||
|19.30 |
|||
| 0.381 |
|||
| 56 |
|||
| |
|||
|- |
|||
| Ge |
|||
| Rear |
|||
| C |
|||
| 10.0 |
|||
| 3 |
|||
|5.35 |
|||
| 0.516 |
|||
| 80 |
|||
| |
|||
|- |
|||
| Ni |
|||
| Front |
|||
| C |
|||
| 0.5 |
|||
| 2 |
|||
|8.91 |
|||
| 0.331 |
|||
| 67 |
|||
| |
|||
|- |
|||
| Pd |
|||
| Rear |
|||
| C |
|||
| 0.9 |
|||
| 4 |
|||
|12.038 |
|||
| 0.357 |
|||
| 48 |
|||
| |
|||
|- |
|||
| Pt |
|||
| Front |
|||
| C |
|||
| 0.4 |
|||
| 1 |
|||
|21.40 |
|||
| 0.245 |
|||
| 67 |
|||
| |
|||
|- |
|||
| Ti |
|||
| Front |
|||
| C |
|||
| 5.0 |
|||
| 4 |
|||
|4.50 |
|||
| 0.628 |
|||
| 67 |
|||
| |
|||
|- |
|||
| Si |
|||
| |
|||
| |
|||
| 10.0 |
|||
| |
|||
|2.32 |
|||
| 0.712 |
|||
| 100 |
|||
|Not available |
|||
|- |
|||
|} |
|||
=[[E-Beam 4 (CHA)]]= |
=[[E-Beam 4 (CHA)]]= |
||
==Materials Table (E-Beam #4)== |
|||
{| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center; font-size: 95%" class="collapsible wikitable" |
|||
|- |
|||
! colspan=6 width=1300 height=35 bgcolor="#D0E7FF" align="center"|<div style="font-size: 150%;">Materials Table</div> |
|||
|- bgcolor="#D0E7FF" |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Material''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Density, g/cm3''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Z Ratio''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Master tooling, %''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Process Gain, A/sec/%pwr''' |
|||
! width="100" bgcolor="#D0E7FF" align="center" | '''Comments''' |
|||
|- |
|||
|Ag |
|||
| 10.50 |
|||
| 0.529 |
|||
|110 |
|||
|10.0 |
|||
|Tony, could you please check this? |
|||
|- |
|||
|Al |
|||
| 2.70 |
|||
| 1.080 |
|||
|165 |
|||
|6.0 |
|||
| |
|||
|- |
|||
|Au |
|||
|19.30 |
|||
| 0.381 |
|||
|120 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Co |
|||
|8.90 |
|||
| 0.343 |
|||
|150 |
|||
|5.0 |
|||
| |
|||
|- |
|||
|Cr |
|||
|7.20 |
|||
| 0.305 |
|||
|140 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Fe |
|||
|7.86 |
|||
| 0.349 |
|||
|165 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Ge |
|||
|5.35 |
|||
| 0.516 |
|||
|126 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Hf |
|||
|13.09 |
|||
| 0.360 |
|||
|150 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Ir |
|||
|22.40 |
|||
| 0.129 |
|||
|130 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Ni |
|||
|8.91 |
|||
| 0.331 |
|||
|150 |
|||
|5.0 |
|||
| |
|||
|- |
|||
|NiCr |
|||
|8.23 |
|||
| 0.321 |
|||
|140 |
|||
|10.0 |
|||
|density should be 8.50, z ratio 0.3258 |
|||
|- |
|||
|NiFe |
|||
|8.70 |
|||
|1.000 |
|||
|100 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Pd |
|||
|12.038 |
|||
|0.357 |
|||
|112 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Pt |
|||
|21.40 |
|||
|0.245 |
|||
|130 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Ru |
|||
|12.362 |
|||
|0.182 |
|||
|100 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Ti |
|||
|4.50 |
|||
|0.628 |
|||
|183 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|Zr |
|||
|6.49 |
|||
|0.600 |
|||
|150 |
|||
|10.0 |
|||
| |
|||
|- |
|||
|} |
|||
Revision as of 18:02, 7 September 2012
Back to Vacuum Deposition Recipes.
E-Beam 1 (Sharon)
Materials Table (E-Beam #1)
Materials Table
| |||||||
|---|---|---|---|---|---|---|---|
| Material | Position | Hearth / Crucible | Film Number | Density | Z Ratio | Tooling | Comments |
| Ag | 4 | C | 5 | 10.5 | 0.524 | 140 | Brian, could you please check this table? Z ratio should be 0.529 |
| Al | 2 | C | 6 | 2.7 | 1.080 | 118 | |
| Al2O3 | 1 | C | 6 | 3.97 | 0.50 | 169 | Z-ratio should be 0.336 |
| Au | 4 | C | 4 | 19.3 | 0.381 | 138 | Bazookas can be used at 20-30Å/sec. |
| AuGe | 3 | C | 5 | 17.63 | 0.397 | 151 | Composition unpredictable unless you practically empty the crucible. |
| Cr | 3 | H | 6 | 7.2 | 0.305 | 140 | Do not evaporate more than 200Å of Cr in the E-Beam evaporator. |
| Fe | 7.86 | 0.349 | |||||
| Ge | 3 | C | 6 | 5.35 | 0.516 | 130 | |
| MgO | 1 | 6 | 3.58 | 0.411 | |||
| Mo | 10.2 | 0.257 | |||||
| Ni | 1 | H | 1 | 8.91 | 0.331 | 140 | Prone to spitting. Cool down for 15 minutes before venting. |
| NiCr | 1 | H | 6 | 8.23 | 0.321 | Density 8.52, z ratio 0.3258 | |
| Nb | 4 | C | 6 | 8.57 | 0.516 ( should be 0.492) | Cool down for at least 35 minutes before venting. | |
| Pd | 1 | H | 9 | 12.0 | 0.357 | 140 | |
| Pt | 1 | C | 8 | 21.40 | 0.245 | 140 | Prone to spitting. Evaporate at 1.5Å/sec or less. |
| Si | 2 | H | 2 | 2.32 | 0.712 | 150 | Cool down very slowly after evaporating lest you crack the source. |
| SiO | C | 6 | 2.13 | 0.87 | 132 | ||
| SiO2 | 1 | C | 6 | 2.2 ( should be 2.648) | 1.07 ( should be 1.000) | 140 | Please change the crystal and the upper mirror after evaporating oxide. |
| SrF2 | 1 | C | 6 | 4.28 | 0.727 | 140 | |
| Ta | 1 | H | 6 | 16.6 | 0.262 | Requires extremely high current. Minimum 35 minute cool down. Hearth #3 may be used. Call me before you try Ta. | |
| W | 1 | C | 6 | 19.3 | 0.163 | 138 | |
| Ti | 3 | H | 3 | 4.50 | 0.628 | 139 | |
E-Beam 2 (Custom)
Materials Table (E-Beam #2)
Materials Table
| ||||
|---|---|---|---|---|
| Material | Density, g/cm3 | Z Ratio | Tooling factor, % | Comments |
| Al2O3 | 3.97 | 0.336 | 140.0 | Tony could you please check this? |
| GeO2 | 6.24 | 1.000 | 139.0 | |
| ITO | 7.16 | 1.000 | 139.0 | Density should be 6.43-7.14, z ratio? |
| MgO | 3.58 | 0.411 | 157.6 | OK |
| Si | 2.32 | 0.712 | 150.0 | |
| SiO2 | 2.20 | 1.070 | 157.6 | Should be 2.648, z ratio 1.000 |
| SiOx | 2.13 | 0.87 | 130.0 | |
| SrF2 | 4.28 | 0.727 | 140.0 | |
| Ta2O5 | 8.2 | 0.30 | 157.6 | |
| TiO2 | 4.26 | 0.400 | 139.0 | |
E-Beam 3 (Temescal)
Materials Table (E-Beam #3)
Materials Table
| ||||||||
|---|---|---|---|---|---|---|---|---|
| Material | Gun | Hearth /Crucible | Process Gain, A/sec/%pwr | Film Number | Density, g/cm3 | Z Ratio | Tooling, % | Comments |
| Ag | Rear | C | 10.0 | 2 | 10.50 | 0.529 | 67 | |
| Al | Rear | C | 10.0 | 1 | 2.70 | 1.080 | 53 | |
| Au | Front | C | 2.0 | 3 | 19.30 | 0.381 | 56 | |
| Ge | Rear | C | 10.0 | 3 | 5.35 | 0.516 | 80 | |
| Ni | Front | C | 0.5 | 2 | 8.91 | 0.331 | 67 | |
| Pd | Rear | C | 0.9 | 4 | 12.038 | 0.357 | 48 | |
| Pt | Front | C | 0.4 | 1 | 21.40 | 0.245 | 67 | |
| Ti | Front | C | 5.0 | 4 | 4.50 | 0.628 | 67 | |
| Si | 10.0 | 2.32 | 0.712 | 100 | Not available | |||
E-Beam 4 (CHA)
Materials Table (E-Beam #4)
Materials Table
| |||||
|---|---|---|---|---|---|
| Material | Density, g/cm3 | Z Ratio | Master tooling, % | Process Gain, A/sec/%pwr | Comments |
| Ag | 10.50 | 0.529 | 110 | 10.0 | Tony, could you please check this? |
| Al | 2.70 | 1.080 | 165 | 6.0 | |
| Au | 19.30 | 0.381 | 120 | 10.0 | |
| Co | 8.90 | 0.343 | 150 | 5.0 | |
| Cr | 7.20 | 0.305 | 140 | 10.0 | |
| Fe | 7.86 | 0.349 | 165 | 10.0 | |
| Ge | 5.35 | 0.516 | 126 | 10.0 | |
| Hf | 13.09 | 0.360 | 150 | 10.0 | |
| Ir | 22.40 | 0.129 | 130 | 10.0 | |
| Ni | 8.91 | 0.331 | 150 | 5.0 | |
| NiCr | 8.23 | 0.321 | 140 | 10.0 | density should be 8.50, z ratio 0.3258 |
| NiFe | 8.70 | 1.000 | 100 | 10.0 | |
| Pd | 12.038 | 0.357 | 112 | 10.0 | |
| Pt | 21.40 | 0.245 | 130 | 10.0 | |
| Ru | 12.362 | 0.182 | 100 | 10.0 | |
| Ti | 4.50 | 0.628 | 183 | 10.0 | |
| Zr | 6.49 | 0.600 | 150 | 10.0 | |