PECVD 2 (Advanced Vacuum)

From UCSB Nanofab Wiki
Revision as of 17:56, 29 September 2025 by John d (talk | contribs) (Recipes & Data)
Jump to navigation Jump to search
PECVD 2 (Advanced Vacuum)
Location Bay 2
Tool Type Vacuum Deposition
Manufacturer Plasma-Therm
Description Vision 310 Advanced Vacuum PECVD

Primary Supervisor Michael Barreraz
(805) 893-4147
mikebarreraz@ece.ucsb.edu

Secondary Supervisor

Don Freeborn


Recipes Vacuum Deposition Recipes

SignupMonkey: Sign up for this tool


About

  • Films/Gases: This open-load system is dedicated to PECVD of SiO2, SiNx, SiOxNy, and a-Si using Silane (2%SiH4, 98% He), N2O, NH3, and N2 gases.
  • Size: The sample electrode has a 270mm diameter useable area, allowing for multiple 4” wafer depositions in a single run.
  • Temperature: Standard operating temperature is 300C, but can be user changed for temps ranging anywhere from 250 to 350C.
  • Low-Stress Si3N4: The system is equipped with a dual generator, dual frequency option for growth of Low-stress Nitride films.
    • These films alternate between thin (<10nm) compressive and tensile layers.
    • The Low-Stress Si3N4 film recipe are tested approx. monthly, and kept within ±100MPa. Data can be found at Recipes (below) > Low-Stress Nitride.

See Also

Documentation

Recipes & Data

SPC chart example
Example Process Control Charts (SPC) for thin-film DepCals.
screenshot of surfscan particle count
Example particle counts taken on each film.
Screenshot of Filmetrics F50 wafermap of typical DepCals film
Example of DepCals Thickness/Refractive Index uniformity measurement (4% shown here).

Process Control Data