ASML Stepper 3 Error Recovery, Troubleshooting and Calibration
Jump to navigation
Jump to search
Error, Calibration, Error logging Procedures for Users
Error Log full
- Close logging window
- Open new console window: Rt click on Desktop, select (ASML)
- At % prompt, type logging (all lowercase!)
- Resize window and place in upper right area of screen as normal.
IQC (Image Quality Control) Calibration Check
- Main Menu > Test Manager > Run Tests > Metrology Verification > Image Quality Control
- Accept (Upper Button)
- Perform Measurement
- Choose System Correction
- # Cycles = 1
- # Measurements = 3
- # Corrections = 1
- M "Measure" mode only (does Not update internal lookup tables)
- (UCSB numbers for test 1 3 1 M)
- Accept to run the test
- If | focus correction | < 50nm, system OK, note number in log book
- If | focus correction | > 50nm, run test again, if number keeps dropping, keep running until number stable
- If | focus correction | is still > 50nm, discuss with engineer
Wafer Handler Issues during batch operation
- Lift black cover behind keyboard
- Press AWH button for 1 sec and release (resets Wafer Handler)
- Wait until wafer handler “system booting” message is done and warning window in bottom left of main computer console has disappeared. Log window should say it booted succesfully. Takes a few minutes.
- (DO NOT do FAST STARTUP from error message, go to main menu!)
- Exit to main menu
- Go to Start/Stop
- Fast Startup
- Start - will reinitialize the wafer handler system
- Once complete, back to main menu
- If wafers in machine, go to material handler (Mat Hdl).
- Go to Remove Wafers menu and Collect Wafers
- Back to main menu when finished.
- If system says wafer is missing, contact engineer
Laser Refill (or laser calling for gas injection)
- Main Menu > Test Manager
- Open Cmd Handler (Cmd Hdl) (Top Menu)
- Illumination and Projection > Illumination
- New Laser Fill
- Use Laser Start instead if laser asking for gas injection
- Illumination and Projection > Illumination