File:Filmmetrics F50.jpg
Tool Type Inspection, Test and Characterization
Location Bay 6
Supervisor Ning Cao
Supervisor Phone (805) 893-4689
Supervisor E-Mail ningcao@ece.ucsb.edu
Description Thin-Film Reflectometry
Manufacturer Filmetrics
Model Filmetrics F50


About

This tool is for thickness and optical property measurements of films on substrates, including mapping of thin-films on full wafers. The technique used is white light reflection. Data is taken with normal incidence reflection of white light (190 nm – 1700 nm) from the surface using a Deuterium (UV) and Halogen (Vis-nIR) lamps. The motorized stage automatically acquires optical reflection spectra at programmed points across the wafer. The data is modeled at each point, and the optical parameters (thickness/refractive index/absorption) are adjusted to give a best least-squared fit to the data. The accuracy of the technique will depend on the thickness of the film and the optical models used for the fitting of the data. For a more complete description go to Filmetrics.

Equipment Specifications

  • 190-1700 nm reflection spectrum
  • 10 Å to 150 µm thickness, n, and k measurements
  • Accepts wafers up to 150mm for motorized mapping.
  • Small substrates also possible, ≥ 2-3mm
  • Data can be saved
  • Can model up to three layers with accuracy