Test Data of etching SiO2 with CHF3/CF4-ICP1
| Date | Sample# | Etch Rate (nm/min) | Etch Selectivity (SiO2/PR) | Comments | SEM Images |
| 11/18/2022 | ND_Pan1_111822 | 132.6 | 1.25 | [30D][CS] | |
| 11/07/2022 | ND_Pan1_110722 | 125.7 | 1.35 | [30D] [CS] | |
| 10/21/2022 | ND_Pan1_102122 | 114.3 | 1.15 | [30D] [CS] | |
| 10/10/2022 | ND_Pan1_101022 | 127.4 | 1.25 | [30D] [CS] | |
| 10/3/2022 | ND_Pan1_100322 | 132.9 | 1.19 | High etch rate and selectivity.
May be due to angled ridge. |
[30D] [CS] |
| 9/26/2022 | ND_Pan1n_092622 | 111.1 | 0.90 | This is the new wafer | [30D] [CS] |
| 9/26/2022 | ND_Pan1o_092622 | 114.3 | 1.01 | This is the old wafer | [45D] [CS] |
| 9/12/2022 | ND_Pan1_091222 | 128.6 | 1.33 | New Si wafer, higher etch rate/
selectivity |
[30D] [CS] |
| 9/01/2022 | ND_Pan1_090122 | 120.9 | 1.05 | [30D] [CS] | |
| 8/22/2022 | ND_Pan1_082222 | 112.9 | 1.08 | Etch Rate seems low
and selectivity on the lower side |
[45D] [CS] |
| 8/5/2022 | ND_Pan 1_080522 | 115.1 | 0.90 | [30D] [CS] | |
| 7/27/2022 | ND_Pan1_072722 | 134.6 | 1.12 | [CS] [30D] | |
| 5/10/2022 | NP_ICP1_07 | 136 | 1.34 | [1] [2] | |
| 4/26/2022 | NP_ICP1_06 | 139.4 | 1.36 | [1] [2] | |
| 4/20/2022 | NP_ICP1_05 | 140 | 1.13 | [1] [2] | |
| 4/13/2022 | NP_ICP1_04 | 140.3 | 1.24 | [1] [2] | |
| 3/29/2022 | NP_ICP1_03 | 136.9 | 1.19 | [1][2] | |
| 3/8/2022 | NP_ICP1_02 | 133.7 | 1.12 | [1] [2] | |
| 3/2/2022 | NP_ICP1_01 | 141.4 | 1.20 | [1] [2] | |
| 1/7/2021 | I12101 | 118 | 1.12 | [1] | |
| 3/3/2020 | I12004 | 110 | 1.05 | [2] | |
| Sidewall profile on 2/28/2020 etches is slanted, chamber will be wet-cleaned on 3/2/2020 and re-tested. Data for 3/3 shows etch returned to "normal". | |||||
| 2/28/2020 | I12003 | 119 | 1.17 | 56.6 | [3] |
| 1/23/2020 | I12002 | 109 | 1.16 | [4] | |
| Rate on 1/13/2020 is low by ~20%, so chamber was wet-cleaned on 1/21/2020. Data below for 1/23 shows rate returned to "normal". | |||||
| 1/13/2020 | I12001 | 78.0 | 1.06 | unusual - two regions | [5] |
| 5/29/2019 | I11903 | 105 | 1.41 | [6] | |
| 1/28/2019 | I11901 | 110 | 1.35 | see SEM → | [7] |
OLD Etch Test Data
Alternate SiO2 etch recipe with O2 included.
| ICP#1: 0.5Pa, 50/500W, CHF3/CF4/O2=35/5/10sccm, time=210 sec | |||||
| Date | Sample# | Etch Rate (nm/min) | Etch Selectivity (SiO2/PR) | Averaged Sidewall Angle (o) | SEM Images |
| 1/28/2019 | I11902 | 78.1 | 0.63 | [8] | |
| 5/29/2019 | I11904 | 71.1 | 0.58 | [9] | |