Difference between revisions of "E-Beam Evaporation Recipes"
Jump to navigation
Jump to search
Line 289: | Line 289: | ||
=[[E-Beam 3 (Temescal)]]= |
=[[E-Beam 3 (Temescal)]]= |
||
+ | ==Materials Table (E-Beam #3)== |
||
+ | {| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center; font-size: 95%" class="collapsible wikitable" |
||
+ | |- |
||
+ | ! colspan=9 width=1300 height=35 bgcolor="#D0E7FF" align="center"|<div style="font-size: 150%;">Materials Table</div> |
||
+ | |- bgcolor="#D0E7FF" |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Material''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Gun''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Hearth /Crucible''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Process Gain, A/sec/%pwr''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Film Number''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Density, g/cm3''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Z Ratio''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Tooling, %''' |
||
+ | ! width="100" bgcolor="#D0E7FF" align="center" | '''Comments''' |
||
+ | |- |
||
+ | |Ag |
||
+ | | Rear |
||
+ | | C |
||
+ | | 10.0 |
||
+ | | 2 |
||
+ | |10.50 |
||
+ | | 0.529 |
||
+ | | 67 |
||
+ | | |
||
+ | |- |
||
+ | | Al |
||
+ | | Rear |
||
+ | | C |
||
+ | | 10.0 |
||
+ | | 1 |
||
+ | |2.70 |
||
+ | | 1.080 |
||
+ | | 53 |
||
+ | | |
||
+ | |- |
||
+ | | Au |
||
+ | | Front |
||
+ | | C |
||
+ | | 2.0 |
||
+ | | 3 |
||
+ | |19.30 |
||
+ | | 0.381 |
||
+ | | 56 |
||
+ | | |
||
+ | |- |
||
+ | | Ge |
||
+ | | Rear |
||
+ | | C |
||
+ | | 10.0 |
||
+ | | 3 |
||
+ | |5.35 |
||
+ | | 0.516 |
||
+ | | 80 |
||
+ | | |
||
+ | |- |
||
+ | | Ni |
||
+ | | Front |
||
+ | | C |
||
+ | | 0.5 |
||
+ | | 2 |
||
+ | |8.91 |
||
+ | | 0.331 |
||
+ | | 67 |
||
+ | | |
||
+ | |- |
||
+ | | Pd |
||
+ | | Rear |
||
+ | | C |
||
+ | | 0.9 |
||
+ | | 4 |
||
+ | |12.038 |
||
+ | | 0.357 |
||
+ | | 48 |
||
+ | | |
||
+ | |- |
||
+ | | Pt |
||
+ | | Front |
||
+ | | C |
||
+ | | 0.4 |
||
+ | | 1 |
||
+ | |21.40 |
||
+ | | 0.245 |
||
+ | | 67 |
||
+ | | |
||
+ | |- |
||
+ | | Ti |
||
+ | | Front |
||
+ | | C |
||
+ | | 5.0 |
||
+ | | 4 |
||
+ | |4.50 |
||
+ | | 0.628 |
||
+ | | 67 |
||
+ | | |
||
+ | |- |
||
+ | | Si |
||
+ | | |
||
+ | | |
||
+ | | 10.0 |
||
+ | | |
||
+ | |2.32 |
||
+ | | 0.712 |
||
+ | | 100 |
||
+ | |Not available |
||
+ | |- |
||
+ | |} |
||
=[[E-Beam 4 (CHA)]]= |
=[[E-Beam 4 (CHA)]]= |
||
+ | ==Materials Table (E-Beam #4)== |
||
+ | {| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center; font-size: 95%" class="collapsible wikitable" |
||
+ | |- |
||
+ | ! colspan=6 width=1300 height=35 bgcolor="#D0E7FF" align="center"|<div style="font-size: 150%;">Materials Table</div> |
||
+ | |- bgcolor="#D0E7FF" |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Material''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Density, g/cm3''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Z Ratio''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Master tooling, %''' |
||
+ | ! width="45" bgcolor="#D0E7FF" align="center" | '''Process Gain, A/sec/%pwr''' |
||
+ | ! width="100" bgcolor="#D0E7FF" align="center" | '''Comments''' |
||
+ | |- |
||
+ | |Ag |
||
+ | | 10.50 |
||
+ | | 0.529 |
||
+ | |110 |
||
+ | |10.0 |
||
+ | |Tony, could you please check this? |
||
+ | |- |
||
+ | |Al |
||
+ | | 2.70 |
||
+ | | 1.080 |
||
+ | |165 |
||
+ | |6.0 |
||
+ | | |
||
+ | |- |
||
+ | |Au |
||
+ | |19.30 |
||
+ | | 0.381 |
||
+ | |120 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Co |
||
+ | |8.90 |
||
+ | | 0.343 |
||
+ | |150 |
||
+ | |5.0 |
||
+ | | |
||
+ | |- |
||
+ | |Cr |
||
+ | |7.20 |
||
+ | | 0.305 |
||
+ | |140 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Fe |
||
+ | |7.86 |
||
+ | | 0.349 |
||
+ | |165 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Ge |
||
+ | |5.35 |
||
+ | | 0.516 |
||
+ | |126 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Hf |
||
+ | |13.09 |
||
+ | | 0.360 |
||
+ | |150 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Ir |
||
+ | |22.40 |
||
+ | | 0.129 |
||
+ | |130 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Ni |
||
+ | |8.91 |
||
+ | | 0.331 |
||
+ | |150 |
||
+ | |5.0 |
||
+ | | |
||
+ | |- |
||
+ | |NiCr |
||
+ | |8.23 |
||
+ | | 0.321 |
||
+ | |140 |
||
+ | |10.0 |
||
+ | |density should be 8.50, z ratio 0.3258 |
||
+ | |- |
||
+ | |NiFe |
||
+ | |8.70 |
||
+ | |1.000 |
||
+ | |100 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Pd |
||
+ | |12.038 |
||
+ | |0.357 |
||
+ | |112 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Pt |
||
+ | |21.40 |
||
+ | |0.245 |
||
+ | |130 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Ru |
||
+ | |12.362 |
||
+ | |0.182 |
||
+ | |100 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Ti |
||
+ | |4.50 |
||
+ | |0.628 |
||
+ | |183 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |Zr |
||
+ | |6.49 |
||
+ | |0.600 |
||
+ | |150 |
||
+ | |10.0 |
||
+ | | |
||
+ | |- |
||
+ | |} |
Revision as of 11:02, 7 September 2012
Back to Vacuum Deposition Recipes.
E-Beam 1 (Sharon)
Materials Table (E-Beam #1)
Materials Table
| |||||||
---|---|---|---|---|---|---|---|
Material | Position | Hearth / Crucible | Film Number | Density | Z Ratio | Tooling | Comments |
Ag | 4 | C | 5 | 10.5 | 0.524 | 140 | Brian, could you please check this table? Z ratio should be 0.529 |
Al | 2 | C | 6 | 2.7 | 1.080 | 118 | |
Al2O3 | 1 | C | 6 | 3.97 | 0.50 | 169 | Z-ratio should be 0.336 |
Au | 4 | C | 4 | 19.3 | 0.381 | 138 | Bazookas can be used at 20-30Å/sec. |
AuGe | 3 | C | 5 | 17.63 | 0.397 | 151 | Composition unpredictable unless you practically empty the crucible. |
Cr | 3 | H | 6 | 7.2 | 0.305 | 140 | Do not evaporate more than 200Å of Cr in the E-Beam evaporator. |
Fe | 7.86 | 0.349 | |||||
Ge | 3 | C | 6 | 5.35 | 0.516 | 130 | |
MgO | 1 | 6 | 3.58 | 0.411 | |||
Mo | 10.2 | 0.257 | |||||
Ni | 1 | H | 1 | 8.91 | 0.331 | 140 | Prone to spitting. Cool down for 15 minutes before venting. |
NiCr | 1 | H | 6 | 8.23 | 0.321 | Density 8.52, z ratio 0.3258 | |
Nb | 4 | C | 6 | 8.57 | 0.516 ( should be 0.492) | Cool down for at least 35 minutes before venting. | |
Pd | 1 | H | 9 | 12.0 | 0.357 | 140 | |
Pt | 1 | C | 8 | 21.40 | 0.245 | 140 | Prone to spitting. Evaporate at 1.5Å/sec or less. |
Si | 2 | H | 2 | 2.32 | 0.712 | 150 | Cool down very slowly after evaporating lest you crack the source. |
SiO | C | 6 | 2.13 | 0.87 | 132 | ||
SiO2 | 1 | C | 6 | 2.2 ( should be 2.648) | 1.07 ( should be 1.000) | 140 | Please change the crystal and the upper mirror after evaporating oxide. |
SrF2 | 1 | C | 6 | 4.28 | 0.727 | 140 | |
Ta | 1 | H | 6 | 16.6 | 0.262 | Requires extremely high current. Minimum 35 minute cool down. Hearth #3 may be used. Call me before you try Ta. | |
W | 1 | C | 6 | 19.3 | 0.163 | 138 | |
Ti | 3 | H | 3 | 4.50 | 0.628 | 139 |
E-Beam 2 (Custom)
Materials Table (E-Beam #2)
Materials Table
| ||||
---|---|---|---|---|
Material | Density, g/cm3 | Z Ratio | Tooling factor, % | Comments |
Al2O3 | 3.97 | 0.336 | 140.0 | Tony could you please check this? |
GeO2 | 6.24 | 1.000 | 139.0 | |
ITO | 7.16 | 1.000 | 139.0 | Density should be 6.43-7.14, z ratio? |
MgO | 3.58 | 0.411 | 157.6 | OK |
Si | 2.32 | 0.712 | 150.0 | |
SiO2 | 2.20 | 1.070 | 157.6 | Should be 2.648, z ratio 1.000 |
SiOx | 2.13 | 0.87 | 130.0 | |
SrF2 | 4.28 | 0.727 | 140.0 | |
Ta2O5 | 8.2 | 0.30 | 157.6 | |
TiO2 | 4.26 | 0.400 | 139.0 |
E-Beam 3 (Temescal)
Materials Table (E-Beam #3)
Materials Table
| ||||||||
---|---|---|---|---|---|---|---|---|
Material | Gun | Hearth /Crucible | Process Gain, A/sec/%pwr | Film Number | Density, g/cm3 | Z Ratio | Tooling, % | Comments |
Ag | Rear | C | 10.0 | 2 | 10.50 | 0.529 | 67 | |
Al | Rear | C | 10.0 | 1 | 2.70 | 1.080 | 53 | |
Au | Front | C | 2.0 | 3 | 19.30 | 0.381 | 56 | |
Ge | Rear | C | 10.0 | 3 | 5.35 | 0.516 | 80 | |
Ni | Front | C | 0.5 | 2 | 8.91 | 0.331 | 67 | |
Pd | Rear | C | 0.9 | 4 | 12.038 | 0.357 | 48 | |
Pt | Front | C | 0.4 | 1 | 21.40 | 0.245 | 67 | |
Ti | Front | C | 5.0 | 4 | 4.50 | 0.628 | 67 | |
Si | 10.0 | 2.32 | 0.712 | 100 | Not available |
E-Beam 4 (CHA)
Materials Table (E-Beam #4)
Materials Table
| |||||
---|---|---|---|---|---|
Material | Density, g/cm3 | Z Ratio | Master tooling, % | Process Gain, A/sec/%pwr | Comments |
Ag | 10.50 | 0.529 | 110 | 10.0 | Tony, could you please check this? |
Al | 2.70 | 1.080 | 165 | 6.0 | |
Au | 19.30 | 0.381 | 120 | 10.0 | |
Co | 8.90 | 0.343 | 150 | 5.0 | |
Cr | 7.20 | 0.305 | 140 | 10.0 | |
Fe | 7.86 | 0.349 | 165 | 10.0 | |
Ge | 5.35 | 0.516 | 126 | 10.0 | |
Hf | 13.09 | 0.360 | 150 | 10.0 | |
Ir | 22.40 | 0.129 | 130 | 10.0 | |
Ni | 8.91 | 0.331 | 150 | 5.0 | |
NiCr | 8.23 | 0.321 | 140 | 10.0 | density should be 8.50, z ratio 0.3258 |
NiFe | 8.70 | 1.000 | 100 | 10.0 | |
Pd | 12.038 | 0.357 | 112 | 10.0 | |
Pt | 21.40 | 0.245 | 130 | 10.0 | |
Ru | 12.362 | 0.182 | 100 | 10.0 | |
Ti | 4.50 | 0.628 | 183 | 10.0 | |
Zr | 6.49 | 0.600 | 150 | 10.0 |