Difference between revisions of "Wet Etching Recipes"
(→The Master Table of Wet Etching (Include All Materials): corrected "BHF 49%" --> "BHF") |
(→Gold Plating Bench (Technic SEMCON 1000): added basic recipe steps for seed/plate/etch) |
||
(32 intermediate revisions by 3 users not shown) | |||
Line 1: | Line 1: | ||
+ | __TOC__ |
||
− | == References == |
||
− | # [http://ieeexplore.ieee.org/abstract/document/546406/ Etch rates for Micromachining Processing (IEEE Jnl. MEMS, 1996)] - includes tables of etch rates of numerous metals vs. various wet and dry etchants. |
||
− | # [http://ieeexplore.ieee.org/abstract/document/1257354/ Etch rates for micromachining-Part II (IEEE Jnl. MEMS, 2003)] - expanded tables containing resists, dielectrics, metals and semiconductors vs. many wet etch chemicals. |
||
− | # [http://www.sciencedirect.com/science/article/pii/S0927796X00000279 Guide to references on III±V semiconductor chemical etching] - exhaustive list of wet etchants for etching various semiconductors, including selective etches. |
||
− | # [http://transene.com/etch-compatibility/ Transene's Chemical Compatibility Chart] provides a useful quick-reference for which Transene etchants attack which materials. |
||
− | ## As a side-note, [http://transene.com/ Transene] provides many pre-mixed solutions that you can order, saving you the time and uncertainty of measuring/mixing such chemicals yourself. Make sure you check with us before ordering so we know how to handle the chemical before it arrives. |
||
+ | ==Chemicals Available== |
||
− | ==Compound Semiconductor Etching== |
||
− | [http://www.sciencedirect.com/science/article/pii/S0927796X00000279 Guide to references on III±V semiconductor chemical etching] |
||
+ | *[[Chemical List|'''The Chemical Lists''']] show stocked chemicals, photolithography chemicals, and how to bring new chemicals. |
||
− | Please add any confirmed etches from this reference to the {{HLink|Wet Etching Recipes|The Master Table of Wet Etching (Include All Materials)}}. |
||
− | |||
− | ==Metal Etching== |
||
− | *[[media:Ta and Cr E-beam deposition and wet etch test.pdf|Selective Wet Etch of Cr over Ta using Cr Etchant]] |
||
− | *[[media:ITO Deposition-250C-Ebeam2-HCl-Wet-Etch.pdf|Wet Etch of ITO using Heated, Diluted HCl Solution]] |
||
− | |||
− | ==Silicon etching== |
||
− | [http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=546406 Etch rates for micromachining processing] |
||
− | |||
− | [http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=1257354 Etch rates for micromachining processing-part II] |
||
− | |||
− | Please add any confirmed etches from this reference to the {{HLink|Wet Etching Recipes|The Master Table of Wet Etching (Include All Materials)}}. |
||
− | |||
− | ==Dielectric etching== |
||
− | |||
− | ==Organic removal== |
||
− | |||
− | ==Gold Plating== |
||
− | |||
− | ==Chemi-Mechanical Polishing (CMP)== |
||
− | |||
− | |||
− | ==Example Wet Etching Table== |
||
− | How to use the Master Table of Wet Etching: |
||
− | |||
− | When entering a new etch into the table make a row for every etchant used in the solution such that the information can be sorted by etchant. For example, the InP etch HCl:H3PO4(1:3) and H3PO4:HCl(3:1). Likewise, if etch is known to be selective to multiple materials the etch should have a row for each material. For example HCl:H3PO4(1:3) is selective to both InGaAs and InGaAsP. |
||
− | |||
− | This multiple entry method may seem laborious for the person entering a new etch, however the power of sorting by selective materials and chemicals in a table with all materials is great. |
||
+ | ==Table of Wet Etching Recipes== |
||
+ | ''Use the ↑ ↓ Arrows in the header row to sort the entire table by material, selectivity, etchant etc.'' |
||
{| class="wikitable sortable" |
{| class="wikitable sortable" |
||
|- |
|- |
||
− | ! |
+ | !Material!!Etchant!!Rate (nm/min)!!Anisotropy!!Selective to!!Selectivity!!Ref.!!Notes!!Confirmed By/Date |
|- |
|- |
||
+ | |Photoresist, polymers/organics |
||
− | | InP|| HCl:H3PO4(1:3)|| ~1000 || Highly || InGaAs || High || [http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] || Example || Jon Doe || Example |
||
+ | |H2SO4:H2O2 = 3:1 |
||
+ | [[Wet Etching Recipes#Organic%20removal|Piranha Solution]] |
||
+ | |typ. 5-10min etch for polymer residue |
||
+ | | |
||
+ | |Cr, W, Au, Pt, Si, SiO2, SiN |
||
+ | | |
||
+ | | |
||
+ | |'''''Dangerous''''' boiling hazard - see '''''[[Wet Etching Recipes#Piranha%20Solution|Piranha Solution]]''''' |
||
+ | section below. Etches Ti |
||
+ | |[[Demis D. John]], 2017 |
||
|- |
|- |
||
+ | |InP |
||
− | | InP|| HCl:H3PO4(1:3)|| ~1000 || Highly || InGaAsP || High || [http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] || Example || Jon Doe || Example |
||
+ | |H3PO4:HCl = 3:1 |
||
+ | |~1000 |
||
+ | |Highly |
||
+ | |InGaAsP |
||
+ | |High |
||
+ | |[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] |
||
+ | | |
||
+ | | |
||
|- |
|- |
||
+ | |InGaAsP |
||
− | | InP|| H3PO4:HCl(3:1)|| ~1000 || Highly || InGaAs || High || [http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] || Example || Jon Doe || Example |
||
+ | |''To Be Added'' |
||
+ | | |
||
+ | | |
||
+ | |InP |
||
+ | |High |
||
+ | | |
||
+ | | |
||
+ | | |
||
|- |
|- |
||
− | | |
+ | |InP||H3PO4:HCl = 3:1||~1000||Highly||InGaAs||High||[http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)]|| |
− | | |
+ | | |
− | |||
− | ==The Master Table of Wet Etching (Include '''All''' Materials)== |
||
− | |||
− | {| class="wikitable sortable" |
||
|- |
|- |
||
+ | |InP |
||
− | ! Material !! Etchant !! Rate (nm/min) !! Anisotropy !! Selective to !! Selectivity !! Ref. !! Notes !! Confirmed by !! Extra column |
||
+ | |HCl:H2O = 3:1 |
||
+ | |~5000 |
||
+ | | |
||
+ | |InGaAs |
||
+ | ~200nm stop-etch |
||
+ | |High |
||
+ | | |
||
+ | |Bubbles while etching |
||
+ | | |
||
|- |
|- |
||
+ | |InGaAs |
||
− | | InP|| HCl:H3PO4(1:3)|| ~1000 || Highly || InGaAs || High || [http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] || Example || Jon Doe || Example |
||
+ | |H2SO4:H2O2:H2O = 1:1:10 |
||
+ | |~600 |
||
+ | | |
||
+ | |InP |
||
+ | |High |
||
+ | | |
||
+ | |Exothermic, may reduce selectivity if hot |
||
+ | | |
||
|- |
|- |
||
+ | |GaAs |
||
− | | InP|| HCl:H3PO4(1:3)|| ~1000 || Highly || InGaAsP || High || [http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] || Example || Jon Doe || Example |
||
+ | |NH4OH:H2O2 = 1:30 |
||
+ | | |
||
+ | | - |
||
+ | |AlGaAs, |
||
+ | Al > 80% |
||
+ | |||
+ | ~200nm stop-etch |
||
+ | |High |
||
+ | | |
||
+ | | |
||
+ | |Garrett Cole |
||
|- |
|- |
||
+ | |AlGaAs, |
||
− | | InP|| H3PO4:HCl(3:1)|| ~1000 || Highly || InGaAs || High || [http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] || Example || Jon Doe || Example |
||
+ | Al ≥80% |
||
+ | |HF:H2O = 1:20 |
||
+ | | |
||
+ | | - |
||
+ | |GaAs |
||
+ | |High |
||
+ | | |
||
+ | | |
||
+ | |Garrett Cole |
||
|- |
|- |
||
+ | |Oxide of InP |
||
− | | InP|| H3PO4:HCl(3:1)|| ~1000 || Highly || InGaAsP || High || [http://tel.archives-ouvertes.fr/docs/00/76/94/02/PDF/VA2_LAMPONI_MARCO_15032012.pdf Lamponi (p.102)] || Example || Jon Doe || Example |
||
+ | |NH4OH:H2O = 1:10 |
||
+ | |1min to remove |
||
+ | | |
||
+ | |InP |
||
+ | |unknown |
||
+ | | |
||
+ | | |
||
+ | |[[Ning Cao]] |
||
|- |
|- |
||
+ | |Oxide of GaAs |
||
− | | Al2O3 ''(ALD Plasma 300C)''|| 300MIF Developer || ~1.6 || || || High || Measured in-house || Rate slows with time, Selective to most non-Al Materials || JTB || Example |
||
+ | |HCl:H2O = 1:10 |
||
+ | |1min to remove |
||
+ | | |
||
+ | |GaAs |
||
+ | |unknown |
||
+ | | |
||
+ | | |
||
+ | |[[Demis D. John]] |
||
|- |
|- |
||
− | | |
+ | |Al2O3 ''(ALD Plasma 300C)''||Developer: 300MIF||~1.6||None |
+ | |Most non-Al Materials. |
||
+ | |High||Measured in-house||Rate slows with time.||JTB |
||
|- |
|- |
||
− | | |
+ | |Al2O3 ''(ALD Plasma 300C)''||Developer: 400K||~2.2||None |
+ | |Most non-Al Materials. |
||
+ | |High||Measured in-house||Rate slows with time.||JTB |
||
|- |
|- |
||
− | | |
+ | |Al2O3 ''(ALD Plasma 300C)''||Developer: 400K (1:4)||~1.6||None |
+ | |Most non-Al Materials. |
||
+ | |High||Measured in-house||Rate slows with time.||JTB |
||
|- |
|- |
||
+ | |Al2O3 ''(ALD Plasma 300C)''||NH4OH:H2O2:H2O (1:2:50)||~<0.5|| || || ||Measured in-house||Rate slows with time||JTB |
||
− | |[https://www.nanotech.ucsb.edu/wiki/index.php/Sputtering_Recipes#Al2O3_deposition_.28IBD.29 Al2O3 ''(IBD)''] |
||
+ | |- |
||
− | |Buffered HF |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Al2O3_deposition_.28IBD.29 Al2O3 ''(IBD)''] |
||
− | |167 |
||
+ | |HF ("Buffered HF Improved", Transene) |
||
− | | |
||
+ | |~170 |
||
+ | |None |
||
|Photoresist |
|Photoresist |
||
|High |
|High |
||
|Measured in-house |
|Measured in-house |
||
+ | |May need to increase adhesion with thin SiO2 layer, and 100°C baked HMDS. |
||
− | | |
||
− | |Biljana Stamenic |
+ | |Biljana Stamenic, |
− | + | 2017-12 |
|
|- |
|- |
||
− | |[https:// |
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Al2O3_deposition_.28IBD.29 Al2O3 ''(IBD)''] |
− | |726 MiF |
+ | |Developer: 726 MiF |
|3.5 |
|3.5 |
||
+ | |None |
||
+ | |Most non-Al Materials. |
||
+ | |High |
||
+ | |Measured in-house |
||
| |
| |
||
+ | |Demis D. John, |
||
− | | |
||
+ | 2017-11 |
||
− | | |
||
+ | |- |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Al2O3_Deposition_.28Sputter_4.29 Al2O3 ''(AJA#4)''] |
||
+ | |Developer: 300 MiF |
||
+ | |4.30 |
||
+ | |None |
||
+ | |Most non-Al Materials. |
||
+ | |High |
||
|Measured in-house |
|Measured in-house |
||
| |
| |
||
|Demis D. John |
|Demis D. John |
||
+ | 2018-02 |
||
− | |2017-11 |
||
|- |
|- |
||
− | |[https:// |
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiO2_deposition_.28PECVD_.231.29 SiO2 ''(PECVD #1)''] |
+ | |HF ("Buffered HF Improved", Transene) |
||
− | |300 MiF Developer |
||
+ | |~500 |
||
− | |4.30 |
||
+ | |None |
||
− | | |
||
+ | |Photoresist |
||
+ | |High |
||
+ | |Measured in-house |
||
+ | |May need to increase adhesion with 100°C baked HMDS. |
||
+ | |Biljana Stamenic |
||
+ | 2017 |
||
+ | |- |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiO2_deposition_.28PECVD_.232.29 SiO2 ''(PECVD #2)''] |
||
+ | |HF ("Buffered HF Improved", Transene) |
||
+ | |~500 |
||
+ | |None |
||
+ | |Photoresist |
||
+ | |High |
||
+ | |Measured in-house |
||
+ | |May need to increase adhesion with 100°C baked HMDS. |
||
+ | |Biljana Stamenic |
||
+ | 2017 |
||
+ | |- |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#SiO2_deposition_.28IBD.29 SiO2 (IBD)] |
||
+ | |HF ("Buffered HF Improved", Transene) |
||
+ | |~350 |
||
+ | |None |
||
+ | |Photoresist |
||
+ | |High |
||
+ | |Measured in-house |
||
| |
| |
||
+ | |Biljana Stamenic |
||
+ | 2016 |
||
+ | |- |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiN_deposition_.28PECVD_.231.29 Si3N4 (PECVD#1)] |
||
+ | |HF ("Buffered HF Improved", Transene) |
||
+ | |85 |
||
+ | |None |
||
+ | |Photoresist |
||
+ | |High |
||
+ | |Measured in-house |
||
| |
| |
||
+ | |Biljana Stamenic |
||
+ | 2018-04 |
||
+ | |- |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#SiN_deposition_.28PECVD_.232.29 Si3N4 (PECVD#2)] |
||
+ | |HF ("Buffered HF Improved", Transene) |
||
+ | |35–45 |
||
+ | |None |
||
+ | |Photoresist |
||
+ | |High |
||
|Measured in-house |
|Measured in-house |
||
| |
| |
||
+ | |Biljana Stamenic |
||
− | |Demis D. John |
||
− | + | 2018-05 |
|
|- |
|- |
||
− | |[https:// |
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=PECVD_Recipes#Low-Stress_SiN_deposition_.28PECVD_.232.29 Si3N4 Low-Stress (PECVD#2)] |
− | |Buffered HF |
+ | |HF ("Buffered HF Improved", Transene) |
+ | |35–50 |
||
− | |~100 |
||
+ | |None |
||
+ | |Photoresist |
||
+ | |High |
||
+ | |Measured in-house |
||
| |
| |
||
+ | |Biljana Stamenic |
||
+ | 2018-05 |
||
+ | |- |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Si3N4_deposition_.28IBD.29 Si3N4 (IBD)] |
||
+ | |HF ("Buffered HF Improved", Transene) |
||
+ | |5–15 |
||
+ | |None |
||
|Photoresist |
|Photoresist |
||
|High |
|High |
||
Line 115: | Line 233: | ||
| |
| |
||
|Biljana Stamenic |
|Biljana Stamenic |
||
+ | 2014 |
||
− | |2017 |
||
|- |
|- |
||
− | |[https:// |
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#Ta2O5_deposition_.28IBD.29 Ta2O5 (IBD)] |
− | |Buffered HF |
+ | |HF ("Buffered HF Improved", Transene) |
+ | |0.4 |
||
− | |~500 |
||
+ | |None |
||
+ | |Photoresist |
||
+ | |High |
||
+ | |Measured in-house |
||
| |
| |
||
+ | |Biljana Stamenic |
||
+ | 2016-12 |
||
+ | |- |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/index.php?title=Sputtering_Recipes#TiO2_deposition_.28IBD.29 TiO2 (IBD)] |
||
+ | |HF ("Buffered HF Improved", Transene) |
||
+ | |1.0–2.0 |
||
+ | |None |
||
|Photoresist |
|Photoresist |
||
|High |
|High |
||
Line 126: | Line 255: | ||
| |
| |
||
|Biljana Stamenic |
|Biljana Stamenic |
||
+ | 2014-12 |
||
− | |2017 |
||
+ | |- |
||
+ | |Si (<100> crystalline) |
||
+ | |KOH (45%) @ 87°C |
||
+ | |~730 |
||
+ | |High, Crystallographic, ~55° |
||
+ | |Low-Stress Si3N4 - either [[PECVD Recipes#Low-Stress SiN deposition .28PECVD .232.29|PECVD #2]] or Commercial LPCVD Si3N4 |
||
+ | Other Si3N4 also OK. |
||
+ | |LS-SiN: High |
||
+ | PR etches quickly, SiO2 etches slowly. |
||
+ | |Measured In-House |
||
+ | - Search online. |
||
+ | |Use the Covered, Heated vertical bath ([[Wet Benches#Wafer Toxic Corrosive Benches|Dedi cated bath in Bay 4]]). Slight Bubbler. |
||
+ | |Brian Thibeault |
||
+ | 2017 |
||
+ | |- |
||
+ | !Material!!Etchant!!Rate (nm/min)!!Anisotropy!!Selective to!!Selectivity!!Ref.!!Notes!!Confirmed By/Date |
||
|} |
|} |
||
+ | ==Wet Etching References== |
||
+ | |||
+ | #[http://ieeexplore.ieee.org/abstract/document/546406/ Etch rates for Micromachining Processing (IEEE Jnl. MEMS, 1996)] - includes tables of etch rates of numerous metals vs. various wet and dry etchants. |
||
+ | #[http://ieeexplore.ieee.org/abstract/document/1257354/ Etch rates for micromachining-Part II (IEEE Jnl. MEMS, 2003)] - expanded tables containing resists, dielectrics, metals and semiconductors vs. many wet etch chemicals. |
||
+ | #[http://www.sciencedirect.com/science/article/pii/S0927796X00000279 Guide to references on III±V semiconductor chemical etching] - exhaustive list of wet etchants for etching various semiconductors, including selective etches. |
||
+ | #[http://transene.com/etch-compatibility/ Transene's Chemical Compatibility Chart] provides a useful quick-reference for which Transene etchants attack which materials. |
||
+ | ##As a side-note, [http://transene.com/ Transene] provides many pre-mixed solutions that you can order, saving you the time and uncertainty of measuring/mixing such chemicals yourself. Make sure you check with us before ordering so we know how to handle the chemical before it arrives. |
||
+ | |||
+ | ===Compound Semiconductor Etching=== |
||
+ | [http://www.sciencedirect.com/science/article/pii/S0927796X00000279 Guide to references on III±V semiconductor chemical etching] |
||
+ | |||
+ | Please add any confirmed etches from this reference to the {{HLink|Wet Etching Recipes|The Master Table of Wet Etching (Include All Materials)}}. |
||
+ | |||
+ | ===Metal Etching=== |
||
+ | |||
+ | *[//wiki.nanotech.ucsb.edu/w/images/c/c3/Ta_and_Cr_E-beam_deposition_and_wet_etch_test.pdf Selective Wet Etch of Cr over Ta using Cr Etchant] |
||
+ | *[//wiki.nanotech.ucsb.edu/w/images/d/dc/ITO_Deposition-250C-Ebeam2-HCl-Wet-Etch.pdf Wet Etch of ITO using Heated, Diluted HCl Solution] |
||
+ | |||
+ | ===Silicon etching=== |
||
+ | [http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=546406 Etch rates for micromachining processing] |
||
+ | |||
+ | [http://ieeexplore.ieee.org/xpls/abs_all.jsp?arnumber=1257354 Etch rates for micromachining processing-part II] |
||
+ | |||
+ | Please add any confirmed etches from this reference to the {{HLink|Wet Etching Recipes|The Master Table of Wet Etching (Include All Materials)}}. |
||
+ | |||
+ | ==Organic removal== |
||
+ | |||
+ | ===Piranha Solution=== |
||
+ | |||
+ | *'''Careful!''' Read about how to prepare and handle this safely: |
||
+ | **[http://web.mit.edu/cortiz/www/PiranhaSafety.doc MIT's Piranha Solution safety document] |
||
+ | **[https://www.ehs.harvard.edu/sites/default/files/lab_safety_guideline_piranha_etch.pdf Harvard EHS's Handling Document] |
||
+ | **Used for etching away Photoresist residues after dry etching. |
||
+ | |||
+ | ===PureStrip (Transene)=== |
||
+ | |||
+ | *Heat to @ 70°C |
||
+ | **[[Wet Benches#Wafer Toxic Corrosive Benches|Vertical heated bath on Wafer Toxic-Corrosive bench in Bay 4]] |
||
+ | **After a few days heated, it loses potency - must drain + replenish with fresh solution. |
||
+ | **~30-90min will remove stubborn, microscopic PR residues from dry etching. |
||
+ | |||
+ | ==[[Gold Plating Bench|Gold Plating Bench (Technic SEMCON 1000)]]== |
||
+ | Standard plating recipes are described and taught during the equipment training for this tool. |
||
+ | |||
+ | Electroplating first requires a Gold seed layer to be present on all surfaces to be plated. Common ways to produce this are to: |
||
+ | |||
+ | * [[Tool List#Sputter Deposition|Sputter-coat]] a thin (~10nm) Au layer on all surfaces of the wafer, being careful to consider shadowing effects during the dep (eg. in high-aspect ratio trenches). |
||
+ | * Perform photolithography to protect (and prevent plating) in desired areas - areas with no photoresist will be plated. |
||
+ | * Perform the electroplating on the [[Gold Plating Bench|Technic SemCon]], contacting the seed layer with electrodes and executing the program for the desired current/time to achieve the plating thickness (typically microns). |
||
+ | * Strip the photoresist using standard solvents. |
||
+ | * Use the [[CAIBE (Oxford Ion Mill)|Oxford Ion Mill]] to blanket etch the 10nm seed layer all over the wafer (the 10nm removed from the plated regions will be negligible). |
||
+ | |||
+ | ==[[Chemical-Mechanical Polisher (Logitech)|Chemi-Mechanical Polishing (CMP)]]== |
||
+ | ''To Be Added'' |
||
+ | |||
+ | ==[[Mechanical Polisher (Allied)|Mechanical Polishing (Allied)]]== |
||
+ | ''To Be Added'' |
||
[[Category:Processing]] |
[[Category:Processing]] |
Latest revision as of 10:34, 1 April 2024
Chemicals Available
- The Chemical Lists show stocked chemicals, photolithography chemicals, and how to bring new chemicals.
Table of Wet Etching Recipes
Use the ↑ ↓ Arrows in the header row to sort the entire table by material, selectivity, etchant etc.
Material | Etchant | Rate (nm/min) | Anisotropy | Selective to | Selectivity | Ref. | Notes | Confirmed By/Date |
---|---|---|---|---|---|---|---|---|
Photoresist, polymers/organics | H2SO4:H2O2 = 3:1 | typ. 5-10min etch for polymer residue | Cr, W, Au, Pt, Si, SiO2, SiN | Dangerous boiling hazard - see Piranha Solution
section below. Etches Ti |
Demis D. John, 2017 | |||
InP | H3PO4:HCl = 3:1 | ~1000 | Highly | InGaAsP | High | Lamponi (p.102) | ||
InGaAsP | To Be Added | InP | High | |||||
InP | H3PO4:HCl = 3:1 | ~1000 | Highly | InGaAs | High | Lamponi (p.102) | ||
InP | HCl:H2O = 3:1 | ~5000 | InGaAs
~200nm stop-etch |
High | Bubbles while etching | |||
InGaAs | H2SO4:H2O2:H2O = 1:1:10 | ~600 | InP | High | Exothermic, may reduce selectivity if hot | |||
GaAs | NH4OH:H2O2 = 1:30 | - | AlGaAs,
Al > 80% ~200nm stop-etch |
High | Garrett Cole | |||
AlGaAs,
Al ≥80% |
HF:H2O = 1:20 | - | GaAs | High | Garrett Cole | |||
Oxide of InP | NH4OH:H2O = 1:10 | 1min to remove | InP | unknown | Ning Cao | |||
Oxide of GaAs | HCl:H2O = 1:10 | 1min to remove | GaAs | unknown | Demis D. John | |||
Al2O3 (ALD Plasma 300C) | Developer: 300MIF | ~1.6 | None | Most non-Al Materials. | High | Measured in-house | Rate slows with time. | JTB |
Al2O3 (ALD Plasma 300C) | Developer: 400K | ~2.2 | None | Most non-Al Materials. | High | Measured in-house | Rate slows with time. | JTB |
Al2O3 (ALD Plasma 300C) | Developer: 400K (1:4) | ~1.6 | None | Most non-Al Materials. | High | Measured in-house | Rate slows with time. | JTB |
Al2O3 (ALD Plasma 300C) | NH4OH:H2O2:H2O (1:2:50) | ~<0.5 | Measured in-house | Rate slows with time | JTB | |||
Al2O3 (IBD) | HF ("Buffered HF Improved", Transene) | ~170 | None | Photoresist | High | Measured in-house | May need to increase adhesion with thin SiO2 layer, and 100°C baked HMDS. | Biljana Stamenic,
2017-12 |
Al2O3 (IBD) | Developer: 726 MiF | 3.5 | None | Most non-Al Materials. | High | Measured in-house | Demis D. John,
2017-11 | |
Al2O3 (AJA#4) | Developer: 300 MiF | 4.30 | None | Most non-Al Materials. | High | Measured in-house | Demis D. John
2018-02 | |
SiO2 (PECVD #1) | HF ("Buffered HF Improved", Transene) | ~500 | None | Photoresist | High | Measured in-house | May need to increase adhesion with 100°C baked HMDS. | Biljana Stamenic
2017 |
SiO2 (PECVD #2) | HF ("Buffered HF Improved", Transene) | ~500 | None | Photoresist | High | Measured in-house | May need to increase adhesion with 100°C baked HMDS. | Biljana Stamenic
2017 |
SiO2 (IBD) | HF ("Buffered HF Improved", Transene) | ~350 | None | Photoresist | High | Measured in-house | Biljana Stamenic
2016 | |
Si3N4 (PECVD#1) | HF ("Buffered HF Improved", Transene) | 85 | None | Photoresist | High | Measured in-house | Biljana Stamenic
2018-04 | |
Si3N4 (PECVD#2) | HF ("Buffered HF Improved", Transene) | 35–45 | None | Photoresist | High | Measured in-house | Biljana Stamenic
2018-05 | |
Si3N4 Low-Stress (PECVD#2) | HF ("Buffered HF Improved", Transene) | 35–50 | None | Photoresist | High | Measured in-house | Biljana Stamenic
2018-05 | |
Si3N4 (IBD) | HF ("Buffered HF Improved", Transene) | 5–15 | None | Photoresist | High | Measured in-house | Biljana Stamenic
2014 | |
Ta2O5 (IBD) | HF ("Buffered HF Improved", Transene) | 0.4 | None | Photoresist | High | Measured in-house | Biljana Stamenic
2016-12 | |
TiO2 (IBD) | HF ("Buffered HF Improved", Transene) | 1.0–2.0 | None | Photoresist | High | Measured in-house | Biljana Stamenic
2014-12 | |
Si (<100> crystalline) | KOH (45%) @ 87°C | ~730 | High, Crystallographic, ~55° | Low-Stress Si3N4 - either PECVD #2 or Commercial LPCVD Si3N4
Other Si3N4 also OK. |
LS-SiN: High
PR etches quickly, SiO2 etches slowly. |
Measured In-House
- Search online. |
Use the Covered, Heated vertical bath (Dedi cated bath in Bay 4). Slight Bubbler. | Brian Thibeault
2017 |
Material | Etchant | Rate (nm/min) | Anisotropy | Selective to | Selectivity | Ref. | Notes | Confirmed By/Date |
Wet Etching References
- Etch rates for Micromachining Processing (IEEE Jnl. MEMS, 1996) - includes tables of etch rates of numerous metals vs. various wet and dry etchants.
- Etch rates for micromachining-Part II (IEEE Jnl. MEMS, 2003) - expanded tables containing resists, dielectrics, metals and semiconductors vs. many wet etch chemicals.
- Guide to references on III±V semiconductor chemical etching - exhaustive list of wet etchants for etching various semiconductors, including selective etches.
- Transene's Chemical Compatibility Chart provides a useful quick-reference for which Transene etchants attack which materials.
- As a side-note, Transene provides many pre-mixed solutions that you can order, saving you the time and uncertainty of measuring/mixing such chemicals yourself. Make sure you check with us before ordering so we know how to handle the chemical before it arrives.
Compound Semiconductor Etching
Guide to references on III±V semiconductor chemical etching
Please add any confirmed etches from this reference to the The Master Table of Wet Etching (Include All Materials).
Metal Etching
- Selective Wet Etch of Cr over Ta using Cr Etchant
- Wet Etch of ITO using Heated, Diluted HCl Solution
Silicon etching
Etch rates for micromachining processing
Etch rates for micromachining processing-part II
Please add any confirmed etches from this reference to the The Master Table of Wet Etching (Include All Materials).
Organic removal
Piranha Solution
- Careful! Read about how to prepare and handle this safely:
- MIT's Piranha Solution safety document
- Harvard EHS's Handling Document
- Used for etching away Photoresist residues after dry etching.
PureStrip (Transene)
- Heat to @ 70°C
- Vertical heated bath on Wafer Toxic-Corrosive bench in Bay 4
- After a few days heated, it loses potency - must drain + replenish with fresh solution.
- ~30-90min will remove stubborn, microscopic PR residues from dry etching.
Gold Plating Bench (Technic SEMCON 1000)
Standard plating recipes are described and taught during the equipment training for this tool.
Electroplating first requires a Gold seed layer to be present on all surfaces to be plated. Common ways to produce this are to:
- Sputter-coat a thin (~10nm) Au layer on all surfaces of the wafer, being careful to consider shadowing effects during the dep (eg. in high-aspect ratio trenches).
- Perform photolithography to protect (and prevent plating) in desired areas - areas with no photoresist will be plated.
- Perform the electroplating on the Technic SemCon, contacting the seed layer with electrodes and executing the program for the desired current/time to achieve the plating thickness (typically microns).
- Strip the photoresist using standard solvents.
- Use the Oxford Ion Mill to blanket etch the 10nm seed layer all over the wafer (the 10nm removed from the plated regions will be negligible).
Chemi-Mechanical Polishing (CMP)
To Be Added
Mechanical Polishing (Allied)
To Be Added