Fluorescence Microscope (Olympus MX51): Difference between revisions
Jump to navigation
Jump to search
(→About) |
(updated details / FLuorescence filters) |
||
Line 13: | Line 13: | ||
=Detailed Specifications= |
=Detailed Specifications= |
||
* Digital image capture & calibrated measurement via integrated Olympus Streams Essentials software. |
|||
* Objectives: 5x, 10x, 20x, 50x, 100x, 150x |
|||
* Reflected (Episcopic) & Transmitted (Diascopic) Illumination modes |
|||
* Filters via rotatable selector: |
|||
** '''BF''': Bright Field normal top-illuminated imaging |
|||
** '''DF''': Dark-Field imaging |
|||
** '''DIC''': Differential Interference Contrast "Nomarski" Imaging. Also requires "DIC-R" prism to be inserted. Polarization manually selected on DIC-R prism. |
|||
* Fluorescence Filters (see [https://www.olympus-lifescience.com/en/microscope-resource/primer/techniques/fluorescence/anatomy/fluoromicroanatomy/ Olympus tutorial]): |
|||
** '''''You must contact Tony Bosch for training on Fluorescence before using this feature!!!''''' |
|||
** '''WB''': "Blue" Fluorescence Filter. Excitation: 450-480nm / Dichroic 500nm / Barrier 515nm+ |
|||
** '''WG''': "Green" Fluorescence Filter. Excitation: 510-550nm / Dichroic 570nm / Barrier 590nm+ |
|||
** '''WU''': "Ultraviolet" Fluorescence Filter. Excitation: 330-385nm / Dichroic 400nm / Barrier 420nm+ |
|||
=Documentation= |
=Documentation= |
Revision as of 23:39, 18 December 2018
|
About
System for general purpose imaging. The system has bright field, dark-field, and DIC imaging modes. Screen captures and measurements on samples can be made through use of Olympus' Streams Essential Software, installed on the attached computer.
Detailed Specifications
- Digital image capture & calibrated measurement via integrated Olympus Streams Essentials software.
- Objectives: 5x, 10x, 20x, 50x, 100x, 150x
- Reflected (Episcopic) & Transmitted (Diascopic) Illumination modes
- Filters via rotatable selector:
- BF: Bright Field normal top-illuminated imaging
- DF: Dark-Field imaging
- DIC: Differential Interference Contrast "Nomarski" Imaging. Also requires "DIC-R" prism to be inserted. Polarization manually selected on DIC-R prism.
- Fluorescence Filters (see Olympus tutorial):
- You must contact Tony Bosch for training on Fluorescence before using this feature!!!
- WB: "Blue" Fluorescence Filter. Excitation: 450-480nm / Dichroic 500nm / Barrier 515nm+
- WG: "Green" Fluorescence Filter. Excitation: 510-550nm / Dichroic 570nm / Barrier 590nm+
- WU: "Ultraviolet" Fluorescence Filter. Excitation: 330-385nm / Dichroic 400nm / Barrier 420nm+