Holographic Lith/PL Setup (Custom): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
m (→‎About: added term "holography" for searching, changed PR to THMR and linked)
(link to PL system page)
Line 14: Line 14:


[[Lithography Recipes#Chemicals Stocked .2B Datasheets|THMR-3600HP]] resist spin-coated to ~ 80 nm thickness is used for grating exposure. 2-D gratings can be formed by rotating the sample and doing multiple exposures. Total exposure times are controlled by a manual shutter and are generally several minutes in length.
[[Lithography Recipes#Chemicals Stocked .2B Datasheets|THMR-3600HP]] resist spin-coated to ~ 80 nm thickness is used for grating exposure. 2-D gratings can be formed by rotating the sample and doing multiple exposures. Total exposure times are controlled by a manual shutter and are generally several minutes in length.

A photoluminescence measurement setup is co-located on the optical table. Please see the page for the PL system for more info:
* [[Photoluminescence PL Setup (Custom)]]


=Detailed Specifications=
=Detailed Specifications=

Revision as of 15:06, 9 October 2020

Holographic Lith/PL Setup (Custom)
Holograph.jpg
Tool Type Lithography
Location Bay 6
Supervisor Ning Cao
Supervisor Phone (805) 893-4689
Supervisor E-Mail ningcao@ece.ucsb.edu
Description Custom Built Interference Lithography
Manufacturer Kimmon
Sign up for this tool


About

The interference lithography system (aka. Holography) at UCSB uses a 15mW, single-mode, 325 nm HeCd laser that is filtered and expanded by pinhole filters to produce the large area exposure beam. The system uses a simple mirror configuration with a fixed 90 degree angle between the mirror and sample. The entire mirror/sample assembly is rotated in 0.1 degree increments to change the grating pitch from ~ 200 nm to ~ 280 nm (35 to 55 degrees) over an ~ 2 cm x 2 cm exposure area.

THMR-3600HP resist spin-coated to ~ 80 nm thickness is used for grating exposure. 2-D gratings can be formed by rotating the sample and doing multiple exposures. Total exposure times are controlled by a manual shutter and are generally several minutes in length.

A photoluminescence measurement setup is co-located on the optical table. Please see the page for the PL system for more info:

Detailed Specifications

  • 15 mW single TEM mode HeCd laser
  • ~ 2 cm x 2 cm uniform exposure area
  • ~ several minute exposure times
  • Grating period adjustable from ~200 to ~280 nm with a single stage