| Line 10: | Line 10: | ||
|SEM Images |
|SEM Images |
||
|- |
|- |
||
| |
|3/30/2022 |
||
| ⚫ | |||
| ⚫ | |||
| |
|164 |
||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
|1.23 |
|1.23 |
||
| |
| |
||
|[https://wiki.nanotech.ucsb.edu/wiki/images/ |
|[https://wiki.nanotech.ucsb.edu/w/images/c/c3/ICP2_03_45D_002.jpg <nowiki>[1]</nowiki>] [https://wiki.nanotech.ucsb.edu/w/images/7/74/ICP2_03_CS_003.jpg <nowiki>[2]</nowiki>] |
||
|- |
|- |
||
|3/ |
|3/8/2022 |
||
| ⚫ | |||
| ⚫ | |||
| |
|144 |
||
|1. |
|1.02 |
||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
|7/18/2019 |
|||
| ⚫ | |||
|162 |
|||
| ⚫ | |||
| |
| |
||
|[https://wiki.nanotech.ucsb.edu/wiki/images/ |
|[https://wiki.nanotech.ucsb.edu/w/images/f/f4/ICP2_02_45D_001.jpg <nowiki>[1]</nowiki>] [https://wiki.nanotech.ucsb.edu/w/images/5/54/ICP2_02_CS_004.jpg <nowiki>[2]</nowiki>] |
||
|- |
|- |
||
| |
|3/2/2022 |
||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
|1. |
|1.29 |
||
| |
| |
||
|[https://wiki.nanotech.ucsb.edu/wiki/images/ |
|[https://wiki.nanotech.ucsb.edu/w/images/2/2e/ICP2_01_45D_002.jpg <nowiki>[1]</nowiki>] [https://wiki.nanotech.ucsb.edu/w/images/6/69/ICP2_01_CS_007.jpg <nowiki>[2]</nowiki>] |
||
|- |
|- |
||
|8/9/ |
|8/9/2021 |
||
| ⚫ | |||
| ⚫ | |||
| |
|140 |
||
|0. |
|0.97 |
||
| ⚫ | |||
| ⚫ | |||
|[https://wiki.nanotech.ucsb.edu/w/images/ |
|[https://wiki.nanotech.ucsb.edu/w/images/2/26/I2210508.pdf] |
||
|- |
|- |
||
| |
|8/9/2021 |
||
| ⚫ | |||
| ⚫ | |||
| |
|147 |
||
|1. |
|1.06 |
||
| ⚫ | |||
| ⚫ | |||
|[https://wiki.nanotech.ucsb.edu/w/images/ |
|[https://wiki.nanotech.ucsb.edu/w/images/2/28/I2210411.pdf] |
||
| ⚫ | |||
|5/19/2021 |
|||
| ⚫ | |||
|163 |
|||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
|- |
|- |
||
|7/21/2021 |
|7/21/2021 |
||
| Line 73: | Line 52: | ||
|[https://wiki.nanotech.ucsb.edu/w/images/c/c3/I2210308.pdf] |
|[https://wiki.nanotech.ucsb.edu/w/images/c/c3/I2210308.pdf] |
||
|- |
|- |
||
| |
|5/19/2021 |
||
| ⚫ | |||
| ⚫ | |||
| |
|163 |
||
|1. |
|1.11 |
||
| ⚫ | |||
| ⚫ | |||
|[https://wiki.nanotech.ucsb.edu/w/images/ |
|[https://wiki.nanotech.ucsb.edu/w/images/4/4c/I2210214.pdf] |
||
|- |
|- |
||
| |
|1/7/2021 |
||
| ⚫ | |||
| ⚫ | |||
| |
|144 |
||
| |
|1.20 |
||
| ⚫ | |||
| ⚫ | |||
|[https://wiki.nanotech.ucsb.edu/w/images/ |
|[https://wiki.nanotech.ucsb.edu/w/images/0/0c/I2210102.pdf] |
||
|- |
|- |
||
| |
|8/9/2020 |
||
| ⚫ | |||
| ⚫ | |||
|102 |
|||
| ⚫ | |||
| |
|0.86 |
||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
|1/16/2020 |
|||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| |
| |
||
|[https://wiki.nanotech.ucsb.edu |
|[https://wiki.nanotech.ucsb.edu/wiki/images/0/09/I2200122.pdf] |
||
|- |
|- |
||
| |
|7/18/2019 |
||
| ⚫ | |||
| ⚫ | |||
| |
|162 |
||
|1. |
|1.37 |
||
| |
| |
||
|[https://wiki.nanotech.ucsb.edu |
|[https://wiki.nanotech.ucsb.edu/wiki/images/2/23/I2190506.pdf] |
||
|- |
|- |
||
|3/ |
|3/6/2019 |
||
| ⚫ | |||
| ⚫ | |||
| |
|151 |
||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
|1.23 |
|1.23 |
||
| |
| |
||
|[https://wiki.nanotech.ucsb.edu |
|[https://wiki.nanotech.ucsb.edu/wiki/images/f/f9/SiO2_Etch_using_ICP2-no_O2-a.pdf] |
||
| ⚫ | |||
|10/5/2018 |
|||
| ⚫ | |||
|160 |
|||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
|}<br /> |
|}<br /> |
||
===Alternate Data (not updated)=== |
===Alternate Data (not updated)=== |
||
Revision as of 20:16, 13 April 2022
| ICP#2: 0.5Pa, 50/900W, CHF3/CF4=10/30 sccm, time=210 sec | |||||
| Date | Sample# | Etch Rate (nm/min) | Etch Selectivity (SiO2/PR) | Averaged Sidewall Angle (o) | SEM Images |
| 3/30/2022 | NP_ICP2_03 | 164 | 1.23 | [1] [2] | |
| 3/8/2022 | NP_ICP2_02 | 144 | 1.02 | [1] [2] | |
| 3/2/2022 | NP_ICP2_01 | 169.6 | 1.29 | [1] [2] | |
| 8/9/2021 | I22105 | 140 | 0.97 | After etching diamond sample for 1 hour using Cl2/Ar. Found chamber/etches are ok. | [1] |
| 8/9/2021 | I22104 | 147 | 1.06 | Before etching diamond sample for 1 hour using Cl2/Ar | [2] |
| 7/21/2021 | I22103 | 134 | 1.09 | Investigating reports of low etch rate | [3] |
| 5/19/2021 | I22102 | 163 | 1.11 | Etch time=130 sec | [4] |
| 1/7/2021 | I22101 | 144 | 1.20 | [5] | |
| 8/9/2020 | I22002 | 102 | 0.86 | caused by air leaking to CHF3 channel | [6] |
| 1/16/2020 | I22001 | 149 | 1.21 | [7] | |
| 7/18/2019 | I21905 | 162 | 1.37 | [8] | |
| 3/6/2019 | I21904 | 151 | 1.23 | 85.6 | [9] |
| 1/28/2019 | I21901 | 146 | 1.23 | [10] | |
| 10/5/2018 | SiO2#02 | 160 | 1.2 | 82.1 | [11] |
Alternate Data (not updated)