{{tool2|{{PAGENAME}} |
||
|picture=Thermal2.jpg |
|picture=Thermal2.jpg |
||
|type = Vacuum Deposition |
|type = Vacuum Deposition |
||
|super= |
|super= Michael Barreraz |
||
|super2= Don Freeborn |
|||
|location=Bay 3 |
|location=Bay 3 |
||
|description = ? |
|description = ? |
||
Line 10: | Line 11: | ||
}} |
}} |
||
= About = |
== About == |
||
Thermal evaporator #2 is the designated "Solder" evaporator. The tool is used primarily for solder materials including Gold, Indium and Tin. See the Recipes section below for other materials used. |
|||
Use this tool for materials with low melting temperatures, or for materials that have a high risk of contamination in the e-beam evaporators. |
|||
= Materials Table = |
|||
== Detailed Specifications == |
|||
{| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center; font-size: 95%" class="collapsible collapsed wikitable" |
|||
Wafers up to 12" can be mounted. |
|||
|- |
|||
! colspan=6 width=1300 height=35 bgcolor="#D0E7FF" align="center"|<div style="font-size: 150%;">Materials Table</div> |
|||
==Documentation== |
|||
|- bgcolor="#D0E7FF" |
|||
*[//wiki.nanotech.ucsb.edu/w/images/b/b3/Thermal_Evaporator2.pdf Operating Instructions] |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Material''' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Symbol''' |
|||
== Recipes == |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Density, g/cm3''' |
|||
* [[Thermal Evaporation Recipes#Thermal Evaporator 2|Recipes > Thermal Evaporation Recipes > Thermal Evaporator 2]] |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Z Ratio''' |
|||
** ''Visit this page for the materials table and evaporation parameters for all materials available.'' |
|||
! width="45" bgcolor="#D0E7FF" align="center" | '''Tooling, %''' |
|||
! width="100" bgcolor="#D0E7FF" align="center" | '''Comments''' |
|||
|- |
|||
|Aluminum |
|||
|Al |
|||
|2.70 |
|||
|1.08 |
|||
| |
|||
| |
|||
|- |
|||
|Antimony |
|||
|Sb |
|||
|6.620 |
|||
|0.768 |
|||
| |
|||
| |
|||
|- |
|||
|Arsenic |
|||
|As |
|||
|5.73 |
|||
|0.996 |
|||
| |
|||
| |
|||
|- |
|||
|Beryllium |
|||
|Be |
|||
|1.85 |
|||
|0.543 |
|||
| |
|||
| |
|||
|- |
|||
|Boron |
|||
|B |
|||
|2.54 |
|||
|0.389 |
|||
| |
|||
| |
|||
|- |
|||
|Cadmium |
|||
|Cd |
|||
|8.64 |
|||
|0.682 |
|||
| |
|||
| |
|||
|- |
|||
|Cadmium sulfide |
|||
|CdS |
|||
|4.83 |
|||
|1.02 |
|||
| |
|||
| |
|||
|- |
|||
|Cadmium telluride |
|||
|CdTe |
|||
|5.85 |
|||
|0.980 |
|||
| |
|||
| |
|||
|- |
|||
|Calcium fluoride |
|||
|CaF<sub>2</sub> |
|||
|3.18 |
|||
|0.775 |
|||
| |
|||
|Used often |
|||
|- |
|||
|Carbon (graphite) |
|||
|C |
|||
|2.25 |
|||
|3.26 |
|||
| |
|||
| |
|||
|- |
|||
|Chromium |
|||
|Cr |
|||
|7.20 |
|||
|0.305 |
|||
|155 |
|||
|Used often |
|||
|- |
|||
|Cobalt |
|||
|Co |
|||
|8.71 |
|||
|0.343 |
|||
| |
|||
| |
|||
|- |
|||
|Copper |
|||
|Cu |
|||
|8.93 |
|||
|0.437 |
|||
| |
|||
| |
|||
|- |
|||
|Gallium |
|||
|Ga |
|||
|5.93 |
|||
|0.593 |
|||
| |
|||
| |
|||
|- |
|||
|Gallium Arsenide |
|||
|GaAs |
|||
|5.31 |
|||
|1.59 |
|||
| |
|||
| |
|||
|- |
|||
|Germanium |
|||
|Ge |
|||
|5.35 |
|||
|0.516 |
|||
| |
|||
| |
|||
|- |
|||
|Gold |
|||
|Au |
|||
|19.30 |
|||
|0.381 |
|||
| |
|||
|Used often |
|||
|- |
|||
|Indium |
|||
|In |
|||
|7.30 |
|||
|0.841 |
|||
|120 |
|||
|USed often |
|||
|- |
|||
|Indium antimonide |
|||
|InSb |
|||
|5.76 |
|||
|0.769 |
|||
| |
|||
| |
|||
|- |
|||
|Iridium |
|||
|Ir |
|||
|22.4 |
|||
|0.129 |
|||
| |
|||
| |
|||
|- |
|||
|Iron |
|||
|Fe |
|||
|7.86 |
|||
|0.349 |
|||
| |
|||
| |
|||
|- |
|||
|Lead |
|||
|Pb |
|||
|11.3 |
|||
|1.13 |
|||
| |
|||
| |
|||
|- |
|||
|Lead sulfide |
|||
|PbS |
|||
|7.50 |
|||
|0.566 |
|||
| |
|||
| |
|||
|- |
|||
|Lithium fluoride |
|||
|LiF |
|||
|2.638 |
|||
|0.778 |
|||
| |
|||
| |
|||
|- |
|||
|Magnesium |
|||
|Mg |
|||
|1.74 |
|||
|1.61 |
|||
| |
|||
| |
|||
|- |
|||
|Magnesium oxide |
|||
|MgO |
|||
|3.58 |
|||
|0.411 |
|||
| |
|||
| |
|||
|- |
|||
|Magnesium Fluoride |
|||
|MgF<sub>2</sub> |
|||
|3.00 ( should be 3.18) |
|||
|0.427( should be 0.637) |
|||
| |
|||
| |
|||
|- |
|||
|Manganese |
|||
|Mn |
|||
|7.20 |
|||
|0.377 |
|||
| |
|||
| |
|||
|- |
|||
|Manganese(II) Sulfide |
|||
|MnS |
|||
|3.99 |
|||
|0.94 |
|||
| |
|||
| |
|||
|- |
|||
|Mercury |
|||
|Hg |
|||
|13.46 |
|||
|0.74 |
|||
| |
|||
| |
|||
|- |
|||
|Molybdenum |
|||
|Mo |
|||
|10.2 |
|||
|0.257 |
|||
| |
|||
| |
|||
|- |
|||
|Nickel |
|||
|Ni |
|||
|8.91 |
|||
|0.331 |
|||
| |
|||
|Used often |
|||
|- |
|||
|Niobium |
|||
|Nb |
|||
|8.578 |
|||
|0.492 |
|||
| |
|||
| |
|||
|- |
|||
|Niobium(V) Oxide |
|||
|Nb<sub>2</sub>O<sub>5</sub> |
|||
|4.47 |
|||
|1.00 |
|||
| |
|||
|z ratio has not been established |
|||
|- |
|||
|Palladium |
|||
|Pd |
|||
|12.0 |
|||
|0.357 |
|||
| |
|||
| |
|||
|- |
|||
|Platinum |
|||
|Pt |
|||
|21.4 |
|||
|0.245 |
|||
| |
|||
| |
|||
|- |
|||
|Potasium Cloride |
|||
|KCl |
|||
|1.98 |
|||
|2.05 |
|||
| |
|||
| |
|||
|- |
|||
|Selenium |
|||
|Se |
|||
|4.82 |
|||
|0.864 |
|||
| |
|||
| |
|||
|- |
|||
|Silicon |
|||
|Si |
|||
|2.32 |
|||
|0.712 |
|||
| |
|||
| |
|||
|- |
|||
|Silicon(II) Oxide |
|||
|SiO |
|||
|2.13 |
|||
|0.87 |
|||
| |
|||
| |
|||
|- |
|||
|Silicon dioxide (fused quarz) |
|||
|SiO<sub>2</sub> |
|||
|2.20 |
|||
|1.07 |
|||
| |
|||
| |
|||
|- |
|||
|Silver |
|||
|Ag |
|||
|10.5 |
|||
|0.529 |
|||
| |
|||
| |
|||
|- |
|||
|Silver bromide |
|||
|AgBr |
|||
|6.47 |
|||
|1.18 |
|||
| |
|||
| |
|||
|- |
|||
|Silver chloride |
|||
|AgCl |
|||
|5.56 |
|||
|1.32 |
|||
| |
|||
| |
|||
|- |
|||
|Sodium |
|||
|Na |
|||
|0.97 |
|||
|4.8 |
|||
| |
|||
| |
|||
|- |
|||
|Sodium chloride |
|||
|NaCl |
|||
|2.17 |
|||
|1.57 |
|||
| |
|||
| |
|||
|- |
|||
|Tantalum |
|||
|Ta |
|||
|16.6 |
|||
|0.262 |
|||
| |
|||
| |
|||
|- |
|||
|Tantalum(IV) oxide |
|||
|Ta<sub>2</sub>O<sub>5</sub> |
|||
|8.2 |
|||
|0.30 |
|||
| |
|||
| |
|||
|- |
|||
|Tin |
|||
|Sn |
|||
|7.30 |
|||
|0.724 |
|||
| |
|||
|Used often |
|||
|- |
|||
|Titanium |
|||
|Ti |
|||
|4.50 |
|||
|0.628 |
|||
| |
|||
|Used often |
|||
|- |
|||
|Titanium (IV) Oxide |
|||
|TiO<sub>2</sub> |
|||
|4.26 |
|||
|0.40 |
|||
| |
|||
| |
|||
|- |
|||
|Titanium Oxide |
|||
|TiO |
|||
|4.90 |
|||
|N/A |
|||
| |
|||
| |
|||
|- |
|||
|Tangsten |
|||
|W |
|||
|19.3 |
|||
|0.163 |
|||
| |
|||
| |
|||
|- |
|||
|Tangsten carbide |
|||
|WC |
|||
|15.6 |
|||
|0.151 |
|||
| |
|||
| |
|||
|- |
|||
|Zinc |
|||
|Zn |
|||
|7.04 |
|||
|0.514 |
|||
| |
|||
|Used often; Density should be 7.14 |
|||
|- |
|||
|Zinc Oxide |
|||
|ZnO |
|||
|5.61 |
|||
|0.556 |
|||
| |
|||
| |
|||
|- |
|||
|Zinc Sulphide |
|||
|ZnS |
|||
|4.09 |
|||
|0.775 |
|||
| |
|||
| |
|||
|- |
|||
|Zirconium |
|||
|Zr |
|||
|6.51 |
|||
|0.60 |
|||
| |
|||
|Should be 6.40 |
|||
|- |
|||
|Zirconium Oxide |
|||
|ZrO<sub>2</sub> |
|||
|5.6 |
|||
| |
|||
| |
|||
|Should be 5.49 |
|||
|- |
Latest revision as of 17:29, 30 August 2022
|
About
Thermal evaporator #2 is the designated "Solder" evaporator. The tool is used primarily for solder materials including Gold, Indium and Tin. See the Recipes section below for other materials used.
Use this tool for materials with low melting temperatures, or for materials that have a high risk of contamination in the e-beam evaporators.
Detailed Specifications
Wafers up to 12" can be mounted.
Documentation
Recipes
- Recipes > Thermal Evaporation Recipes > Thermal Evaporator 2
- Visit this page for the materials table and evaporation parameters for all materials available.