Important Guidelines

●      Wafer Thickness: Do not use wafers thicker than 1 mm.

●      Log all scans and issues in the logbook.

●      Always log off after completing your session.


Procedure

1. Login Instructions

·      Access Code: fab (new password from 2025).

●      If the keyboard isn’t functional, use the virtual keyboard on the monitor.

●      Double-click the SCAN icon to launch the software.


2. Loading a Recipe

  1. Navigate to Recipe > Load (upper-left corner).
  2. Select one of the standard recipes for 6-inch wafers:

○      STD_UCSB_6inch_G4: For thicknesses 0.16–1.6 µm.

○      STD_UCSB_6inch_G2: For thicknesses 1.6–28 µm.

  1. Click OK. The recipe name will appear at the top of the screen.

3. Customizing a Recipe

  1. Select Recipe > Save as and name the recipe (e.g., your name or company name).
  2. Go to Recipe > View/Modify to adjust parameters:

○      Wafer diameter

○      Edge exclusion

○      Gain

○      Threshold

  1. Under Bin Split, click Default Bin to apply changes.
  2. Save the updated recipe.

4. Load Wafers into the Cassette

  1. Use the white cassette for 6-inch wafers.
  2. Load wafers with the face up; the major flat can face either direction.
  3. Secure the cassette in the right indexer (specific for 6-inch wafers).
  4. On the monitor, select CASS (Cassette) and wait for the tool to detect the wafers.

5. Start Scanning

To Scan a Single Wafer:

●      Select the desired wafer from the monitor.

●      Click START (upper menu bar).

●      Wait for the scan to finish and record the results.

To Scan Multiple Wafers:

●      Select AUTO mode.

●      Click START to begin scanning all wafers in the cassette.


6. Completion

  1. Once the scan is complete:

○      Press ENTER on the small laptop to save the scan as a photo.

○      Location: Desktop/Local Disk(C)/Users/Public/Public Documentation.

○      Photos can be accessed/downloaded via the Nanofab server.

○      Note: Full data-saving functionality will be available in January 2025.

  1. Click HOME, select RECIPE/Exit, and log off.

7. Log Book

●      Ensure the cassette is properly seated in the right indexer.

●      Always log off after completing your session.