Rapid Thermal Processor (SSI Solaris 150): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
No edit summary
(added table of max temp vs time)
 
(9 intermediate revisions by 4 users not shown)
Line 1: Line 1:
{{tool
{{tool
|super = Brian Lingg
|super = Bill Millerski
|picture=RTP-Solaris-150.jpg
|picture=RTP-Solaris-150.jpg
|type = Thermal Processing
|type = Thermal Processing
Line 12: Line 12:


==About==
==About==
The Solaris 150 is a manual loading RTP system for R&D and pre-production. The Solaris 150 can process up to 152.4mm substrates at a temperature range from RT- 1250 degrees. The unique temperature measurement system of the Solaris requires virtually no calibration for different wafer types and backside emmissivity differences.
The Solaris 150 is a manual loading “Rapid Therm Process” (RPT) or “Rapid Thermal Anneal” (RTA) system for R&D and pre-production. The Solaris 150 can process up to 152.4mm substrates at a temperature range from RT- 1200 degrees. The unique temperature measurement system of the Solaris requires virtually no calibration for different wafer types and backside emmissivity differences.


The Solaris uses a unique PID process controller that ensures accurate temperature stability and uniformity. The system can accommodate four interlocked MFCs for gas mixing and forming gas processing. The Solaris is designed for silicon implant annealing and monitoring and compound semiconductor implant activation and ohmic alloying.
The Solaris uses a unique PID process controller that ensures accurate temperature stability and uniformity. The system can accommodate four interlocked MFCs for gas mixing and forming gas processing. The Solaris is designed for silicon implant annealing and monitoring and compound semiconductor implant activation and ohmic alloying.


==Detailed Specificiations==
==Detailed Specificiations==
Wafer handling: Manual loading of wafer into the oven, single wafer processing.


Wafer sizes: 2", 3", 4", 5" and 6" wafers and pieces.
*Wafer handling: Manual loading of wafer into the oven, single wafer processing.
*Wafer sizes: 2", 3", 4", 5" and 6" wafers and pieces.
*Ramp up rate: 10-200°C per second, user-controllable.
*Recommended steady state duration: 0-600 seconds per step.
*Ramp down rate: Temperature Dependent, max 150°C per second.
*Steady state temperature range: 300C - 1200°C
*Thermocouple temperature accuracy: + 2.5C
*Temperature repeatability: + 3C or better at 1150C wafer-to-wafer.
*Temperature uniformity: + 5C across a 6" (150 mm) wafer at 1150C.
*Gasses available: Nitrogen, Oxygen, Argon, Forming gas (90% Nitrogen / 10% Hydrogen).


==Operating Instructions==
Ramp up rate: 10-200C per second, user-controllable.


*[https://wiki.nanofab.ucsb.edu/w/images/7/7a/RTA_SSI_Operating_Instructions.pdf Solaris 150 Operating Procedure]
Recommended steady state duration: 0-600 seconds per step.
*[https://wiki.nanotech.ucsb.edu/wiki/images/6/66/Solaris_150_install_manual_software3_0_revQ.pdf User manual]

Ramp down rate: Temperature Dependent, max 150°C per second.

Recommended steady state temperature range: 300C - 1250°C

Thermocouple temperature accuracy: + 2.5C

Temperature repeatability: + 3C or better at 1150C wafer-to-wafer.

Temperature uniformity: + 5C across a 6" (150 mm) wafer at 1150C.

==Operating Instructions==


== Process Info ==
*[[SSI Solaris 150 - Operating Procedure|Solaris 150 Operating Procedure]]
{| class="wikitable"
*[https://www.nanotech.ucsb.edu/wiki/images/6/66/Solaris_150_install_manual_software3_0_revQ.pdf User manual]
|+Maximum Temperature vs Time thresholds
!Temp
!Time
|-
|1200°C
|2.0 min
|-
|1150°C
|2.5 min
|-
|1100°C
|5.0 min
|-
|1050°C
|5.5 min
|-
|1000°C
|6.0 min
|-
|950°C
|6.5 min
|-
|900°C
|7.5 min
|-
|850°C
|8.5 min
|-
|800°C
|10 min
|-
|750°C
|12 min
|-
|700°C
|15 min
|-
|650°C
|20 min
|-
|600°C
|30 min
|-
|550°C
|45 min
|-
|500°C
|60 min
|-
|450°C
|76 min
|-
|400°C
|85 min
|}

Latest revision as of 19:30, 14 March 2025

Rapid Thermal Processor (SSI Solaris 150)
RTP-Solaris-150.jpg
Tool Type Thermal Processing
Location Bay 5
Supervisor Bill Millerski
Supervisor Phone (805) 893-2655
Supervisor E-Mail wmillerski@ucsb.edu
Description Rapid Thermal Annealing (RTA)
Manufacturer SSI
Model Solaris 150
Thermal Processing Recipes
Sign up for this tool



About

The Solaris 150 is a manual loading “Rapid Therm Process” (RPT) or “Rapid Thermal Anneal” (RTA) system for R&D and pre-production. The Solaris 150 can process up to 152.4mm substrates at a temperature range from RT- 1200 degrees. The unique temperature measurement system of the Solaris requires virtually no calibration for different wafer types and backside emmissivity differences.

The Solaris uses a unique PID process controller that ensures accurate temperature stability and uniformity. The system can accommodate four interlocked MFCs for gas mixing and forming gas processing. The Solaris is designed for silicon implant annealing and monitoring and compound semiconductor implant activation and ohmic alloying.

Detailed Specificiations

  • Wafer handling: Manual loading of wafer into the oven, single wafer processing.
  • Wafer sizes: 2", 3", 4", 5" and 6" wafers and pieces.
  • Ramp up rate: 10-200°C per second, user-controllable.
  • Recommended steady state duration: 0-600 seconds per step.
  • Ramp down rate: Temperature Dependent, max 150°C per second.
  • Steady state temperature range: 300C - 1200°C
  • Thermocouple temperature accuracy: + 2.5C
  • Temperature repeatability: + 3C or better at 1150C wafer-to-wafer.
  • Temperature uniformity: + 5C across a 6" (150 mm) wafer at 1150C.
  • Gasses available: Nitrogen, Oxygen, Argon, Forming gas (90% Nitrogen / 10% Hydrogen).

Operating Instructions

Process Info

Maximum Temperature vs Time thresholds
Temp Time
1200°C 2.0 min
1150°C 2.5 min
1100°C 5.0 min
1050°C 5.5 min
1000°C 6.0 min
950°C 6.5 min
900°C 7.5 min
850°C 8.5 min
800°C 10 min
750°C 12 min
700°C 15 min
650°C 20 min
600°C 30 min
550°C 45 min
500°C 60 min
450°C 76 min
400°C 85 min