==About== |
||
The Model 7100 Series is a semi-automatic dicing saw. Semiconductor, glass, and plastic substrates of all types can be automatically or manually cut. This dicing Saw is optimized for multi-angle dicing of tight tolerance substrates up to 200 mm diameter in size. |
|||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
| ⚫ | |||
== Operating Procedures == |
|||
* [[ADT 7100 - Initial Setup Before Cutting|Setting up the Saw before Cutting]] |
|||
* [[ADT 7100 - Standard Dicing Procedure - Programmed Cut Map|Standard Dicing Procedure - Programmed Cut Map]] |
|||
* [[ADT 7100 - Standard Dicing Procedure - Single Cuts|Standard Dicing Procedure - Single Cuts]] |
|||
* [[ADT 7100 - Recovering an Old Recipe (2019)|Recovering an Old Recipe]] |
|||
*Maximum Wafer Size: 8" |
|||
*Parts mounted to UV-release tape for dicing |
|||
*Automated cut maps at multiple angles (0° and 90° typical) |
|||
**"Polygon" cutting to approximate Coring wafers is possible. |
|||
*~Few micron alignment to on-wafer features. |
|||
*Thermocarbon Resnoid dicing blades provided by Staff |
|||
== |
==Operating Procedures== |
||
* Packaging Recipes > [https://wiki.nanotech.ucsb.edu/wiki/index.php/Packaging_Recipes?venotify=created#Dicing_Saw_Recipes_.28ADT_7100.29 Dicing Saw Recipes (ADT 7100)] |
|||
*[https://wiki.nanofab.ucsb.edu/w/images/c/c5/ADT_SOP_Rev_K.pdf ADT Dicing Saw Standard Operating Procedure] |
|||
| ⚫ | |||
*[https://wiki.nanofab.ucsb.edu/w/images/5/5c/Post_Dicing_Tape_Expansion_SOP.pdf Post Dicing Tape Expansion] |
|||
*[https://wiki.nanofab.ucsb.edu/w/images/a/ac/Ultron_1042R-B_Film_Specs.pdf Ultron Systems, Inc. 1042R-B Anti-Static Ultraviolet Film TDS] |
|||
*[https://wiki.nanofab.ucsb.edu/w/images/5/59/Recipe_Importing_New_Software.pdf Recovering an Old Recipe (2023)] |
|||
==Recipes== |
|||
*Recipes > Packaging > '''[[Packaging Recipes#Dicing Saw Recipes .28ADT 7100.29|Dicing Saw Recipes (ADT 7100)]]''' |
|||
| ⚫ | |||
==== Dicing Processing Parameters ==== |
|||
If you are not familiar with dicing, the following document lists the important parameters you need to document before starting your job: |
|||
* [[Media:Dicing - Example Dicing Instructions for UC Santa Barbara v2.pptx|Example Dicing Instructions for UC Santa Barbara v2.pptx]] ''- by Demis D. John'' |
|||
Revision as of 19:29, 14 May 2025
| ||||||||||||||||||||||||||||||||
About
The Model 7100 Series is a semi-automatic dicing saw. Semiconductor, glass, and plastic substrates of all types can be automatically or manually cut. This dicing Saw is optimized for multi-angle dicing of tight tolerance substrates up to 200 mm diameter in size.
Check the Dicing Saw Recipes page for the blades we currently stock.
An ADT WM-966 tape applicator is used to mount samples to UV-release tape for dicing and an Ultron Systems UH104-8 UV lamp system is used to release samples from the tape after dicing.
Contact Staff for blades and dicing frames for your group.
Detailed Specifications
- Maximum Wafer Size: 8"
- Parts mounted to UV-release tape for dicing
- Automated cut maps at multiple angles (0° and 90° typical)
- "Polygon" cutting to approximate Coring wafers is possible.
- ~Few micron alignment to on-wafer features.
- Thermocarbon Resnoid dicing blades provided by Staff
Operating Procedures
- ADT Dicing Saw Standard Operating Procedure
- Post Dicing Tape Expansion
- Ultron Systems, Inc. 1042R-B Anti-Static Ultraviolet Film TDS
- Recovering an Old Recipe (2023)
Recipes
- Recipes > Packaging > Dicing Saw Recipes (ADT 7100)
Be sure to also see the recipes for protecting your sample from dicing dust, and mounting/unmounting.
Dicing Processing Parameters
If you are not familiar with dicing, the following document lists the important parameters you need to document before starting your job: