PECVD 2 (Advanced Vacuum): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
Biljana (talk | contribs)
Barreraz (talk | contribs)
 
(46 intermediate revisions by 3 users not shown)
Line 1: Line 1:
{{tool|{{PAGENAME}}
{{tool2|{{PAGENAME}}
|picture=PECVD2.jpg
|picture=PECVD2.jpg
|type = Vacuum Deposition
|type = Vacuum Deposition
|super= Mike Silva
|super= Michael Barreraz
|super2= Don Freeborn
|phone=(805)839-3918x219
|phone=(805)839-7975
|location=Bay 2
|location=Bay 2
|email=silva@ece.ucsb.edu
|email=silva@ece.ucsb.edu
|description = Vision 310 Advanced Vacuum PECVD
|description = Vision 310 Advanced Vacuum PECVD
|manufacturer = Veeco
|manufacturer = Plasma-Therm
|materials =
|materials =
|toolid=15
|toolid=15
}}
}}
= About =
==About==


This open-load system is dedicated to PECVD of SiO<sub>2</sub>, SiN<sub>x</sub>, SiO<sub>x</sub>N<sub>y</sub>, and a-Si using Silane (2%SiH<sub>4</sub>, 98% He), N<sub>2</sub>O, NH<sub>3</sub>, and N<sub>2</sub> gases. The sample electrode has a 270mm diameter useable area, allowing for multiple 4” wafer depositions in a single run. Standard operating temperature is 300C, but can be user changed for temps ranging anywhere from 250 to 350C. The system is equipped with a dual generator, dual frequency option for growth of low-stress Nitride films.
*'''Films/Gases''': This open-load system is dedicated to PECVD of '''SiO<sub>2</sub>, SiN<sub>x</sub>, SiO<sub>x</sub>N<sub>y</sub>, and a-Si''' using Silane (2%SiH<sub>4</sub>, 98% He), N<sub>2</sub>O, NH<sub>3</sub>, and N<sub>2</sub> gases.
*'''Size''': The sample electrode has a 270mm diameter useable area, allowing for multiple 4” wafer depositions in a single run.
*'''Temperature''': Standard operating temperature is 300C, but can be user changed for temps ranging anywhere from 250 to 350C.
*'''Low-Stress Si<sub>3</sub>N<sub>4</sub>''': The system is equipped with a dual generator, dual frequency option for growth of Low-stress Nitride films.
**These films alternate between thin (<10nm) compressive and tensile layers.
**The Low-Stress Si3N4 film recipe are tested approx. monthly, and kept within ±100MPa. Data can be found at Recipes (below) > [[PECVD Recipes#Historical Data 5|Low-Stress Nitride]].

==See Also==


=See Also=
*[http://www.advanced-vacuum.se/Ny-sida-8.html Vision 310 Product Page]
*[http://www.advanced-vacuum.se/Ny-sida-8.html Vision 310 Product Page]

*[[PECVD Recipes]]
==Documentation==

*
*
*
*
*
*
*
*
*
*
*[https://wiki.nanofab.ucsb.edu/w/images/9/9a/PECVD2_SOP_Final_V45.47_3-6-26-compressed.pdf Operating Instructions]
*[https://wiki.nanofab.ucsb.edu/w/images/6/68/PECVD2_STD_LS_Si3N4v4recipe_depositiontime.pdf Modifying Deposition Time in "STD LS-Si3N4v4" Recipe]
*[https://wiki.nanofab.ucsb.edu/w/images/1/1b/PECVD2_Recipe_Temperature_Change.pdf Modifying Deposition Temperature]

*[[Wafer Coating Process Traveler]]
*For particle counting method, see the [[Wafer scanning process traveler|Surfscan Scanning Procedure]]

==Recipes & Data==
[[File:PECVD_SPC_Chart_Example.png|alt=SPC chart example|thumb|228x228px|Example Process Control Charts (SPC) for thin-film DepCals.]]
[[File:Surfscan_230113A7G2_after_low_particles.jpg|alt=screenshot of surfscan particle count|thumb|205x205px|Example particle counts taken on each film.]]
[[File:PECVD1_SiO2_F50_WaferMap_example.jpg|alt=Screenshot of Filmetrics F50 wafermap of typical DepCals film|thumb|215x215px|Example of DepCals Thickness/Refractive Index uniformity measurement (4% shown here).]]
*Standard Recipes: [[PECVD Recipes#PECVD 2 .28Advanced Vacuum.29|'''Recipes > Deposition > <u>PECVD #2</u>''']]

*A list of ''all available'' deposited films can be found here: [[Vacuum Deposition Recipes|Recipes > Vacuum Deposition Recipes]]

===Process Control Data===

*Process Control Charts: '''[[Process Group - Process Control Data#PECVD #1 (PlasmaTherm 790) - Process Control|Process Control > PECVD#1]]'''
**[https://docs.google.com/spreadsheets/d/1iSW1eAAg824y9PYYLG9aiaw53PEJ-f9ofylpVlCDq9Y/edit#gid=272916741 Plots of all data]
**[https://docs.google.com/spreadsheets/d/1iSW1eAAg824y9PYYLG9aiaw53PEJ-f9ofylpVlCDq9Y/edit#gid=1313651154 SiO<sub>2</sub> Data]
**[https://docs.google.com/spreadsheets/d/1iSW1eAAg824y9PYYLG9aiaw53PEJ-f9ofylpVlCDq9Y/edit#gid=773875841 Si<sub>3</sub>N<sub>4</sub> Data]
**[https://docs.google.com/spreadsheets/d/1iSW1eAAg824y9PYYLG9aiaw53PEJ-f9ofylpVlCDq9Y/edit#gid=584923738 Low-Stress Si<sub>3</sub>N<sub>4</sub> Data]
***[https://docs.google.com/spreadsheets/d/1iSW1eAAg824y9PYYLG9aiaw53PEJ-f9ofylpVlCDq9Y/edit#gid=203400760 Plots of Low-Stress Si<sub>3</sub>N<sub>4</sub> Data]

Latest revision as of 18:55, 6 March 2026

PECVD 2 (Advanced Vacuum)
Location Bay 2
Tool Type Vacuum Deposition
Manufacturer Plasma-Therm
Description Vision 310 Advanced Vacuum PECVD

Primary Supervisor Michael Barreraz
(805) 893-4147
mikebarreraz@ece.ucsb.edu

Secondary Supervisor

Don Freeborn


Recipes Vacuum Deposition Recipes

SignupMonkey: Sign up for this tool


About

  • Films/Gases: This open-load system is dedicated to PECVD of SiO2, SiNx, SiOxNy, and a-Si using Silane (2%SiH4, 98% He), N2O, NH3, and N2 gases.
  • Size: The sample electrode has a 270mm diameter useable area, allowing for multiple 4” wafer depositions in a single run.
  • Temperature: Standard operating temperature is 300C, but can be user changed for temps ranging anywhere from 250 to 350C.
  • Low-Stress Si3N4: The system is equipped with a dual generator, dual frequency option for growth of Low-stress Nitride films.
    • These films alternate between thin (<10nm) compressive and tensile layers.
    • The Low-Stress Si3N4 film recipe are tested approx. monthly, and kept within ±100MPa. Data can be found at Recipes (below) > Low-Stress Nitride.

See Also

Documentation

Recipes & Data

SPC chart example
Example Process Control Charts (SPC) for thin-film DepCals.
screenshot of surfscan particle count
Example particle counts taken on each film.
Screenshot of Filmetrics F50 wafermap of typical DepCals film
Example of DepCals Thickness/Refractive Index uniformity measurement (4% shown here).

Process Control Data