Holographic Lith/PL Setup (Custom): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
link to PL system page
 
(8 intermediate revisions by the same user not shown)
Line 1: Line 1:
{{tool|{{PAGENAME}}
{{tool2|{{PAGENAME}}
|picture=Holograph.jpg
|picture=Holograph.jpg
|type = Lithography
|type = Lithography
|super= Ning Cao
|super= Demis D. John
|super2 = Brian Thibeault
|phone=(805) 839-5689
|location=Bay 6
|location=Bay 6
|description = Custom Interference Lithography
|email=Ningcao@ece.ucsb.edu
|description = Custom Built Interference Lithography
|manufacturer = Custom-Built Free-Space Optics
|manufacturer = Kimmon
|toolid=35
|toolid=35
}}
}}
{{ToolActions|
⚫
= About =
|toolid = 35
|InstructionsURL =
|ProcessControlURL =
|TrainingURL =
}}
⚫
==About==
===Holography Lithography===
The interference lithography system (aka. Holography) at UCSB uses a 15mW, single-mode, 325 nm HeCd laser that is filtered and expanded by pinhole filters to produce the large area exposure beam. The system uses a simple mirror configuration with a fixed 90 degree angle between the mirror and sample. The entire mirror/sample assembly is rotated in 0.1 degree increments to change the grating pitch from ~ 200 nm to ~ 280 nm (35 to 55 degrees) over an ~ 2 cm x 2 cm exposure area.
The interference lithography system (aka. Holography) at UCSB uses a 15mW, single-mode, 325 nm HeCd laser that is filtered and expanded by pinhole filters to produce the large area exposure beam. The system uses a simple mirror configuration with a fixed 90 degree angle between the mirror and sample. The entire mirror/sample assembly is rotated in 0.1 degree increments to change the grating pitch from ~ 200 nm to ~ 280 nm (35 to 55 degrees) over an ~ 2 cm x 2 cm exposure area.


[[Lithography Recipes#Chemicals Stocked .2B Datasheets|THMR-3600HP]] resist spin-coated to ~ 80 nm thickness is used for grating exposure. 2-D gratings can be formed by rotating the sample and doing multiple exposures. Total exposure times are controlled by a manual shutter and are generally several minutes in length.
[https://wiki.nanofab.ucsb.edu/w/images/3/33/XHRiC-Anti-Reflective-Coating.pdf XHRiC] bottom anti-reflection coating (BARC) & [[Lithography Recipes#Chemicals Stocked .2B Datasheets|THMR-3600HP]] imaging resist spin-coated to ~80 nm thickness is used for grating exposure. 2-D gratings can be formed by rotating the sample and doing multiple exposures. Total exposure times are controlled by a manual shutter and are generally several minutes in length.


⚫
===Photoluminescence Measurement===
A photoluminescence measurement setup is co-located on the optical table. Please see the page for the PL system for more info:
A photoluminescence measurement setup is co-located on the optical table. Please see the page for the PL system for more info:
⚫
* [[Photoluminescence PL Setup (Custom)]]


*[[Photoluminescence PL Setup (Custom)|'''Photoluminescence PL Setup (Custom)''']]
⚫
=Detailed Specifications=

⚫
*15 mW single TEM mode HeCd laser
⚫
==Detailed Specifications==
⚫
*~ 2 cm x 2 cm uniform exposure area

⚫
*~ several minute exposure times
⚫
*15 mW single TEM mode HeCd laser (Kimmon)
⚫
*Grating period adjustable from ~200 to ~280 nm with a single stage
*Max sample size: ~1.5 inch square uniform exposure area
⚫
*~2 cm x 2 cm uniform exposure area
⚫
*Several-minute exposure times (manual shutter)
⚫
*Grating period adjustable from ~200 to ~280 nm with rotation stage
*Manual optics alignment.
*1-D and 2-D gratings possible, see recipes below.

==Operating Procedures==
* [https://docs.google.com/document/d/1PFmkN9Br5jH1cdzAzTS0ZqL4g3SGOnbV/edit?usp=sharing&ouid=117965716847601392764&rtpof=true&sd=true Holography Std. Operating Procedures] - ''(by Gopi Meena)''

==Recipes==

*[[Lithography Recipes#Holography%20Recipes|Lithography Recipes > Holography Recipes]] - developed recipes for 1D (line/space) and 2D (pillars) gratings.

Latest revision as of 09:37, 6 December 2025

Holographic Lith/PL Setup (Custom)
Location Bay 6
Tool Type Lithography
Manufacturer Custom-Built Free-Space Optics
Description Custom Interference Lithography

Primary Supervisor Demis D. John
(805) 893-5934
demis@ucsb.edu

Secondary Supervisor

Brian Thibeault


Recipes

SignupMonkey: Sign up for this tool


Instructions + Procedures

Sign Up

Get Trained

Maintenance
Messages
Report Issue


About

Holography Lithography

The interference lithography system (aka. Holography) at UCSB uses a 15mW, single-mode, 325 nm HeCd laser that is filtered and expanded by pinhole filters to produce the large area exposure beam. The system uses a simple mirror configuration with a fixed 90 degree angle between the mirror and sample. The entire mirror/sample assembly is rotated in 0.1 degree increments to change the grating pitch from ~ 200 nm to ~ 280 nm (35 to 55 degrees) over an ~ 2 cm x 2 cm exposure area.

XHRiC bottom anti-reflection coating (BARC) & THMR-3600HP imaging resist spin-coated to ~80 nm thickness is used for grating exposure. 2-D gratings can be formed by rotating the sample and doing multiple exposures. Total exposure times are controlled by a manual shutter and are generally several minutes in length.

Photoluminescence Measurement

A photoluminescence measurement setup is co-located on the optical table. Please see the page for the PL system for more info:

Detailed Specifications

  • 15 mW single TEM mode HeCd laser (Kimmon)
  • Max sample size: ~1.5 inch square uniform exposure area
  • ~2 cm x 2 cm uniform exposure area
  • Several-minute exposure times (manual shutter)
  • Grating period adjustable from ~200 to ~280 nm with rotation stage
  • Manual optics alignment.
  • 1-D and 2-D gratings possible, see recipes below.

Operating Procedures

Recipes