SEM Sample Coater (Hummer): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
Barreraz (talk | contribs)
Instructions
→Operating Procedures: fixed bullet points, listed approx film thickness
 
(One intermediate revision by the same user not shown)
Line 11: Line 11:
}}
}}
=About=
=About=
This system is for coating Pd/Au thin layers onto samples that will charge during FESEM inspection.
This system is for coating thin (1-10nm) AuPd layers onto samples that will reduce charging during FESEM inspection.


===When to use===
===When to use===
Line 18: Line 18:
==Operating Procedures==
==Operating Procedures==


⚫
* [https://wiki.nanotech.ucsb.edu/w/images/5/5d/SEM_sample_coater_%28hummer%29_instructions.pdf SEM Sample Coater (Hummer) Instructions]


⚫
*[[Hummer SEM Sample Coater - Techniques to reduce charging in SEMs|Techniques to reduce charging in SEMs using Hummer Coater]]
⚫
[https://wiki.nanotech.ucsb.edu/w/images/5/5d/SEM_sample_coater_%28hummer%29_instructions.pdf SEM Sample Coater (Hummer) Instructions]

⚫
*[[Hummer SEM Sample Coater - Techniques to reduce charging in SEMs|Techniques to reduce charging in SEMs]]


<br />
<br />

Latest revision as of 00:17, 2 August 2024

SEM Sample Coater (Hummer)
Location Bay 1
Tool Type Inspection, Test and Characterization
Manufacturer Hummer

Primary Supervisor Michael Barreraz
(805) 893-4147
mikebarreraz@ece.ucsb.edu

Secondary Supervisor

Tony Bosch


Recipes


About

This system is for coating thin (1-10nm) AuPd layers onto samples that will reduce charging during FESEM inspection.

When to use

If you tried SEM'ing a sample, and the sample appears to drift/move during imaging/capture, or astigmatism is constantly changing, then your sample may be charging up and deflecting the electron beams. This is common for imaging insulators such as substrates made of glass (SiO2), or substrates coated in glass (SiO2, SiN) or thick photoresist (≥1µm). Thin layers (<<1µm) on top of conductive substrates (eg. 200nm SiO2 on Silicon substrate) can often allow electron beams and fields to penetrate to the conductor, possibly not requiring AuPd coating.

Operating Procedures