Atomic Force Microscope (Bruker ICON): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
Zwarburg (talk | contribs)
No edit summary
linked to high-aspect-ratio trench schematic
 
(13 intermediate revisions by 3 users not shown)
Line 1: Line 1:
{{tool|{{PAGENAME}}
{{tool2|{{PAGENAME}}
|picture=VeecoNanoman.jpg
|picture=ICON_AFM.jpg
|type = Inspection, Test and Characterization
|type = Inspection, Test and Characterization
|super= Bill Mitchell
|super= Aidan Hopkins
|super2= Bill Mitchell
|phone= (805)893-4974
|phone= (805)893-4974
|location=Bay 7
|location=Bay 5
|email=mitchell@ece.ucsb.edu
|email=mitchell@ece.ucsb.edu
|description = Dimension 3100 Nanoman
|description = Bruker ICON AFM
|manufacturer = Veeco Metrology, LLC
|manufacturer = Bruker Nano, Inc
|materials =
|materials =
|toolid= 1
}}
}}
= About =
= About =
The Bruker ICON AFM utilizes the latest paradigm in Atomic Force Microscopy - '''PeakForce Tapping®,''' a method that combines the best features of the legacy Contact Mode and Tapping Mode imaging techniques, namely direct force control and intermittent surface contact to reduce damaging lateral forces, to precisely control the probe-to-surface force interaction as pixel-to-pixel force curve measurements. This control results in the most consistent, highest resolution AFM imaging technique that can be used on a wide range of sample types.
The Dimension 3100 Nanoman AFM from Veeco provides a variety of high resolution surface imaging techniques and the ability to manipulate or create nanoscale structures directly. Techniques available for imaging include contact mode AFM, tapping mode AFM, Scanning Tunnelling AFM, Conductive AFM, and Scanning Capacitance Microscopy. With the Nanoman option, the X and Y deflecting piezo-elements are independently controlled, allowing for precise and direct placement of the tip anywhere within the field. Direct manipulation of particles on the surface is then possible by dragging the tip in the desired direction. This feedback control, coupled with direct control of conductive tip and substrate voltages, allows for direct-write oxidation on a variety of surfaces to create or modify nanostructures through local anodization.

Note that the ICON AFM also can also utilize the legacy ContactMode and TappingMode techniques for imaging if required.

In PeakForce Tapping®, the probe periodically taps the sample and the pN-level interaction force is measured directly by the deflection of the cantilever. Through superior force control, the feedback loop keeps the peak force constant, down to pN range, in both air and fluid, which is significantly lower than is typically used with other modes, such as TappingMode™, in the 1-10 nN range. PeakForce Tapping® enables the researcher to precisely control probe-to-sample interaction, providing the lowest available imaging forces.


=Detailed Specifications=
=Detailed Specifications=
*PeakForce Tapping® imaging for superior resolution and consistency
*Conductive AFM modules for nA-microAmp current measurements
*Legacy Contact Mode and Tapping Mode imaging also available
*Tunneling AFM modulefor pA-nA current measurements
*Substrate size range: small pieces (>2mm) up to 4" wafers
⚫
*Scanning Capacitance Microscopy
*Computer controlled XY stage for superior sample positioning; piezo Z scan range ~ 12um
*Resolution: Sub-nm height-measurement capability; X-Y resolution tip dependent
*Basic electrical measurement techniques are available but only using Contact Mode imaging.
*Registration tolerance to a known mark: Field size dependent
⚫
**Scanning Capacitance Microscopy (SCM)
*Minimum substrate size: small pieces
**Conductive AFM (CAFM) for high currents, Tunneling AFM (TUNA) for low currents
*Largest substrate size: 100 mm wafer
*Oxidation Line widths: Call for info.


=See Also=
=See Also=
*[https://www.bruker.com/products/surface-and-dimensional-analysis/atomic-force-microscopes/dimension-icon/overview.html Bruker ICON AFM overview]
*[http://www.veeco.com/ Veeco Website]
*[https://docs.google.com/drawings/d/1WLvuu-BD0pzroAbuUE52mt5bO5VWN3biWDdG27dM86o/edit?usp=sharing Limitations on measuring high-aspect ratio trenches] - what happens when your trench is too deep for the probe tip? You mostly measure the tip's shape!

Latest revision as of 19:56, 16 June 2026

Atomic Force Microscope (Bruker ICON)
Location Bay 5
Tool Type Inspection, Test and Characterization
Manufacturer Bruker Nano, Inc
Description Bruker ICON AFM

Primary Supervisor Aidan Hopkins
(805) 893-2343
hopkins@ece.ucsb.edu

Secondary Supervisor

Bill Mitchell


Recipes

SignupMonkey: Sign up for this tool


About

The Bruker ICON AFM utilizes the latest paradigm in Atomic Force Microscopy - PeakForce Tapping®, a method that combines the best features of the legacy Contact Mode and Tapping Mode imaging techniques, namely direct force control and intermittent surface contact to reduce damaging lateral forces, to precisely control the probe-to-surface force interaction as pixel-to-pixel force curve measurements. This control results in the most consistent, highest resolution AFM imaging technique that can be used on a wide range of sample types.

Note that the ICON AFM also can also utilize the legacy ContactMode and TappingMode techniques for imaging if required.

In PeakForce Tapping®, the probe periodically taps the sample and the pN-level interaction force is measured directly by the deflection of the cantilever. Through superior force control, the feedback loop keeps the peak force constant, down to pN range, in both air and fluid, which is significantly lower than is typically used with other modes, such as TappingMode™, in the 1-10 nN range. PeakForce Tapping® enables the researcher to precisely control probe-to-sample interaction, providing the lowest available imaging forces.

Detailed Specifications

  • PeakForce Tapping® imaging for superior resolution and consistency
  • Legacy Contact Mode and Tapping Mode imaging also available
  • Substrate size range: small pieces (>2mm) up to 4" wafers
  • Computer controlled XY stage for superior sample positioning; piezo Z scan range ~ 12um
  • Basic electrical measurement techniques are available but only using Contact Mode imaging.
    • Scanning Capacitance Microscopy (SCM)
    • Conductive AFM (CAFM) for high currents, Tunneling AFM (TUNA) for low currents