The three stack Tystar 8” furnace is used primarily for 3 processes. The processes are dedicated for one tube each:
The three stack Tystar 8” furnace is used primarily for 3 processes. The processes are dedicated for one tube each:
# SOG curing - Tube 1
* '''Tube #1''': SOG curing
# Dry or wet oxidation of silicon - Tubes 2 and 3
* '''Tubes #2 and #3''': Dry or wet oxidation of silicon
# General furnace annealing - Tube 3
* '''Tube #3''': General furnace annealing


Each process tube can accomodate up to one hundred 8” wafers per cycle. We have boats for 2", 3", 4", 6", 8" and irregular shaped pieces. The maximum temperature is 1050°C for the system. Gases used are O<sub>2</sub>, Steam from DI-H<sub>2</sub>O, N<sub>2</sub>.
Each process tube can accomodate up to one hundred 8” wafers per cycle. We have boats for 2", 3", 4", 6", 8" and irregular shaped pieces. The maximum temperature is 1050°C for the system. Gases used are O<sub>2</sub>, Steam from DI-H<sub>2</sub>O, N<sub>2</sub>.
Line 22: Line 22:
=Process Information=
=Process Information=


Recipe Characterization Data, such as thermal oxidation times, can be found on the recipe page:
⚫
Use the [http://cleanroom.byu.edu/OxideTimeCalc BYU] or [http://www.lelandstanfordjunior.com/thermaloxide.html Stanford] Thermal Oxidation Calculators to determine the time and temperature that will be necessary for your process needs. You can "calibrate" your oxidations to the Stanford calculator by adjusting the ''Partial Pressure'' on the calculator to match your experimental data.
* [[Thermal Processing Recipes|Thermal Processing Recipes: Tystar 8300]]
⚫
Use the [http://cleanroom.byu.edu/OxideTimeCalc BYU] or [http://www.lelandstanfordjunior.com/thermaloxide.html Stanford Leland Jr.] Thermal Oxidation Calculators to determine the time and temperature that will be necessary for your process needs. You can "calibrate" your oxidations to the Stanford calculator by adjusting the ''Partial Pressure'' on the calculator to match your experimental data.


Keep in mind that all process must be 30 minutes in length at a minimum. Processes less than 30 minutes will suffer from poor uniformity because the process tube will not have sufficient time to saturate with O<sub>2</sub> or DI-H<sub>2</sub>O.
Keep in mind that all process must be 30 minutes in length at a minimum. Processes less than 30 minutes will suffer from poor uniformity because the process tube will not have sufficient time to saturate with O<sub>2</sub> or DI-H<sub>2</sub>O.
Line 28: Line 30:
=Recipes=
=Recipes=


The following are the available recipes on each furnace tube:
⚫
Tube 1:

⚫
'''Tube 1:'''
* SOG425.001 - ''Spin-On Glass Cure''
* SOG425.001 - ''Spin-On Glass Cure''
* ALGAAS.001 - ''Oxidation of AlGaAs''
* ALGAAS.001 - ''Oxidation of AlGaAs''
Tube 2:
'''Tube 2:'''
* WET1050.002 - ''WetOx at 1050°C''
* WET1050.002 - ''WetOx at 1050°C''
* DRY1050.002 - ''DryOx at 1050°C''
* DRY1050.002 - ''DryOx at 1050°C''
* WETVAR.002 - ''WetOx, variable temp.''
* WETVAR.002 - ''WetOx, variable temp.''
* DRYVAR.002 - ''DryOx, variable temp.''
* DRYVAR.002 - ''DryOx, variable temp.''
Tube 3:
'''Tube 3:'''
* WET1050.003 - ''WetOx at 1050°C''
* WET1050.003 - ''WetOx at 1050°C''
* DRY1050.003 - ''DryOx at 1050°C''
* DRY1050.003 - ''DryOx at 1050°C''
Line 49: Line 53:
*[http://www.tystar.com/ Tystar] - Manufacturer of the tool
*[http://www.tystar.com/ Tystar] - Manufacturer of the tool
*[http://cleanroom.byu.edu/OxideThickCalc Silicon Thermal Oxide Thickness Calculator (BYU)] - Use this on-line calculator to calculate times for silicon oxidation.
*[http://cleanroom.byu.edu/OxideThickCalc Silicon Thermal Oxide Thickness Calculator (BYU)] - Use this on-line calculator to calculate times for silicon oxidation.
*[http://www.lelandstanfordjunior.com/thermaloxide.html Advanced Silicon Thermal Oxide Thickness Calculator (Stanford)] - Another thermal oxide calculator, with flexibility to vary ''partial pressure'' parameter to calibrate to your own process.
*[http://www.lelandstanfordjunior.com/thermaloxide.html Advanced Silicon Thermal Oxide Thickness Calculator (Stanford Leland Jr.)] - Another thermal oxide calculator, with flexibility to vary ''partial pressure'' parameter to calibrate to your own process.


=Operational Instructions=
=Operational Instructions=

Revision as of 19:28, 9 April 2018

Tube Furnace (Tystar 8300)
Tool Type Thermal Processing
Location Bay 4
Supervisor Tony Bosch
Supervisor Phone (805) 893-3486
Supervisor E-Mail bosch@ece.ucsb.edu
Description Tystar 8" 3-Tube Oxidation/Annealing System
Manufacturer Tystar Corporation
Model Tystar 8300
Sign up for this tool


About

The three stack Tystar 8” furnace is used primarily for 3 processes. The processes are dedicated for one tube each:

  • Tube #1: SOG curing
  • Tubes #2 and #3: Dry or wet oxidation of silicon
  • Tube #3: General furnace annealing

Each process tube can accomodate up to one hundred 8” wafers per cycle. We have boats for 2", 3", 4", 6", 8" and irregular shaped pieces. The maximum temperature is 1050°C for the system. Gases used are O2, Steam from DI-H2O, N2.

Process Information

Recipe Characterization Data, such as thermal oxidation times, can be found on the recipe page:

Use the BYU or Stanford Leland Jr. Thermal Oxidation Calculators to determine the time and temperature that will be necessary for your process needs. You can "calibrate" your oxidations to the Stanford calculator by adjusting the Partial Pressure on the calculator to match your experimental data.

Keep in mind that all process must be 30 minutes in length at a minimum. Processes less than 30 minutes will suffer from poor uniformity because the process tube will not have sufficient time to saturate with O2 or DI-H2O.

Recipes

The following are the available recipes on each furnace tube:

Tube 1:

  • SOG425.001 - Spin-On Glass Cure
  • ALGAAS.001 - Oxidation of AlGaAs

Tube 2:

  • WET1050.002 - WetOx at 1050°C
  • DRY1050.002 - DryOx at 1050°C
  • WETVAR.002 - WetOx, variable temp.
  • DRYVAR.002 - DryOx, variable temp.

Tube 3:

  • WET1050.003 - WetOx at 1050°C
  • DRY1050.003 - DryOx at 1050°C
  • WETVAR.003 - WetOx, variable temp.
  • DRYVAR.003 - DryOx, variable temp.
  • ANNEAL.003 - Anneal with variable time and temperature

Useful Information