ICP Etching Recipes: Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
→New test recipe: testing 2
Ningcao (talk | contribs)
added angled SiO2 sidewall recipe
Line 66: Line 66:
*[[media:33-Etching SiO2 with Vertical Side-wall.pdf|SiO<sub>2</sub> Vertical Etch Recipe#2 - CF<sub>4</sub>/CHF<sub>3</sub>]]
*[[media:33-Etching SiO2 with Vertical Side-wall.pdf|SiO<sub>2</sub> Vertical Etch Recipe#2 - CF<sub>4</sub>/CHF<sub>3</sub>]]
*[[media:Panasonic2-ICP-Plasma-Etch-SiO2-nanoscale-rev1.pdf|SiO<sub>2</sub> Nanoscale Etch Recipe - CHF<sub>3</sub>/O<sub>2</sub>]]
*[[media:Panasonic2-ICP-Plasma-Etch-SiO2-nanoscale-rev1.pdf|SiO<sub>2</sub> Nanoscale Etch Recipe - CHF<sub>3</sub>/O<sub>2</sub>]]
*[https://www.nanotech.ucsb.edu/wiki/images/1/1e/05-SiO2_Nano-structure_Etch.pdf Angled SiO2 sidewall recipe]
*[[TEST recipe]]


==SiN<sub>x</sub> Etching (Panasonic 2)==
==SiN<sub>x</sub> Etching (Panasonic 2)==

Revision as of 17:46, 6 April 2018

Back to Dry Etching Recipes.

DSEIII_(PlasmaTherm/Deep_Silicon_Etcher)

High Rate Bosch Etch (DSEIII)

Single-Step Low Etch Rate Smooth Sidewall Process (DSEIII)

Si Deep RIE (PlasmaTherm/Bosch Etch)

Bosch and Release Etch (Si Deep RIE)

  • Bosch and Release Processes
    • Ideal for deep (>>1µm), vertical etching of Silicon. Through-wafer etches are possible (requires carrier wafer).
    • Etch rate depends on area of exposed silicon being etched.
    • Al2O3 mask (ALD or Sputter) has >9000:1 selectivity
    • SiO2 (PECVD) mask has ~100:1 selectivity
    • Thermal SiO2 has ~300:1 selectivity.

Single-step Si Etching (not Bosch Process!) (Si Deep RIE)

ICP Etch 1 (Panasonic E626I)

SiO2 Etching (Panasonic 1)

SiNx Etching (Panasonic 1)

Al Etch (Panasonic 1)

Cr Etch (Panasonic 1)

Ti Etch (Panasonic 1)

W-TiW Etch (Panasonic 1)

GaAs-AlGaAs Etch (Panasonic 1)

GaN Etch (Panasonic 1)

SiC Etch (Panasonic 1)

Sapphire Etch (Panasonic 1)

ICP Etch 2 (Panasonic E640)

Recipes starting points for materials without processes listed can be obtained from Panasonic1 recipe files. The chambers are slightly different, but essentially the same, requiring only small program changes to obtain similar results.

SiO2 Etching (Panasonic 2)

SiNx Etching (Panasonic 2)

Al Etch (Panasonic 2)

Al2O3 Etching (Panasonic 2)

ALD Al2O3 Etch Rates in BCl3 Chemistry (click for plots of etch rate)

Contributed by Brian Markman, 2018

  • BCl3 = 30sccm
  • Pressure = 0.50 Pa
  • ICP Source RF = 500
  • Bias RF = 50W or 250W (250W can burn PR)
  • Cooling He Flow/Pressure = 15.0 sccm / 400 Pa
  • Etch Rate 50W: 0.66nm/sec
  • Etch Rate 250W: 1.0 nm/sec

GaAs Etch (Panasonic 2)

Photoresist and ARC etching

Basic recipes for etching photoresist and Anti-Reflection Coating (ARC) underlayers are as follows:

ARC Etching: DUV-42P or AR6

  • O2 = 40 sccm // 0.5 Pa
  • ICP = 75W // RF = 75W
  • 45 sec for full etching of DUV-42P (same for AR6)

UV6-0.8 Etching

Works very well for photoresist stripping

  • O2 = 40 sccm // 1.0 Pa
  • ICP = 350W // RF = 100W
  • Etch Rate = 518.5nm / 1min
  • 2m30sec to fully remove with ~50% overetch