+ | ==Reticle ID: "'''UCSB-OPC1'''"== |
− | This reticle is always installed in the system, in the "System Reticles" Box #1. |
+ | This reticle is always installed in the system, in the "System Reticles" Box #1. |
+ | The reticle contains alignment markers for various NanoFab lithography systems, along with resolution test structures and patterns for calibrating [https://en.wikipedia.org/wiki/Optical_proximity_correction optical proximity correction] on the system. Some patterns are proprietary to the mask designer, so we can not share the full GDS CAD file. |
||
− | === Alignment Markers === |
||
+ | |||
+ | ===Alignment Markers=== |
||
{| class="wikitable" |
{| class="wikitable" |
||
!Image ID |
!Image ID |
||
Line 24: | Line 26: | ||
has 1.1mm margin on all sides |
has 1.1mm margin on all sides |
||
|White is Chrome, Pattern is Clear[[File:Stepper align - Screen Shot 2018-07-23 at 11.24.31 AM.png|frameless|213x213px]] |
|White is Chrome, Pattern is Clear[[File:Stepper align - Screen Shot 2018-07-23 at 11.24.31 AM.png|frameless|213x213px]] |
||
+ | [[Media:GCA stepper align.gds|<small>CAD File (GDS):GCA_stepper_align.gds</small>]] |
||
|- |
|- |
||
|E-Beam Litho Alignment Mark - Positive |
|E-Beam Litho Alignment Mark - Positive |
||
Line 32: | Line 35: | ||
0.925mm margin on all sides |
0.925mm margin on all sides |
||
− | |||
− | >> Use this for Dicing Alginment Guides |
||
|White is Chrome, Pattern is Clear |
|White is Chrome, Pattern is Clear |
||
[[File:GlobalMulti POS - Screen Shot 2018-07-23 at 11.17.23 AM.png|frameless|145x145px]] |
[[File:GlobalMulti POS - Screen Shot 2018-07-23 at 11.17.23 AM.png|frameless|145x145px]] |
||
+ | |||
+ | [[Media:EBL-GlobalMulti POS.gds|<small>CAD File (GDS): EBL-GlobalMulti_POS.gds</small>]] |
||
|- |
|- |
||
|E-Beam Litho Alignment Mark - Negative |
|E-Beam Litho Alignment Mark - Negative |
||
Line 48: | Line 51: | ||
|Striped area is Clear |
|Striped area is Clear |
||
[[File:GlobalMulti NEG - Screen Shot 2018-07-23 at 11.22.19 AM.png|frameless|115x115px]] |
[[File:GlobalMulti NEG - Screen Shot 2018-07-23 at 11.22.19 AM.png|frameless|115x115px]] |
||
+ | |||
+ | [[Media:EBL-GlobalMulti NEG.gds|<small>CAD File (GDS): EBL-GlobalMulti_NEG.gds</small>]] |
||
|- |
|- |
||
|Contact_Mark |
|Contact_Mark |
||
Line 55: | Line 60: | ||
with 1.1mm margin on all sides |
with 1.1mm margin on all sides |
||
+ | |||
+ | Note the Polarity - will ''expose'' a ~550µm area. |
||
+ | |||
+ | Contact [[Demis D. John|Demis]] or [[Brian Thibeault|Brian]] for the CAD file for the male/female version of this mark. |
||
|White is Chrome, Striped area is Clear |
|White is Chrome, Striped area is Clear |
||
[[File:Align Front - Screen Shot 2018-07-23 at 11.32.16 AM.png|frameless|146x146px]] |
[[File:Align Front - Screen Shot 2018-07-23 at 11.32.16 AM.png|frameless|146x146px]] |
||
+ | |||
+ | [[Media:Contact-AlignFront.gds|<small>CAD File (GDS): Contact-AlignFront.gds</small>]] |
||
+ | |||
+ | [[Media:MA6-FrontBack AlignMarks only.gds|<small>Corresponding male/female alignment marks (GDS): MA6-FrontBack_AlignMarks_only.gds</small>]] |
||
|} |
|} |
||
− | === |
+ | ===Resolution Test Charts=== |
− | The Resolution test charts are repeated all across the reticle, in order to test for lens aberrations. You can have the system |
+ | The Resolution test charts are repeated all across the reticle, in order to test for lens aberrations. You can have the system expose only a single resolution chart, but since they are placed closely together on the reticle, it's very likely that partial shots of adjacent charts will also be exposed. |
− | In addition, the repeating cells allow us to test for the proper [https://en.wikipedia.org/wiki/Optical_proximity_correction optical proximity correction] algorithm. The Five <code>Dense_...</code> patterns are for calibrating the OPC algorithm, and are not for user analysis. |
+ | In addition, the repeating cells allow us to test for the proper [https://en.wikipedia.org/wiki/Optical_proximity_correction optical proximity correction] (OPC) algorithm. The Five <code>Dense_...</code> patterns are for calibrating the OPC algorithm, and are not for user analysis. |
+ | '''''Patterned area is CLEAR/transparent.''''' |
||
− | ==== Calibration Chart Layout ==== |
||
+ | |||
+ | ====Calibration Chart Layout==== |
||
Cell name is "''UCSB_Cal''", with coordinates below pointing to center of this cell.[[File:UCSB cal - Screen Shot 2018-07-23 at 12.06.58 PM.png|alt=Layout of the repeating calibration charts|none|thumb|512x512px|Layout of the repeating calibration charts]] |
Cell name is "''UCSB_Cal''", with coordinates below pointing to center of this cell.[[File:UCSB cal - Screen Shot 2018-07-23 at 12.06.58 PM.png|alt=Layout of the repeating calibration charts|none|thumb|512x512px|Layout of the repeating calibration charts]] |
||
− | ==== |
+ | ====Resolution Chart Schematic==== |
− | "resolution_chart_ORIG" cell in the above. |
+ | "resolution_chart_ORIG" cell in the above. '''''Blue/patterned area is CLEAR/transparent.''''' |
− | [[File:Resolution Chart - Screen Shot 2018-07-23 at 1.36.31 PM.png|alt=Resolution Chart Layout schematic|none|thumb|400x400px|Resolution Chart Layout, with res. test from 2.00µm to 0.130µm]] |
+ | [[File:Resolution Chart - Screen Shot 2018-07-23 at 1.36.31 PM.png|alt=Resolution Chart Layout schematic|none|thumb|400x400px|'''"resolution_chart_ORIG":''' Resolution Chart Layout, with res. test from 2.00µm to 0.130µm]] |
The ""resolution_chart_OPC" version has an optical proximity correction algorithm applied: |
The ""resolution_chart_OPC" version has an optical proximity correction algorithm applied: |
||
− | [[File:Resolution chart OPC - Screen Shot 2018-07-23 at 1.40.48 PM.png|alt=OPC'd Resolution Chart Layout schematic|none|thumb|395x395px|OPC'd Resolution Chart Layout, with res. test from 2.0µm to 0.130µm]] |
+ | [[File:Resolution chart OPC - Screen Shot 2018-07-23 at 1.40.48 PM.png|alt=OPC'd Resolution Chart Layout schematic|none|thumb|395x395px|'''"resolution_chart_OPC":''' OPC'd Resolution Chart Layout, with res. test from 2.0µm to 0.130µm]] |
+ | |||
+ | ====Coords for "resolution_chart" Calibration patterns==== |
||
+ | ''Image coords for each of the "resolution_chart_ORIG" cells. You can pick just one of these for shooting a resolution test structure. The purpose of the many different locations is to check for variations due to lens aberrations. You could just choose one near the center of the plate to test your process, or you could choose a chart that is in a similar location as the pattern you're shooting on your mask plate.'' |
||
+ | |||
+ | ''Note that some portion of the adjacent patterns will likely be exposed as well, due to the patterns not being surrounded by 1mm of chrome. Make sure you set your Cell Size large enough to make sure the bleed-over doesn't overlap with adjacent die.'' |
||
+ | {| class="wikitable" |
||
+ | !Image Size |
||
+ | X , Y |
||
+ | |||
+ | (Wafer, mm) |
||
+ | !Image Shift |
||
+ | X |
||
+ | |||
+ | (Wafer, mm) |
||
+ | !Image Shift |
||
+ | Y |
||
+ | |||
+ | (Wafer, mm) |
||
+ | |- |
||
+ | |0.605 , 1.005 |
||
+ | | -10.115000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | |same for each |
||
+ | | -7.415000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | |" " |
||
+ | | -4.715000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | |" " |
||
+ | | -2.015000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | |" " |
||
+ | |0.685000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | |" " |
||
+ | |3.385000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | |" " |
||
+ | |6.085000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | |" " |
||
+ | |8.785000 |
||
+ | |12.150000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | |9.450000 |
||
+ | |- |
||
+ | | |
||
+ | | colspan="2" |''Alignment Marker'' |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | |9.450000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | |9.450000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | |9.450000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | |9.450000 |
||
+ | |- |
||
+ | | |
||
+ | | colspan="2" |''Alignment Marker'' |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | |9.450000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | | |
||
+ | | -7.415000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |6.085000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | |6.750000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | | |
||
+ | | -7.415000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |6.085000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | |4.050000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | | |
||
+ | | -7.415000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |6.085000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | |1.350000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | | |
||
+ | | -7.415000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |6.085000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | | -1.350000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | | |
||
+ | | -7.415000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |6.085000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | | -4.050000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | | |
||
+ | | -7.415000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |6.085000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | | -6.750000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | | -9.450000 |
||
+ | |- |
||
+ | | |
||
+ | | colspan="2" |''Alignment Marker'' |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | | -9.450000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | | -9.450000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | | -9.450000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | | -9.450000 |
||
+ | |- |
||
+ | | |
||
+ | | colspan="2" |''Alignment Marker'' |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | | -9.450000 |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |- |
||
+ | | |
||
+ | | -10.115000 |
||
+ | | -12.150000 |
||
+ | |- |
||
+ | | |
||
+ | | -7.415000 |
||
+ | | -12.150000 |
||
+ | |- |
||
+ | | |
||
+ | | -4.715000 |
||
+ | | -12.150000 |
||
+ | |- |
||
+ | | |
||
+ | | -2.015000 |
||
+ | | -12.150000 |
||
+ | |- |
||
+ | | |
||
+ | |0.685000 |
||
+ | | -12.150000 |
||
+ | |- |
||
+ | | |
||
+ | |3.385000 |
||
+ | | -12.150000 |
||
+ | |- |
||
+ | | |
||
+ | |6.085000 |
||
+ | | -12.150000 |
||
+ | |- |
||
+ | | |
||
+ | |8.785000 |
||
+ | | -12.150000 |
||
+ | |} |
||
− | ==== |
+ | ====Coords for "UCSB_Cal" Calibration Patterns==== |
Each of the above "''UCSB_Cal''" cells, including all 7 patterns, is repeated on the following coordinates across the plate (coords are to the center of the "''UCSB_Cal''" cell): |
Each of the above "''UCSB_Cal''" cells, including all 7 patterns, is repeated on the following coordinates across the plate (coords are to the center of the "''UCSB_Cal''" cell): |
||
{| class="wikitable" |
{| class="wikitable" |
Latest revision as of 09:35, 1 May 2024
Reticle ID: "UCSB-OPC1"
This reticle is always installed in the system, in the "System Reticles" Box #1.
The reticle contains alignment markers for various NanoFab lithography systems, along with resolution test structures and patterns for calibrating optical proximity correction on the system. Some patterns are proprietary to the mask designer, so we can not share the full GDS CAD file.
Alignment Markers
Image ID | Image Size
X , Y (Wafer, mm) |
Image Shift
X , Y (Wafer, mm) |
Notes/Description | Schematics |
---|---|---|---|---|
GCA_Align | 0.530000 , 0.140000 | -6.750000 , 9.450000 | ImageShift references the center of the -X- "global" mark.
The ==||| "Local" mark is X+200µm to the right has 1.1mm margin on all sides |
White is Chrome, Pattern is Clear |
E-Beam Litho Alignment Mark - Positive | 0.900000 , 0.900000 | -6.750000 , -9.450000 | ImageShift is the center coords of the larger "+" mark
Smaller "+" mark is (0.225,-0.225)mm down-right 0.925mm margin on all sides |
White is Chrome, Pattern is Clear |
E-Beam Litho Alignment Mark - Negative | 0.710000 , 0.710000 | 6.750000 , -9.450000 | ImageShift is the center coords of the larger "+" mark
Smaller "+" mark is (0.225,-0.225)mm down-right Blank (masked) space on left+top sides 1.0mm margin on all sides |
Striped area is Clear |
Contact_Mark | 0.564000 , 0.564000 | 6.750000 , 9.450000 | ImageShift references the center of the contact alignment mark "+"
with 1.1mm margin on all sides Note the Polarity - will expose a ~550µm area. Contact Demis or Brian for the CAD file for the male/female version of this mark. |
White is Chrome, Striped area is Clear
CAD File (GDS): Contact-AlignFront.gds Corresponding male/female alignment marks (GDS): MA6-FrontBack_AlignMarks_only.gds |
Resolution Test Charts
The Resolution test charts are repeated all across the reticle, in order to test for lens aberrations. You can have the system expose only a single resolution chart, but since they are placed closely together on the reticle, it's very likely that partial shots of adjacent charts will also be exposed.
In addition, the repeating cells allow us to test for the proper optical proximity correction (OPC) algorithm. The Five Dense_...
patterns are for calibrating the OPC algorithm, and are not for user analysis.
Patterned area is CLEAR/transparent.
Calibration Chart Layout
Cell name is "UCSB_Cal", with coordinates below pointing to center of this cell.
Resolution Chart Schematic
"resolution_chart_ORIG" cell in the above. Blue/patterned area is CLEAR/transparent.
The ""resolution_chart_OPC" version has an optical proximity correction algorithm applied:
Coords for "resolution_chart" Calibration patterns
Image coords for each of the "resolution_chart_ORIG" cells. You can pick just one of these for shooting a resolution test structure. The purpose of the many different locations is to check for variations due to lens aberrations. You could just choose one near the center of the plate to test your process, or you could choose a chart that is in a similar location as the pattern you're shooting on your mask plate.
Note that some portion of the adjacent patterns will likely be exposed as well, due to the patterns not being surrounded by 1mm of chrome. Make sure you set your Cell Size large enough to make sure the bleed-over doesn't overlap with adjacent die.
Image Size
X , Y (Wafer, mm) |
Image Shift
X (Wafer, mm) |
Image Shift
Y (Wafer, mm) |
---|---|---|
0.605 , 1.005 | -10.115000 | 12.150000 |
same for each | -7.415000 | 12.150000 |
" " | -4.715000 | 12.150000 |
" " | -2.015000 | 12.150000 |
" " | 0.685000 | 12.150000 |
" " | 3.385000 | 12.150000 |
" " | 6.085000 | 12.150000 |
" " | 8.785000 | 12.150000 |
-10.115000 | 9.450000 | |
Alignment Marker | ||
-4.715000 | 9.450000 | |
-2.015000 | 9.450000 | |
0.685000 | 9.450000 | |
3.385000 | 9.450000 | |
Alignment Marker | ||
8.785000 | 9.450000 | |
-10.115000 | 6.750000 | |
-7.415000 | 6.750000 | |
-4.715000 | 6.750000 | |
-2.015000 | 6.750000 | |
0.685000 | 6.750000 | |
3.385000 | 6.750000 | |
6.085000 | 6.750000 | |
8.785000 | 6.750000 | |
-10.115000 | 4.050000 | |
-7.415000 | 4.050000 | |
-4.715000 | 4.050000 | |
-2.015000 | 4.050000 | |
0.685000 | 4.050000 | |
3.385000 | 4.050000 | |
6.085000 | 4.050000 | |
8.785000 | 4.050000 | |
-10.115000 | 1.350000 | |
-7.415000 | 1.350000 | |
-4.715000 | 1.350000 | |
-2.015000 | 1.350000 | |
0.685000 | 1.350000 | |
3.385000 | 1.350000 | |
6.085000 | 1.350000 | |
8.785000 | 1.350000 | |
-10.115000 | -1.350000 | |
-7.415000 | -1.350000 | |
-4.715000 | -1.350000 | |
-2.015000 | -1.350000 | |
0.685000 | -1.350000 | |
3.385000 | -1.350000 | |
6.085000 | -1.350000 | |
8.785000 | -1.350000 | |
-10.115000 | -4.050000 | |
-7.415000 | -4.050000 | |
-4.715000 | -4.050000 | |
-2.015000 | -4.050000 | |
0.685000 | -4.050000 | |
3.385000 | -4.050000 | |
6.085000 | -4.050000 | |
8.785000 | -4.050000 | |
-10.115000 | -6.750000 | |
-7.415000 | -6.750000 | |
-4.715000 | -6.750000 | |
-2.015000 | -6.750000 | |
0.685000 | -6.750000 | |
3.385000 | -6.750000 | |
6.085000 | -6.750000 | |
8.785000 | -6.750000 | |
-10.115000 | -9.450000 | |
Alignment Marker | ||
-4.715000 | -9.450000 | |
-2.015000 | -9.450000 | |
0.685000 | -9.450000 | |
3.385000 | -9.450000 | |
Alignment Marker | ||
8.785000 | -9.450000 | |
-10.115000 | -12.150000 | |
-7.415000 | -12.150000 | |
-4.715000 | -12.150000 | |
-2.015000 | -12.150000 | |
0.685000 | -12.150000 | |
3.385000 | -12.150000 | |
6.085000 | -12.150000 | |
8.785000 | -12.150000 |
Coords for "UCSB_Cal" Calibration Patterns
Each of the above "UCSB_Cal" cells, including all 7 patterns, is repeated on the following coordinates across the plate (coords are to the center of the "UCSB_Cal" cell):
Image Size
X , Y (Wafer, mm) |
Image Shift
X (Wafer, mm) |
Image Shift
Y (Wafer, mm) |
---|---|---|
2.610000 , 2.610000 | -9.450000 | 12.150000 |
same as above | -6.750000 | 12.150000 |
" " | -4.050000 | 12.150000 |
" " | -1.350000 | 12.150000 |
" " | 1.350000 | 12.150000 |
" " | 4.050000 | 12.150000 |
" " | 6.750000 | 12.150000 |
" " | 9.450000 | 12.150000 |
-9.450000 | 9.450000 | |
Alignment Marker | ||
-4.050000 | 9.450000 | |
-1.350000 | 9.450000 | |
1.350000 | 9.450000 | |
4.050000 | 9.450000 | |
Alignment Marker | ||
9.450000 | 9.450000 | |
-9.450000 | 6.750000 | |
-6.750000 | 6.750000 | |
-4.050000 | 6.750000 | |
-1.350000 | 6.750000 | |
1.350000 | 6.750000 | |
4.050000 | 6.750000 | |
6.750000 | 6.750000 | |
9.450000 | 6.750000 | |
-9.450000 | 4.050000 | |
-6.750000 | 4.050000 | |
-4.050000 | 4.050000 | |
-1.350000 | 4.050000 | |
1.350000 | 4.050000 | |
4.050000 | 4.050000 | |
6.750000 | 4.050000 | |
9.450000 | 4.050000 | |
-9.450000 | 1.350000 | |
-6.750000 | 1.350000 | |
-4.050000 | 1.350000 | |
-1.350000 | 1.350000 | |
1.350000 | 1.350000 | |
4.050000 | 1.350000 | |
6.750000 | 1.350000 | |
9.450000 | 1.350000 | |
-9.450000 | -1.350000 | |
-6.750000 | -1.350000 | |
-4.050000 | -1.350000 | |
-1.350000 | -1.350000 | |
1.350000 | -1.350000 | |
4.050000 | -1.350000 | |
6.750000 | -1.350000 | |
9.450000 | -1.350000 | |
-9.450000 | -4.050000 | |
-6.750000 | -4.050000 | |
-4.050000 | -4.050000 | |
-1.350000 | -4.050000 | |
1.350000 | -4.050000 | |
4.050000 | -4.050000 | |
6.750000 | -4.050000 | |
9.450000 | -4.050000 | |
-9.450000 | -6.750000 | |
-6.750000 | -6.750000 | |
-4.050000 | -6.750000 | |
-1.350000 | -6.750000 | |
1.350000 | -6.750000 | |
4.050000 | -6.750000 | |
6.750000 | -6.750000 | |
9.450000 | -6.750000 | |
-9.450000 | -9.450000 | |
Alignment Marker | ||
-4.050000 | -9.450000 | |
-1.350000 | -9.450000 | |
1.350000 | -9.450000 | |
4.050000 | -9.450000 | |
Alignment Marker | ||
9.450000 | -9.450000 | |
-9.450000 | -12.150000 | |
-6.750000 | -12.150000 | |
-4.050000 | -12.150000 | |
-1.350000 | -12.150000 | |
1.350000 | -12.150000 | |
4.050000 | -12.150000 | |
6.750000 | -12.150000 | |
9.450000 | -12.150000 |