Difference between revisions of "Wafer Coating Process Traveler"
Line 11: | Line 11: | ||
** Get rates from [[PECVD Recipes#Thin-Film Properties 6|historical data]]. |
** Get rates from [[PECVD Recipes#Thin-Film Properties 6|historical data]]. |
||
* Cleaning recipe name: Post dep PD, t=VARIABLE sec |
* Cleaning recipe name: Post dep PD, t=VARIABLE sec |
||
+ | ** Get rates for [https://wiki.nanotech.ucsb.edu/wiki/ICP-PECVD_(Unaxis_VLR)#Documentation cleaning time in Operating Procedure] . |
||
− | ** Get rates from cleaning recipe. |
||
There is only one SiO2 seasoning recipe. This same recipe is used for seasoning for both LDR and HDR depositions. The goal in this step is to prepare the chamber and deposit ~200nm of SiO2 on walls. |
There is only one SiO2 seasoning recipe. This same recipe is used for seasoning for both LDR and HDR depositions. The goal in this step is to prepare the chamber and deposit ~200nm of SiO2 on walls. |
||
Revision as of 10:57, 22 April 2020
Unaxis deposition - SiO2 LDR film @250°C
a) Prepare three 4” wafers:
- for seasoning (regular Si wafer ~500nm thick)
- for deposition (your wafer for deposition)
- for cleaning ( Si thick wafer~0.8-1um)
b) Load all wafers in a cassette, making sure major flat is facing outside (toward you). Try to align flats for all three wafers.
c) Edit standard recipes (for seasoning, deposition, and cleaning). You can ONLY change the time in recipes.
- Seasoning recipe name: SiO2 seasoning, t=2min
- Deposition recipe name: SiO2 LDR 250°C, t=VARIABLE sec
- Get rates from historical data.
- Cleaning recipe name: Post dep PD, t=VARIABLE sec
- Get rates for cleaning time in Operating Procedure .
There is only one SiO2 seasoning recipe. This same recipe is used for seasoning for both LDR and HDR depositions. The goal in this step is to prepare the chamber and deposit ~200nm of SiO2 on walls.
d) Run the job (name the job/select the wafer, execute the job)
After loading the wafers and selecting for each wafer correct recipe, you could leave and come back when all runs are finished. You should monitor process at the very beginning to make sure there are no any issues with the run.
e) Unload wafers ONLY when all runs are finished.
Unaxis deposition - 300nm SiO2 HDR film @250°C
a) Prepare three 4” wafers:
- for seasoning (regular Si wafer ~500nm thick)
- for deposition (your wafer for deposition)
- for cleaning ( Si thick wafer~0.8-1um )
b) Load all wafers in a cassette, making sure major flat is facing outside (toward you). Try to align flats for all three wafers.
c) Edit standard recipes (for seasoning, deposition, and cleaning). You can ONLY change the time in recipes.
- Seasoning recipe name: SiO2 seasoning, t=2min
- Deposition recipe name: SiO2 HDR 250°C, t=180 sec
- Cleaning recipe name: Post dep PD, t=900sec
There is only one SiO2 seasoning recipe. This same recipe is used for seasoning for both LDR and HDR depositions. The goal in this step is to prepare the chamber and deposit ~200nm of SiO2 on walls.
d) Run the job (name the job/select the wafer, execute the job)
After loading the wafers and selecting for each wafer correct recipe, you could leave and come back when all runs are finished. You should monitor process at the very beginning to make sure there are no any issues with the run.
e) Unload wafers ONLY when all runs are finished.
Unaxis deposition - 300nm SiN film @250°C
a) Prepare three 4” wafers:
- for seasoning (regular Si wafer ~500nm thick)
- for deposition (your wafer for deposition)
- for cleaning ( Si thick wafer~0.8-1um)
b) Load all wafers in a cassette, making sure major flat is facing outside (toward you). Try to align flats for all three wafers.
c) Edit standard recipes (for seasoning, deposition, and cleaning). You can ONLY change the time in recipes.
- Seasoning recipe name: SiN seasoning, t=5min
- Deposition recipe name: SiN 250°C, t=480 sec
- Cleaning recipe name: Post dep PD, t=1500sec
d) Run the job (name the job/select the wafer, execute the job)
After loading the wafers and selecting for each wafer correct recipe, you could leave and come back when all runs are finished. You should monitor process at the very beginning to make sure there are no any issues with the run.
e) Unload wafers ONLY when all runs are finished.
Unaxis deposition - 300nm SiN LS film @250°C
a) Prepare three 4”wafers:
- for seasoning (regular Si wafer ~500nm thick)
- for deposition (your wafer for deposition)
- for cleaning ( Si thick wafer~0.8-1um )
b) Load all wafers in a cassette, making sure major flat is facing outside (toward you). Try to align flats for all three wafers.
c) Edit standard recipes (for seasoning, deposition, and cleaning). You can ONLY change the time in recipes.
- Seasoning recipe name: SiN seasoning, t=2min
- Deposition recipe name: SiN LS 250°C, t=180 sec
- Cleaning recipe name: Post dep PD, t=1500sec
d) Run the job (name the job/select the wafer, execute the job)
After loading the wafers and selecting for each wafer correct recipe, you could leave and come back when all runs are finished. You should monitor process at the very beginning to make sure there are no any issues with the run.
e) Unload wafers ONLY when all runs are finished.