OLD - PECVD2 Recipes: Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
Biljana (talk | contribs)
m John d moved page PECVD2 Recipes to OLD - PECVD2 Recipes without leaving a redirect: this page is not linked anywhere, looks like 2014 data only
 
(10 intermediate revisions by one other user not shown)
Line 3: Line 3:
==Photos==
==Photos==


*[[media: Dirty platen photos.pdf|Dirty platen]]
*[[media: Photos Sheet1.pdf|Dirty platen]]

⚫
== SiN deposition (PECVD #2) ==
⚫
*[[media:New Advanced PECVD-Nitride2 300C standard recipe.pdf|Nitride2 Standard Recipe]]
⚫
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dDE0R3FlenNPa2txNmRldTczMXZNNnc#gid=sharing Nitride2 Data Devcember 2014]
⚫
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dERkYm85bGtYQVpjVk5GTGJuMkg2anc&usp=drive_web#gid=sharing Nitride2 Thickness uniformity 2014]

*Deposition Rate: ≈ 7.93 nm/min (users must calibrate this prior to critical deps)
*Refractive Index: ≈ 1.961
*Stress ≈ 495MPa
*HF etch rate:~49nm/min


== SiO<sub>2</sub> deposition (PECVD #2) ==
== SiO<sub>2</sub> deposition (PECVD #2) ==
*[[media:Advanced PECVD OXIDE 300C standard recipe OXIDE Standard Recipe.pdf|Oxide Standard Recipe]]
*[[media:NewAdvPECVD OXIDE 300C standard recipe.pdf|Oxide Standard Recipe]]
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dGJZaGtDTVRqa1BRdW5iU1N1Y01jd0E&usp=drive_web#gid=sharing Oxide Data December 2014]
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dGJZaGtDTVRqa1BRdW5iU1N1Y01jd0E&usp=drive_web#gid=sharing Oxide Data 2014]
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dDRjSHEtYUZRUTg4azE2U0JGUTg2M1E#gid=sharing Oxide Thickness uniformity 2014]
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dDRjSHEtYUZRUTg4azE2U0JGUTg2M1E#gid=sharing Oxide Thickness uniformity 2014]


⚫
== SiN deposition (PECVD #2) ==
*Deposition Rate: ≈ 28.64 nm/min (users must calibrate this prior to critical deps)
⚫
*[[media:NewAdvPECVD-Nitride2 300C standard recipe.pdf|Nitride2 Standard Recipe]]
*Refractive Index: ≈ 1.475
⚫
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dDE0R3FlenNPa2txNmRldTczMXZNNnc#gid=sharing Nitride2 Data 2014]
*Stress ≈ -260MPa
⚫
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dERkYm85bGtYQVpjVk5GTGJuMkg2anc&usp=drive_web#gid=sharing Nitride2 Thickness uniformity 2014]
*HF etch rate~623nm/min


==LS SiN deposition (PECVD #2) ==
==LS SiN deposition (PECVD #2) ==
⚫
*[[media:NewAdvPECVD-LS Nitride2 300C standard recipe.pdf|LS Nitride2 Standard Recipe]]

⚫
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dEZvWVhzS1pHUXZkOGcyQWZ4LTNBWGc&usp=drive_web#gid=sharing LS Nitride2 Data 2014]
⚫
*[[media:AdvPECVD-LS Nitride2 300C standard recipe.pdf|LS Nitride2 Standard Recipe]]
⚫
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dEZvWVhzS1pHUXZkOGcyQWZ4LTNBWGc&usp=drive_web#gid=sharing LS Nitride2 Data December 2014]
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dENSak1ZNnVaVTFEQTBzdDJMSDlDTFE&usp=drive_web#gid=1=sharing LS Nitride2 Thickness uniformity 2014]
*[https://docs.google.com/spreadsheet/ccc?key=0AnwBU1s4JQo2dENSak1ZNnVaVTFEQTBzdDJMSDlDTFE&usp=drive_web#gid=1=sharing LS Nitride2 Thickness uniformity 2014]

*Deposition Rate: ≈ 7.87 nm/min (users must calibrate this prior to critical deps)
*Refractive Index: ≈ 1.938
*Stress ≈ 1.76MPa
*HF etch rate~47nm/min