Surface Analysis (KLA/Tencor Surfscan): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
Content deleted Content added
Biljana (talk | contribs)
Biljana (talk | contribs)
 
(24 intermediate revisions by 2 users not shown)
Line 6: Line 6:
|location=Bay 5
|location=Bay 5
|description = Surface Analysis
|description = Surface Analysis
KLA/Tencor Surfscan
|manufacturer = KLA/Tencor
|manufacturer = Tencor
|materials =
|materials =
|toolid=
|toolid=
Line 25: Line 24:


*[https://wiki.nanofab.ucsb.edu/w/images/2/29/Wafer_Particle_Count-Process_Traveler.pdf Wafer Scanning Instructions]
*[https://wiki.nanofab.ucsb.edu/w/images/2/29/Wafer_Particle_Count-Process_Traveler.pdf Wafer Scanning Instructions]
**''This is the procedure Staff uses to calibrate particle counts on our deposition tools. After the scan is complete you need to press ENTER on a small laptop. The Data will be saved on:
**''This is the procedure Staff uses to calibrate particle counts on our deposition tools. After the scan is complete you need to press ENTER on a small laptop. The Data will be saved on: Desktop/Local Disk(C)/Users/Public/Public Documentation] The data will be saved as a photo.''
Desktop/Local Disk (C)/Users/Public/Public Documentation/Users Nanofiles/YOUR FOLDER. The data will be saved as a photo.
''


⚫
*[[Surfscan SOP for small substrates]] (''You must water-mount your small sample or wafer (2inch or 3 inch) to a carrier wafer)''
* Surfscan instructions [wafers different size, and pieces]:


*[[Surfscan SOP for 4inch wafers]]
*[https://drive.google.com/drive/folders/1OcByelh_v8a2CezFyC7w5lTMMdLgVy_a 8inch wafers]
*[https://drive.google.com/drive/folders/1OcByelh_v8a2CezFyC7w5lTMMdLgVy_a 6inch wafers]
*[https://drive.google.com/drive/folders/1OcByelh_v8a2CezFyC7w5lTMMdLgVy_a 4inch wafers]
*[https://drive.google.com/drive/folders/1OcByelh_v8a2CezFyC7w5lTMMdLgVy_a 2inch 3inch and small samples]


*[[Surfscan SOP for 6inch wafers]]
⚫
**''You must water-mount your small sample or wafer(2inch or 3 inch) to a 4-inch wafer.''

⚫
*[[Wafer scanning process traveler|Wafer Particle Count - Process Traveler]]- Very detailed instructions
*[[Surfscan SOP for 8inch wafers]]

⚫
* [[Wafer scanning process traveler|Wafer Particle Count - Process Traveler]]
**'' Very detailed instructions


===Other Documentation===
===Other Documentation===

Latest revision as of 23:04, 20 November 2024

Surface Analysis (KLA/Tencor Surfscan)
Location Bay 5
Tool Type Inspection, Test and Characterization
Manufacturer KLA/Tencor
Description Surface Analysis

Primary Supervisor Biljana Stamenic
(805) 893-4002
biljana@ece.ucsb.edu

Secondary Supervisor

Don Freeborn


Recipes


About

This system uses a laser-based scattering method to count size and distribution of particles (or other scattering defects) on a flat wafer surface.

It can scan wafers in size from 4 to 8 inches. Piece-parts are more difficult but can be scanned with a custom recipe.

4-inch wafers are the most standard size to measure.

For measuring very low particle counts accurately, purchase "low particle count" (LPC) wafers from a Silicon wafer vendor, and keep the wafers in the case and clean at all times until use.

Documentation

Operating Procedures

  • Wafer Scanning Instructions
    • This is the procedure Staff uses to calibrate particle counts on our deposition tools. After the scan is complete you need to press ENTER on a small laptop. The Data will be saved on: Desktop/Local Disk(C)/Users/Public/Public Documentation] The data will be saved as a photo.

Other Documentation

  • Operations Manual
    • For detailed measurement info, it is highly recommended that you read the manual.

Examples

A low-particle 4-inch wafer example:
Gain 4: Small Particles

(0.160µm – 1.60µm)

Gain 2: Large Particles

(1.60µm – 28.0µm)

A high-particle 4-inch wafer example:
Gain 4: Small Particles

(0.160µm – 1.60µm)

Gain 2: Large Particles

(1.60µm – 28.0µm)