{{tool|{{PAGENAME}}
{{tool2|{{PAGENAME}}
|picture=IBD.jpg
|picture=IBD.jpg
|type = Vacuum Deposition
|type = Vacuum Deposition
|super= Mike Silva
|super= Michael Barreraz
|super2= Bill Millerski
|phone=(805)839-3918x219
|location=Bay 3
|location=Bay 4
|email=silva@ece.ucsb.edu
|description = Ion Beam Deposition Tool
|description = Ion Beam Deposition Tool
|manufacturer = Veeco
|manufacturer = Veeco
Line 11: Line 10:
|toolid=14
|toolid=14
}}
}}

= About =
=About=
This tool is designed for high quality, precise, reproducible deposition of dielectric films for optical quality films. Metallic material is ion bombarded from a target material and a reactive ion beam of oxygen and/or nitrogen is aimed at the surface, creating a metal-oxide or nitride on the sample.
This tool is designed for high quality, precise, reproducible deposition of dielectric films for optical quality films. Metallic material is ion bombarded from a target material and a reactive ion beam of oxygen and/or nitrogen is aimed at the surface, creating a metal-oxide or nitride on the sample.
The system is fully computer controlled to facilitate multi-layer stack recipes for high reflectivity or low reflectivity coatings.
The system is fully computer controlled to facilitate multi-layer stack recipes for high reflectivity or low reflectivity coatings.
Line 22: Line 22:
The most common films for High-/Anti-Reflection (HR/AR) coatings are '''SiO<sub>2</sub>''' & '''Ta<sub>2</sub>O<sub>5</sub>''', both of which are extremely stable w/r/to refractive index. Users typically calibrate their dep. rates prior to critical deps. & multi-layer coatings.
The most common films for High-/Anti-Reflection (HR/AR) coatings are '''SiO<sub>2</sub>''' & '''Ta<sub>2</sub>O<sub>5</sub>''', both of which are extremely stable w/r/to refractive index. Users typically calibrate their dep. rates prior to critical deps. & multi-layer coatings.


= Detailed Specifications =
=Detailed Specifications=


⚫
*Full programmable control through GUI, designed for multi-layer optical coatings
*Xenon target bombardment
*Xenon sputter-target bombardment
*Nitrogen and/or Oxygen deposition assist source for metal oxides/nitrides
*Nitrogen and/or Oxygen deposition assist source for metal oxides/nitrides
*Targets: Ta, Si, Al, Ti, ITO
*Currently installed Sputter Targets: '''Ta, Si, Al, Ti''' (ITO target on-hand but not installed)
⚫
*Standard Recipes: Ta<sub>2</sub>O<sub>5</sub>, SiO<sub>2</sub>, Si<sub>3</sub>N<sub>4</sub>, SiO<sub>x</sub>N<sub>y</sub>, TiO<sub>2</sub>, Al<sub>2</sub>O<sub>3</sub>.
⚫
*Pieces up to 6” wafers accepted
**Custom recipes: AlN, TaN, TiN
⚫
*Full programmable control through GUI for multi-layer optical coatings
*Argon sample cleaning/pre-sputtering with low energy/etch rate
⚫
*Standard Recipes: Ta<sub>2</sub>O<sub>5</sub> (1.33 A/s rate), SiO<sub>2</sub> (0.85A/s), Si<sub>3</sub>N<sub>4</sub>, TiO<sub>2</sub>.
** Custom recipes: Al<sub>2</sub>O<sub>3</sub>, SiO<sub>x</sub>N<sub>y</sub>, AlN, TaN
*Base Pressure ≤ 3e-8 Torr
*Base Pressure ≤ 3e-8 Torr
⚫
*Accepts small pieces, up to 6” wafers accepted
⚫
*High quality, High Reflectivity DBR (>99.5%) mirrors demonstrated in blue wavelengths
**Holders available for etched/cleaved facet coating
*High uniformity over 6-inch area, high density films, low hydrogen content. Higher particle count.
⚫
*High quality, High Reflectivity DBR (>99.5%) mirrors demonstrated at UV-Vis & near-IR wavelengths
*Wide-band Anti-Reflection coatings demonstrated in near-IR (~1550nm)
*Wide-band Anti-Reflection coatings demonstrated in near-IR (~1550nm)


=Procedures=
=Documentation=

*[[media:IBDInstructions.pdf|Operating Instructions]]
*{{file|IBDoperationguide1.pdf|Operating Instructions Part 1}}
*{{file|IBDoperationguide2.pdf|Operating Instructions Part 2}}
*[https://wiki.nanofab.ucsb.edu/w/images/1/18/IBD_Troubleshooting_Guide.pdf Troubleshooting Guide] *
*[[IBD: Calibrating Optical Thickness|Method for accurately calibrating multi-layer optical films]]
**For example, for calibrating and depositing Multi-layer DBR gratings, Anti-Reflection coatings etc.
**Uses Fabry-Perot depositions to calibration "optical thickness" directly, similar to VCSEL growth calibrations.
*[https://wiki.nanofab.ucsb.edu/w/images/1/1c/IBD_SOP.pdf IBD Standard Operating procedure] *

==Recipes==

*[[Sputtering_Recipes#Ion_Beam_Deposition_.28Veeco_NEXUS.29|Sputtering Recipes: Ion Beam Deposition (Veeco NEXUS)]]
**''Dep Rates, Refractive Indices, Stress etc. for various available dielectric coatings''

==Schematics/Examples==
[[File:IBD Schematic.jpg|alt=Illustration showing wafer, assist beam and sputter/depo beam|none|thumb|Top-Down Schematic of the Ion Beam Deposition system, where (1) is the substrate/wafer, (2) is the Depo/Sputter ion beam, (3) is the variable sputter target (either Si, Ta, Al, or Ti), and (4) is the Assist Ion Beam (Ar, O<sub>2</sub>, or N<sub>2</sub> beam). ''Image'' ''Author: [[Demis D. John]], 2010'']]
<br />
[[File:IBD 9-period DBR - Reflectivity Spectrum (EMpy).png|alt=Screenshot of optical reflection spectrum|none|thumb|Example reflectivity of a 9-period SiO2/Ta2O5 DBR, with high reflectivity at ~1050nm, measured with the Filmetrics F10-RT]]

Latest revision as of 22:16, 8 May 2025

Ion Beam Deposition (Veeco NEXUS)
Location Bay 4
Tool Type Vacuum Deposition
Manufacturer Veeco
Description Ion Beam Deposition Tool

Primary Supervisor Michael Barreraz
(805) 893-4147
mikebarreraz@ece.ucsb.edu

Secondary Supervisor

Bill Millerski


Recipes Vacuum Deposition Recipes

SignupMonkey: Sign up for this tool


About

This tool is designed for high quality, precise, reproducible deposition of dielectric films for optical quality films. Metallic material is ion bombarded from a target material and a reactive ion beam of oxygen and/or nitrogen is aimed at the surface, creating a metal-oxide or nitride on the sample. The system is fully computer controlled to facilitate multi-layer stack recipes for high reflectivity or low reflectivity coatings. The system is load locked and can handle wafers up to 6” in diameter as well as small pieces. Sample rotation and angling is used to facilitate material quality and allows for sidewall coverage on non-planar surfaces. Uniformity is better than 1.5% over 6" wafers and reproducibility is expected to be within one percent (with pre-dep calibration). The high-energy sputter deposition produces denser films than other techniques, improving optical damage threshold (and thus coating lifetime). This also causes the refractive indices to be slightly higher than comparable stoichiometric films deposited by other techniques.

Four metallic targets can be installed (producing Oxides & Nitrides of each). Ta & Si are always available, the others are rotated out as users need them.

The most common films for High-/Anti-Reflection (HR/AR) coatings are SiO2 & Ta2O5, both of which are extremely stable w/r/to refractive index. Users typically calibrate their dep. rates prior to critical deps. & multi-layer coatings.

Detailed Specifications

  • Full programmable control through GUI, designed for multi-layer optical coatings
  • Xenon sputter-target bombardment
  • Nitrogen and/or Oxygen deposition assist source for metal oxides/nitrides
  • Currently installed Sputter Targets: Ta, Si, Al, Ti (ITO target on-hand but not installed)
  • Standard Recipes: Ta2O5, SiO2, Si3N4, SiOxNy, TiO2, Al2O3.
    • Custom recipes: AlN, TaN, TiN
  • Argon sample cleaning/pre-sputtering with low energy/etch rate
  • Base Pressure ≤ 3e-8 Torr
  • Accepts small pieces, up to 6” wafers accepted
    • Holders available for etched/cleaved facet coating
  • High uniformity over 6-inch area, high density films, low hydrogen content. Higher particle count.
  • High quality, High Reflectivity DBR (>99.5%) mirrors demonstrated at UV-Vis & near-IR wavelengths
  • Wide-band Anti-Reflection coatings demonstrated in near-IR (~1550nm)

Procedures

Recipes

Schematics/Examples

Illustration showing wafer, assist beam and sputter/depo beam
Top-Down Schematic of the Ion Beam Deposition system, where (1) is the substrate/wafer, (2) is the Depo/Sputter ion beam, (3) is the variable sputter target (either Si, Ta, Al, or Ti), and (4) is the Assist Ion Beam (Ar, O2, or N2 beam). Image Author: Demis D. John, 2010


Screenshot of optical reflection spectrum
Example reflectivity of a 9-period SiO2/Ta2O5 DBR, with high reflectivity at ~1050nm, measured with the Filmetrics F10-RT