Ion Beam Deposition (Veeco NEXUS): Difference between revisions
→Schematics/Examples: added example DBR spectrum |
→Documentation: moved DBR calibration into "documentation" section |
||
| Line 38: | Line 38: | ||
*Wide-band Anti-Reflection coatings demonstrated in near-IR (~1550nm) |
*Wide-band Anti-Reflection coatings demonstrated in near-IR (~1550nm) |
||
=Procedures= |
|||
=Documentation= |
|||
*{{file|IBDoperationguide1.pdf|Operating Instructions Part 1}} |
*{{file|IBDoperationguide1.pdf|Operating Instructions Part 1}} |
||
*{{file|IBDoperationguide2.pdf|Operating Instructions Part 2}} |
*{{file|IBDoperationguide2.pdf|Operating Instructions Part 2}} |
||
| ⚫ | |||
| ⚫ | |||
**For example, for calibrating and depositing Multi-layer DBR gratings, Anti-Reflection coatings etc. |
|||
**Uses Fabry-Perot depositions to calibration "optical thickness" directly, similar to VCSEL growth calibrations. |
|||
| ⚫ | |||
==Recipes== |
==Recipes== |
||
*[[Sputtering_Recipes#Ion_Beam_Deposition_.28Veeco_NEXUS.29|Sputtering Recipes: Ion Beam Deposition (Veeco NEXUS)]] |
*[[Sputtering_Recipes#Ion_Beam_Deposition_.28Veeco_NEXUS.29|Sputtering Recipes: Ion Beam Deposition (Veeco NEXUS)]] |
||
**''Dep Rates, Refractive Indices etc. for various available dielectric coatings'' |
**''Dep Rates, Refractive Indices, Stress etc. for various available dielectric coatings'' |
||
| ⚫ | |||
| ⚫ | |||
==Other documents== |
|||
| ⚫ | |||
==Schematics/Examples== |
==Schematics/Examples== |
||
Latest revision as of 22:16, 8 May 2025
| ||||||||||||||||||||||||||||||
About
This tool is designed for high quality, precise, reproducible deposition of dielectric films for optical quality films. Metallic material is ion bombarded from a target material and a reactive ion beam of oxygen and/or nitrogen is aimed at the surface, creating a metal-oxide or nitride on the sample. The system is fully computer controlled to facilitate multi-layer stack recipes for high reflectivity or low reflectivity coatings. The system is load locked and can handle wafers up to 6” in diameter as well as small pieces. Sample rotation and angling is used to facilitate material quality and allows for sidewall coverage on non-planar surfaces. Uniformity is better than 1.5% over 6" wafers and reproducibility is expected to be within one percent (with pre-dep calibration). The high-energy sputter deposition produces denser films than other techniques, improving optical damage threshold (and thus coating lifetime). This also causes the refractive indices to be slightly higher than comparable stoichiometric films deposited by other techniques.
Four metallic targets can be installed (producing Oxides & Nitrides of each). Ta & Si are always available, the others are rotated out as users need them.
The most common films for High-/Anti-Reflection (HR/AR) coatings are SiO2 & Ta2O5, both of which are extremely stable w/r/to refractive index. Users typically calibrate their dep. rates prior to critical deps. & multi-layer coatings.
Detailed Specifications
- Full programmable control through GUI, designed for multi-layer optical coatings
- Xenon sputter-target bombardment
- Nitrogen and/or Oxygen deposition assist source for metal oxides/nitrides
- Currently installed Sputter Targets: Ta, Si, Al, Ti (ITO target on-hand but not installed)
- Standard Recipes: Ta2O5, SiO2, Si3N4, SiOxNy, TiO2, Al2O3.
- Custom recipes: AlN, TaN, TiN
- Argon sample cleaning/pre-sputtering with low energy/etch rate
- Base Pressure ≤ 3e-8 Torr
- Accepts small pieces, up to 6” wafers accepted
- Holders available for etched/cleaved facet coating
- High uniformity over 6-inch area, high density films, low hydrogen content. Higher particle count.
- High quality, High Reflectivity DBR (>99.5%) mirrors demonstrated at UV-Vis & near-IR wavelengths
- Wide-band Anti-Reflection coatings demonstrated in near-IR (~1550nm)
Procedures
- Operating Instructions Part 1
- Operating Instructions Part 2
- Troubleshooting Guide *
- Method for accurately calibrating multi-layer optical films
- For example, for calibrating and depositing Multi-layer DBR gratings, Anti-Reflection coatings etc.
- Uses Fabry-Perot depositions to calibration "optical thickness" directly, similar to VCSEL growth calibrations.
- IBD Standard Operating procedure *
Recipes
- Sputtering Recipes: Ion Beam Deposition (Veeco NEXUS)
- Dep Rates, Refractive Indices, Stress etc. for various available dielectric coatings
Schematics/Examples