PECVD 1 (PlasmaTherm 790)
SiN deposition (PECVD #1)
- SiN Standard Recipe
- SiN Data 2014
- SiN 1000A Thickness uniformity 2014
- Low Stress Si3N4 - Variable Stress Recipes
- Low Stress Si3N4 - Variable Stress Plot
SiO2 deposition (PECVD #1)
SiOxNy deposition (PECVD #1)
- SiOxNy Standard Recipe
- SiOxNy Data 2014
- SiOxNy1000A Thickness uniformity 2014
- SiOxNy Deposition Recipes - Varying N/O Ratio
- SiOxNy Stress/Index vs. O/N Ratio
PECVD 2 (Advanced Vacuum)
SiN deposition (PECVD #2)
- Nitride2 Standard Recipe
- Nitride2 Data (Deposition Rate, Refractive Index, Stress, HF etch rate)
- Nitride2 Data 2014
- Nitride2 Thickness uniformity 2014
- Deposition Rate: ≈ 9.0 nm/min
- Refractive Index: ≈ 1.952
- Stress ≈ 484MPa
SiO2 deposition (PECVD #2)
- Oxide Standard Recipe
- Oxide Data (Deposition Rate, Refractive Index, Stress, HF etch rate )
- Oxide Data 2014
- Oxide Thickness uniformity 2014
- Deposition Rate: ≈ 28.0 nm/min
- Refractive Index: ≈ 1.472
- Stress ≈ -270MPa
LS SiN deposition (PECVD #2)
- LS Nitride2 Standard Recipe
- LS Nitride2 Data (Deposition Rate, Refractive Index, Stress, HF etch rate )
- LS Nitride2 Data 2014
- LS Nitride2 Thickness uniformity 2014
- Deposition Rate: ≈ 8.34 nm/min
- Refractive Index: ≈ 1.932
- Stress ≈ -45MPa
Amorphous-Si deposition (PECVD #2)
ICP-PECVD (Unaxis VLR)
SiN deposition (Unaxis VLR)
SiN (2% SiH4)
SiN (2% SiH4 - No-Ar)
| 50° | 100° | 250° |
|---|---|---|
SiN (2% SiH4 - No-Ar - Extra N2)
| 50° | 100° | 250° |
|---|---|---|
SiN (100% SiH4 )
| 50° | 100° | 250° |
|---|---|---|
SiO2 deposition (Unaxis VLR)
SiO2 (2% SiH4)
| 50° | 100° | 250° |
|---|---|---|
SiO2 (2% SiH4 - No Ar)
| 50° | 100° | 250° |
|---|---|---|
SiO2 (100% SiH4 HDR)
| 50° | 100° | 250° |
|---|---|---|
SiO2 (100% SiH4 LDR)
| 50° | 100° | 250° |
|---|---|---|
Amorphous Si (100%SiH4 Ar He)
| 90° | 250° |
|---|---|