Difference between revisions of "SEM 1 (JEOL IT800SHL)"
(One intermediate revision by the same user not shown) | |||
Line 15: | Line 15: | ||
===Capabilities=== |
===Capabilities=== |
||
The system has multiple detectors, detailed below. |
The system has multiple detectors, detailed below. |
||
− | Low-vacuum mode reduces sample charging by introducing N2 gas into the chamber, without sacrificing imaging quality (using a special vacuum nozzle on the electron column) |
+ | Low-vacuum mode reduces sample charging by introducing N2 gas into the chamber, without sacrificing imaging quality (using a special vacuum nozzle on the electron column). Both of these are useful for imaging low conductivity and insulating materials without the need for conductive layer coatings. |
− | The system can accept a 6” wafer, but only |
+ | The system can accept a 6” wafer, but only 140mm (X) and 80mm (Y) of the wafer is accessible with the stage movement. |
The [[SEM Sample Coater (Hummer)|'''<u>Hummer coater</u>''']] is used to deposit a thin AuPd on your samples, to reduce electrical charging of insulating samples (such as SiO2 substrates, or thick >1µm layers of SiO2 or PR). |
The [[SEM Sample Coater (Hummer)|'''<u>Hummer coater</u>''']] is used to deposit a thin AuPd on your samples, to reduce electrical charging of insulating samples (such as SiO2 substrates, or thick >1µm layers of SiO2 or PR). |
||
− | This SEM also has an Electron-Beam Lithography Nabity system. Contact |
+ | This SEM also has an Electron-Beam Lithography Nabity system. Contact Aidan Hopkins for info. |
==Detailed Specifications== |
==Detailed Specifications== |
||
Line 57: | Line 57: | ||
**Tilt: -5 to 70 degrees (depending on sample holder and offset) |
**Tilt: -5 to 70 degrees (depending on sample holder and offset) |
||
**Rotation: 360 degrees |
**Rotation: 360 degrees |
||
− | |||
− | ===Mechanical=== |
||
− | |||
− | *Max Sample Size: 6-inch wafer |
||
− | *Stage movement: |
||
− | **max: 70 x 50mm |
||
− | **4-inch wafer: limited to ~25x25mm movement area from wafer center. |
||
− | *Tilt: -5° to varies. Tilt is dependent on sample holder, stage height, and offset value. |
||
− | *Rotation: 360° |
||
− | *Specimen holders : |
||
− | **Copper and XYZ Carbon tape available |
||
− | **4-inch wafer with topside clips |
||
− | **1-inch holder for 30°/90°, 45°/90° mounting with tape or clips. |
||
==Operating Procedures== |
==Operating Procedures== |
Latest revision as of 13:01, 16 May 2024
|
About
The JEOL IT800HSL Field Emission Scanning Electron Microscope is used for imaging a variety of samples made in the facility.
Capabilities
The system has multiple detectors, detailed below. Low-vacuum mode reduces sample charging by introducing N2 gas into the chamber, without sacrificing imaging quality (using a special vacuum nozzle on the electron column). Both of these are useful for imaging low conductivity and insulating materials without the need for conductive layer coatings.
The system can accept a 6” wafer, but only 140mm (X) and 80mm (Y) of the wafer is accessible with the stage movement.
The Hummer coater is used to deposit a thin AuPd on your samples, to reduce electrical charging of insulating samples (such as SiO2 substrates, or thick >1µm layers of SiO2 or PR).
This SEM also has an Electron-Beam Lithography Nabity system. Contact Aidan Hopkins for info.
Detailed Specifications
Imaging
- Resolution:
- 0.5nm at 15kV SHL mode
- 0.7nm at 1kV
- 0.9nm at 500V
- Magnification:
- Photo magnification: x10 to x2,000,000 (128mm x 96mm)
- Display magnification: x27 to x5,480,000 (1280pix x 960pix)
- Imaging Modes:
- STD: Standard
- LDF: Large depth of focus
- BD: Beam deceleration
- Applies negative voltage to sample stage to increase effective acceleration without increasing beam acceleration (reducing charging).
- SHL: Super hybrid lens
- Detectors
- SED: Secondary electron detector (low angle)
- UHD: Ultra high resolution detector
- SBED: Scintillated back scatter electron detector
- Inserts between the objective lens and the sample
- LVBED: Low vacuum back scatter electron detector
- LVSED: Low vacuum secondary electron detector
- Accelerating Voltages:
- SEM: 0.01 to 30kV
- Probe currents
- A few pA to 500nA (30kV) 100nA (5kV)
- Specimen stage
- X: 140mm Y: 80mm
- Z: 6mm to 41mm
- Tilt: -5 to 70 degrees (depending on sample holder and offset)
- Rotation: 360 degrees
Operating Procedures
- JEOL IT800SHL Operating Procedure.
- Optimizing Astigmatism (CalTech Nanoscience Institute)
- Stig is the most common cause of blurry images.
- A Common mistake is to optimize stig on flat lines (eg. a cleaved edge or line/space features). This always leads to accidentally skewing the stig in the direction of the lines. Instead, make sure to optimize on a roundish feature, such as a piece of dust/debris.