E-Beam 4 (CHA): Difference between revisions

From UCSB Nanofab Wiki
Jump to navigation Jump to search
(added process control data, changed to heading level 1, renamed to "recipes" section)
 
(22 intermediate revisions by 8 users not shown)
Line 1: Line 1:
{{tool|{{PAGENAME}}
{{tool2|{{PAGENAME}}
|picture=e-beam4.jpg
|picture=e-beam4.jpg
|type = Vacuum Deposition
|type = Vacuum Deposition
|super= Tony Bosch
|super= Michael Barreraz
|super2= Bill Millerski
|phone=(805)839-3918x217
|phone=(805)839-7975
|location=Bay 3
|location=Bay 3
|email=bosch@ece.ucsb.edu
|email=dfreeborn@ece.ucsb.edu
|description = Multi-Wafer Evaporator
|description = Multi-Wafer Evaporator
|manufacturer = CHA Industries
|manufacturer = CHA Industries
Line 11: Line 12:
|materials =
|materials =
|toolid=10
|toolid=10
}}
}}
[[File:EBeam4 Controls.jpeg|thumb|EBeam4's controls]]
= About =
This electron-beam evaporation system is the newest of the lab for metal deposition. This system is a bell-jar type system and has the capability to do up to 10-4” wafers in a lift-off configuration and up to 24-4” wafers in a sidewall coverage configuration. Rotational motion in combination with baffling is used for lift-off and provides roughly 5% uniformity across a 4” wafer. The sidewall coverage fixturing uses full planetary motion to provide coverage over all sidewalls. The system also an 8-pocket e-beam source and an Inficon IC/5 deposition controller that allows for programming of fully automated multiple layer depositions. The metals available for deposition are Al, Ti, Au, Pt, Ni, Pd, Ag, Ge, Fe, NiCr, NiFe and Cr. This system is used for n-type ohmic contact metalization to compound semiconductors, Schottky contacts to semiconductors, bond pads, and other general metalizations. The maximum deposition thickness during a run is limited to 1.0 microns.


== About ==
= Detailed Specifications =
This electron-beam evaporation system is a bell-jar type system and has the capability to do up to 9-4” wafers in a lift-off configuration and up to 18-4” wafers in a sidewall coverage configuration. Rotational motion in combination with baffling is used for lift-off and provides roughly 5% uniformity across a 4” wafer. The sidewall coverage fixturing uses full planetary motion to provide coverage over all sidewalls. The system also an 8-pocket e-beam source and an Inficon IC/5 deposition controller that allows for programming of fully automated multiple layer depositions. The metals available for deposition are Al, Ti, Au, Pt, Ni, Pd, Ag, Ge, Fe, NiCr, NiFe and Cr. This system is used for n-type ohmic contact metalization to compound semiconductors, Schottky contacts to semiconductors, bond pads, and other general metalizations. The maximum deposition thickness during a run is limited to 1.0 microns.

== Detailed Specifications ==
*Temescal 10kV power supply
*Temescal 10kV power supply
*1-Temescal 8-pocket series 260 e-beam sources
*1-Temescal 8-pocket series 260 e-beam sources
*Cryo-pumped system with ~ 1e-7 ultimate base pressure
*Cryo-pumped system with ~ 5e-7 ultimate base pressure
*Rotation with baffle for 5% uniformity over 4” wafer
*Rotation with baffle for 5% uniformity over 4” wafer
*Automatic vacuum sequencing
*Automatic vacuum sequencing
Line 25: Line 28:
*Automatic deposition of multiple layer stacks
*Automatic deposition of multiple layer stacks
*Sample size: Pieces or up to 10-4” wafers for lift-off and 24-4” wafers for sidewall coverage
*Sample size: Pieces or up to 10-4” wafers for lift-off and 24-4” wafers for sidewall coverage
*Metals: Al, Ti, Au, Pt, Ni, Pd, Ag, Ge, Fe, NiCr, NiFe and Cr
*Metals: Al, Ti, Au, Pt, Ni, Pd, Ag, Ge, Fe, NiCr, NiFe, Cr


=Documentation=
== Documentation ==
*[[media:E-Beam-4-Operating-Procedures.pdf|Operating Procedures]]
*[https://wiki.nanofab.ucsb.edu/w/images/4/46/EB4_Operating_Instructions_6-6-24.pdf Operating Procedures]


== Recipes ==


= Materials Table =
=== Materials Table ===
For tables listing all available materials and deposition parameters, please visit the [[E-Beam_Evaporation_Recipes#Materials_Table_(E-Beam #4)|'''E-Beam Recipe Page''']].


=== Process Control Data ===
{| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center; font-size: 95%" class="collapsible collapsed wikitable"
*E-Beam 4 Ti: [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=0#gid=0 Titanium Datasheet] & [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=384597990#gid=384597990 Titanium Plots]
|-
*E-Beam 4 Au: [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=834604706#gid=834604706 Gold Datasheet] & [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=721807140#gid=721807140 Gold Plots]
! colspan=6 width=1300 height=35 bgcolor="#D0E7FF" align="center"|<div style="font-size: 150%;">Materials Table</div>
*E-Beam 4 Cr: [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=1629968393#gid=1629968393 Chromium Datasheet] & [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=701729811#gid=701729811 Chromium Plots]
|- bgcolor="#D0E7FF"
*E-Beam 4 Ni: [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=649905488#gid=649905488 Nickel Datasheet] & [https://docs.google.com/spreadsheets/d/1W7OFMAlRIcbpjm7FsbCh9ZLCCbhEp0bhtAC7UqEmJ5U/edit?gid=1525197973#gid=1525197973 Nickel Plots]
! width="45" bgcolor="#D0E7FF" align="center" | '''Material'''
! width="45" bgcolor="#D0E7FF" align="center" | '''Density, g/cm3'''
! width="45" bgcolor="#D0E7FF" align="center" | '''Z Ratio'''
! width="45" bgcolor="#D0E7FF" align="center" | '''Master tooling, %'''
! width="45" bgcolor="#D0E7FF" align="center" | '''Process Gain, A/sec/%pwr'''
! width="100" bgcolor="#D0E7FF" align="center" | '''Comments'''
|-
|Ag
| 10.50
| 0.529
|110
|10.0
|Tony, could you please check this?
|-
|Al
| 2.70
| 1.080
|165
|6.0
|
|-
|Au
|19.30
| 0.381
|120
|10.0
|
|-
|Co
|8.90
| 0.343
|150
|5.0
|
|-
|Cr
|7.20
| 0.305
|140
|10.0
|
|-
|Fe
|7.86
| 0.349
|165
|10.0
|
|-
|Ge
|5.35
| 0.516
|126
|10.0
|
|-
|Hf
|13.09
| 0.360
|150
|10.0
|
|-
|Ir
|22.40
| 0.129
|130
|10.0
|
|-
|Ni
|8.91
| 0.331
|150
|5.0
|
|-
|NiCr
|8.23
| 0.321
|140
|10.0
|density should be 8.50
|-
|NiFe
|8.70
|1.000
|100
|10.0
|
|-
|Pd
|12.038
|0.357
|112
|10.0
|
|-
|Pt
|21.40
|0.245
|130
|10.0
|
|-
|Ru
|12.362
|0.182
|100
|10.0
|should be 12.45
|-
|Ti
|4.50
|0.628
|183
|10.0
|
|-
|Zr
|6.49
|0.600
|150
|10.0
|should 6.40
|-

Latest revision as of 18:42, 28 February 2025

E-Beam 4 (CHA)
E-beam4.jpg
Location Bay 3
Tool Type Vacuum Deposition
Manufacturer CHA Industries
Model SEC-600-RAP
Description Multi-Wafer Evaporator

Primary Supervisor Michael Barreraz
(805) 893-4147
mikebarreraz@ece.ucsb.edu

Secondary Supervisor

Bill Millerski


Recipes Vacuum Deposition Recipes

SignupMonkey: Sign up for this tool


EBeam4's controls

About

This electron-beam evaporation system is a bell-jar type system and has the capability to do up to 9-4” wafers in a lift-off configuration and up to 18-4” wafers in a sidewall coverage configuration. Rotational motion in combination with baffling is used for lift-off and provides roughly 5% uniformity across a 4” wafer. The sidewall coverage fixturing uses full planetary motion to provide coverage over all sidewalls. The system also an 8-pocket e-beam source and an Inficon IC/5 deposition controller that allows for programming of fully automated multiple layer depositions. The metals available for deposition are Al, Ti, Au, Pt, Ni, Pd, Ag, Ge, Fe, NiCr, NiFe and Cr. This system is used for n-type ohmic contact metalization to compound semiconductors, Schottky contacts to semiconductors, bond pads, and other general metalizations. The maximum deposition thickness during a run is limited to 1.0 microns.

Detailed Specifications

  • Temescal 10kV power supply
  • 1-Temescal 8-pocket series 260 e-beam sources
  • Cryo-pumped system with ~ 5e-7 ultimate base pressure
  • Rotation with baffle for 5% uniformity over 4” wafer
  • Automatic vacuum sequencing
  • Temescal e-beam sweep control
  • Inficon IC/5 programmable crystal thickness monitoring system
  • Automatic deposition of multiple layer stacks
  • Sample size: Pieces or up to 10-4” wafers for lift-off and 24-4” wafers for sidewall coverage
  • Metals: Al, Ti, Au, Pt, Ni, Pd, Ag, Ge, Fe, NiCr, NiFe, Cr

Documentation

Recipes

Materials Table

For tables listing all available materials and deposition parameters, please visit the E-Beam Recipe Page.

Process Control Data