4. Example of dicing different materials: |
4. Example of dicing different materials: |
||
a) Very thick Ga<sub>2</sub>O<sub>3</sub> (660um), used double pass, blade 2. |
a) Very thick Ga<sub>2</sub>O<sub>3</sub> (660um), used double pass, blade 2.187-8C-54RU-3, expected kerf ~(210-230)um, actual kerf ~225um |
||
b) Very thick Sapphire (650um), used double pass, blade 2. |
b) Very thick Sapphire (650um), used double pass, blade 2.187-8C-54RU-3, kerf ~(210-230)um, actual kerf ~240um |
||
5. |
5. Recipe example for Dual Dicing and Kerf for Ga<sub>2</sub>O<sub>3</sub> |
||
<gallery> |
<gallery> |
||
File:Dicing Thick Ga2O3.jpg|Recipe example for Dicing Ga<sub>2</sub>O<sub>3</sub> |
File:Dicing Thick Ga2O3.jpg|Recipe example for Dicing Ga<sub>2</sub>O<sub>3</sub> |
||
File:Kerf 225um 660um thick Ga2O3 dual pass.png|Measured kerf for Ga<sub>2</sub>O<sub>3</sub> |
File:Kerf 225um 660um thick Ga2O3 dual pass.png|Measured kerf for Ga<sub>2</sub>O<sub>3</sub> (with photoresist on surface) |
||
</gallery> |
</gallery> |
||
Author: [[Biljana Stamenic]], 2026-06 |
|||
=== Partial Dicing === |
|||
For partial-depth dicing, you need to do single-pass, dicing into substrate desired depth. Setting up the recipe: |
|||
1. Setup: Enter wafer thickness (example:750um), and tape (90um) |
|||
2. Cut: (if partial cut is only 100um- looking from top of the substrate) |
|||
* Index "0" - Set this to desired index (distance between two cuts, for example: 0.105mm), index "1" - set this to "0" |
|||
* Cut: Depth - Head1 "1"- Set Depth as following [(wafer thickness + tape thickness)- partial cut depth]=(750+90)-100=740um |
|||
3. Height: Check height after every 5 cuts |
|||
4. Example of partial cut: |
|||
a) Very thick Si (750um), used single pass, blade 2.187-2C-9RU-3, cutting speed=0.5mm/sec, spindle speed=30KRPM |
|||
5. Recipe example for Partial dicing into Si substrate |
|||
[[File:Partial dicing.jpeg|thumb|150x150px|none|Cut parameters for partial-depth cut]] |
|||
Author: [[Biljana Stamenic]], 2026-06 |
|||
==[[Wafer Bonder (Logitech WBS7)]]== |
==[[Wafer Bonder (Logitech WBS7)]]== |
||
Latest revision as of 22:41, 11 June 2026
Dicing Saw Recipes (ADT 7100)
Dicing Alignment Instructions
The Process Group often has users fill out these instructions below to fully define a dicing job. This will ensure you have thought about the entire dicing process.
Note that you should design your chips with ≥250µm dicing street width, to avoid the blade cutting into your devices.
It is very helpful to also place alignment guides in the dicing streets, such as crosses at the intersections of dicing streets, such as this:
---- + ---- + ---- + ---
| | |
---- + ---- + ---- + ---
| | |
---- + ---- + ---- + ---
Recommended Dicing Parameters
This table is for our stocked Thermocarbon Resnoid blades.
-2C blades are 2mils/50µm wide, -4C blades are 4mils/100µm wide, and -8C blades are 8mils/200µm wide. Plan for ~10–30µm extra edge clearance to account for kerf, chipping, etc.
Narrower (~30-50µm) Nickel Hubbed blades are often used for even narrower dicing streets, these must be purchased by the user. KnS G1440-Q5H0 work very well, with ~30µm blade width and smaller kerf. You need to insert a shim to use these blades - please contact tool supervisor for how to use these blades.
| Material | Blade P/N | Spindle Speed
(KRPM) |
Cut Speed
(mm/s) |
Mohs Hardness
Scale* |
|---|---|---|---|---|
| Alumina, AlN | 2.187-8C-54RU-3 | 25 | 0.5-2 | 8 |
| Ceramic | 2.187-4C-30RU-3 | 18 | 0.5-2 | 7 - 9 |
| GaAs | 2.187-4C-9RU-3 | 35 | 1-5 | 4.5 |
| GaN (<550um) | 2.187-4C-30RU-3 | 35 | 0.5-3 | 9 |
| GaN (>550um) | 2.187-8C-30RU-3 | 35 | 0.5-2 | 9 |
| Glass/Fused Silica | 2.187-4C-22RU-3 | 25 | 1-5 | 5.3 - 6.5 |
| Ga2O3** | 2.187-8C-54RU-3 | 18 | 0.5 | 9 |
| InP | 2.187-4C-9RU-3 | 35 | 1-5 | 9 |
| Quartz | 2.187-4C-30RU-3 | 25 | 1-5 | 7 |
| Sapphire** | 2.187-8C-54RU-3 | 18 | 0.5-2 | 9 |
| Si | 2.187-2C-9RU-3 | 30 | 1-2 | 7 |
| Si | 2.187-4C-9RU-3 | 35 | 4-10 | 7 |
| Si on Glass | 2.187-4C-9RU-3 | 25 | 1-5 | 7 |
| SiC | 2.187-8C-30RU-3 | 25 | 0.5-2 | 9.5 |
| Ti | 2.187-8C-54RU-3 | 15 | 0.5-2 | 6 |
* If you do not see the material you want to dice listed, refer to the Mohs Hardness scale for Blade P/N.
** Refer to Dual-Pass Dicing process below.
Anatomy of a Blade
Example: 2.187-4C-9RU-3
"2.187": This is the blade Outer Diameter ("OD") in inches (55.55 mm).
"4C": Blade thickness in mils. 4 mil = 100 µm
"9": Diamond particle size in microns. Stocked resin blades have embedded diamond particles. Smaller particles create a smoother kerf, but remove less material and are thus less robust or require slower cutting speeds.
"RU-3". A blade parameter that deals with cut quality vs. robustness (lifetime) of the blade.
Calculated Blade Exposures
| Blade Diam | Flange Diam. | Blade Exposure | |
|---|---|---|---|
| 2.187" (55.55 mm) | 47 mm | 4.275 mm | |
| 2.187" (55.55 mm) | 49 mm | 3.275 mm | |
| 2.187" (55.55 mm) | 51 mm | 2.275 mm | |
| 2.187" (55.55 mm) | 52 mm | 1.775 mm | |
| 2.187" (55.55 mm) | 53 mm | 1.275 mm |
Blade Exposure Calculation

Mounting/Unmounting Samples
The UV-Release Tape dispenser is most-often used for mounting sample for dicing.
The Tape Model installed is Ultron 1042R-B. Data Sheet Here.
- Procedure for mounting sample on UV-Release Tape
- Full Release: 120 sec exposure
- Partial Release for Shipping: 9 sec exposure
Wax-Mounting to Carrier
If your final die size will be very small, eg. <3mm square or so, the chances increase that many die will be ejected into the cooling water stream once you begin the 2nd cut angle, because the total surface area of die contacting the adhesive tape becomes very low
A method to overcome this is to mount your sample with CrystalBond wax onto a Silicon carrier wafer. The wax is a much stronger adhesive. The drawback is that you must dissolve the wax to unmount your die, often resulting in a jumbled pile of small die in Acetone/Isopropanol, which can be difficult to handle/sort afterwards.
| Procedure to mount with wax onto a carrier wafer: |
|---|
|
| Procedure written by Demis D John, 2022-07-04. Please consider the publication policy. |
Dicing Tips
Harder materials will often require larger diamond particle sizes, and thicker blades will last longer if they are overheating and breaking often.
It is not uncommon to have to change a blade in the middle of cutting a wafer - the software is set up to allow this easily without aborting the programmed cuts. The "Height Check Rate" in the recipe will check the blade exposure after this many cuts, using the optical height sensor - this allows you to see how quickly the blade is wearing out (as blade exposure reduces).
Ensuring the cut water jet is hitting at ~7-8 o'clock on the blade and the water jet is being split in two will keep the blade coolest and help prevent breakage. Water sprays should be set to 0.9/0.9/0.9 by default.
For sapphire dicing (very hard material), it is common to use "double-pass dicing", where the substrate is cut at only half depth (eg. cutting only 150µm deep for a 300µm thick substrate) on the first pass, and then re-cut at the full depth. The blade will need to be changed often, so set your "Height Check Rate" to 1 or 2. This can be very time consuming. 200-300µm thick Sapphire substrates are much easier to cut than 650µm thick - often single-pass dicing is adequate for the thinner substrates.
Surface Protection
Photoresist
Users most often use sacrificial photoresists to protect the surface from accumulating dicing dust. The static-buildup of dielectric films causes the dust to adhere strongly. Ensure that the PR thickness will adequately coat all your exposed topography (eg. use a ≥2µm thick PR for protecting 1.5-2.0µm tall etched features).
- Choose a photoresist of appropriate thickness, and spin-coat it & soft-bake it according to a standard recipe.
- Perform your dicing
- Remove the die from the UV release tape (60sec UV Exposure)
- Strip the PR from each die in Acetone and ISO & N2 dry. We recommend to use ACE/ISO squirt bottles on each die individually to ensure the particles on the PR surface don’t land and stick to the chip surface.
Blue Tape
Alternatively our low-tack residue-free Blue tape can be used to protect the die surface.
Blue tape removal is easy for large die, but does require manual removal from each die, and eliminates sample exposure to solvents.
You will need to physically, gently press the blue tape onto the wafer surface, while eliminating bubbles. Do this by placing your wafer directly in front of the blue tape-roll dispenser, then attaching the tape to the table over the wafer (without touching wafer), allowing you to press the tape onto the wafer progressively from one side.
The very edges of the die may accumulate a bit more dicing dust due to the tape delaminating slightly during dicing. Plan for about 50-100µm of edge clearance on each die.
Cut Depth Accuracy
Depth accuracy (distance between bottom of the blade and sample chuck) (a) is only accurate to ~20µm or so, and (b) varies by at least~10µm across the chuck. If you need to leave a certain amount of your substrate uncut, here is a way to improve the accuracy, adds ~30-60min to your process.
Procedure to improve cut depth accuracy
To achieve an accurate depth like <30µm accuracy:
- Do Manual alignment of sample.
- You will do test cuts on the tape only (with sample mounted, but ~50mm away from sample in Y-direction (not X!)) - nominally 90µm thick but I think the plastic thickness varies between ~60-100µm as a guess. Then progressively reduce the depth to se when it hits the tape.
- Do Manual y-offset, with depth (distance between tip of blade & chuck)=0.070mm
- If you don’t see a cut, then [Cancel]
- Adjust recipe to -0.010mm depth, 0.060mm in this case
- Do Manual Y-offset again 10mm away from original cut
- Repeat until you see the blade just scraping the surface. You will see it “skips”, hitting then missing then hitting (this is the tape or chuck height variation)
- That is the depth to hit the tape.
- (To ascertain the variation, keep going down by 10µm increments until you see a “full” cut with no skips.)
- Then in your recipe, add the desired remaining amount
- eg. Depth to hit tape = 0.060mm + 0.070mm remaining uncut sample = 0.130mm cut depth.
- (30µm will also work, you’ll just see it varies by ~20µm across the cut.)
Dual-Pass Dicing
For hard (≥ 9 Mohs Hardness) wafers thicker than 400µm or so, you likely need to do double-pass dicing, where you cut 1/2 of the thickness the first pass, then cut the full wafer thickness on the second pass. Setting up the recipe:
1. Setup: Enter wafer thickness (for example:660um), and tape thickness (100um)
2. Cut:
- Index "0" - Set this to desired index (distance between two cuts), index "1" - set this to "0"
- Cut: Depth - Head1 "0"- set this to half way of actual thickness, for example: 0.300mm
- Cut: Depth - Head1 "1"- set this to cut the full thickness (wafer + cutting 20-30um into tape), for example: 0.070mm
3. Height: Check it after every cut
4. Example of dicing different materials:
a) Very thick Ga2O3 (660um), used double pass, blade 2.187-8C-54RU-3, expected kerf ~(210-230)um, actual kerf ~225um
b) Very thick Sapphire (650um), used double pass, blade 2.187-8C-54RU-3, kerf ~(210-230)um, actual kerf ~240um
5. Recipe example for Dual Dicing and Kerf for Ga2O3
-
Recipe example for Dicing Ga2O3
-
Measured kerf for Ga2O3 (with photoresist on surface)
Author: Biljana Stamenic, 2026-06
Partial Dicing
For partial-depth dicing, you need to do single-pass, dicing into substrate desired depth. Setting up the recipe:
1. Setup: Enter wafer thickness (example:750um), and tape (90um)
2. Cut: (if partial cut is only 100um- looking from top of the substrate)
- Index "0" - Set this to desired index (distance between two cuts, for example: 0.105mm), index "1" - set this to "0"
- Cut: Depth - Head1 "1"- Set Depth as following [(wafer thickness + tape thickness)- partial cut depth]=(750+90)-100=740um
3. Height: Check height after every 5 cuts
4. Example of partial cut:
a) Very thick Si (750um), used single pass, blade 2.187-2C-9RU-3, cutting speed=0.5mm/sec, spindle speed=30KRPM
5. Recipe example for Partial dicing into Si substrate

Author: Biljana Stamenic, 2026-06
Wafer Bonder (Logitech WBS7)
This tool is used for bonding samples to Silicon carrier wafers with CrystalBond wax.
- Wax Mounting Procedure, with bulk Crystalbond Wax - Recommended, works for most applications.
- Wax Mounting Procedure, with Spin-On Crystalbond - if very thin/high uniformity (≤5µm) is required.
Automated Wafer Cleaver (Loomis LSD-155LT)
This tool is used for scribe and break of your samples.
- Recommended recipes, a starting point for most applications.