__NOTOC__ |
|||
== Lift-Off Techniques == |
|||
{| class="wikitable" |
|||
|+ |
|||
!Table of Contents |
|||
|- |
|||
| |
|||
==='''<big>Photolithography Processes</big>'''=== |
|||
#'''UV Optical Lithography''' |
|||
*{{fl|Liftoff-Techniques.pdf|Description/Tutorial}} |
|||
#*[[#PositivePR |'''Stocked Lithography Chemical + Datasheets''']] |
|||
*{{fl|Bi-LayerContactprocesswithPMGI.pdf|Bi-Layer Process with PMGI Underlayer and Contact Aligner}} |
|||
#**''Lists all stocked photolith. chemicals, PRs, strippers, developers, and links to the chemical's application notes/datasheet, which detail the spin curves and nominal processes.'' |
|||
#*[[#Photolithography_Recipes |'''Photo Lithography Recipe section''']] |
|||
#**''Starting recipes (spin, bake, exposure, develop etc.) for all photolith. tools.'' |
|||
#**''Substrate/surface materials/pattern size can affect process parameters. Users may need to run Focus/Exposure Arrays/Matrix (FEA's/FEM's) with these processes to achieve high-resolution.'' |
|||
#**[[Contact Alignment Recipes|<u>Contact Aligner Recipes</u>]] |
|||
#***[[Contact Alignment Recipes#Suss Aligners .28SUSS MJB-3.29|Suss MJB Aligners]] |
|||
#***[[Contact Alignment Recipes#Contact Aligner .28SUSS MA-6.29|Suss MA6]] |
|||
#**[[Stepper Recipes|<u>Stepper Recipes</u>]] |
|||
#***[[Stepper Recipes#Stepper 1 .28GCA 6300.29|Stepper #1: GCA 6300]] (I-Line) |
|||
#***[[Stepper Recipes#Stepper 2 .28AutoStep 200.29|Stepper #2: GCA Autostep 200]] (I-Line) |
|||
#***[[Stepper Recipes#Stepper 3 .28ASML DUV.29|Stepper #3: ASML PAS 5500/300]] (DUV) |
|||
#**[[Direct-Write Lithography Recipes|<u>Direct-Write Recipes</u>]] |
|||
#***[[Direct-Write Lithography Recipes#Maskless Aligner .28Heidelberg MLA150.29|Heidelberg MLA150 (I-Line)]] |
|||
#***[[Direct-Write Lithography Recipes#E-Beam Lithography Recipes (Raith EBPG 5150+)|Raith EBPG 5150+ (Electron-Beam)]] |
|||
#***[[Lithography Recipes#FIB Lithography Recipes .28Raith Velion.29|Raith Velion FIB (Ion Beam)]] |
|||
#**[[Lithography Recipes#Automated Coat.2FDevelop System Recipes .28S-Cubed Flexi.29|Automated Coater Recipes (S-Cubed Flexi)]] |
|||
#[[Lithography Recipes#General Photolithography Techniques|'''General Photolithography Techniques''']] |
|||
#*''Techniques for improving litho. or solving common photolith. problems.'' |
|||
#*[[Photolithography - Improving Adhesion Photoresist Adhesion|HMDS Process for Improving Adhesion]] |
|||
#*[[Photolithography - Manual Edge-Bead Removal Techniques|Edge-Bead Removal Techniques]] |
|||
#*[https://www.microchemicals.com/technical_information/reflow_photoresist.pdf Photoresist reflow (MicroChem)] |
|||
#*[[Lithography Calibration - Analyzing a Focus-Exposure Matrix]] |
|||
#'''[[Lithography Recipes#Lift-Off Recipes|Lift-Off Recipes]]''' |
|||
#*''Verified Recipes for lift-off using various photolith. tools'' |
|||
#*''General educational description of this technique and it's limitations/considerations.'' |
|||
#'''E-beam Lithography''' |
|||
#*[[#E-Beam_Lithography_Recipes |E-Beam Lithography Recipes]] |
|||
#**''Has links to starting recipes. Substrates and patterns play a large role in process parameters.'' |
|||
#*[[#EBLPR |EBL Photoresist Datasheets]] |
|||
#**''Provided for reference, also showing starting recipes and usage info.'' |
|||
#'''[[Lithography Recipes#Holography Recipes|Holography]]''' |
|||
#*''For 1-D and 2-D gratings with 220nm nominal period, available on substrates up to 1 inch square.'' |
|||
#*''Recipes for silicon substrates are provided, and have been translated to other substrates by users.'' |
|||
#*''Datasheets are provided with starting recipes and usage info.'' |
|||
#'''[[Lithography Recipes#Edge-Bead Removal Techniques|Edge-Bead Removal]]''' |
|||
#*''Edge photoresist removal methods needed for clamp-based etchers'' |
|||
#*''Improves resolution for contact lithography'' |
|||
|- |
|||
| |
|||
==='''<big>Photolithography Chemicals/Materials</big>'''=== |
|||
== Chemical Datasheets == |
|||
#'''[[#Underlayers |Underlayers]]''' |
|||
{| |
|||
#*''These are used beneath resists for both adhesive purposes and to enable bi-layer lift-off profiles for use with photoresist.'' |
|||
|- valign="top" |
|||
#*''Datasheets are provided.'' |
|||
| width="400" | |
|||
#'''[[#AntiReflectionCoatings |Anti-Reflection Coatings]]''': |
|||
;Positive Photoresists |
|||
#*[[#Photolithography_Recipes |The Photoresist Recipes]] section contains recipes using these materials. |
|||
#*''Bottom Anti-Reflection Coatings (BARC) are used in the stepper systems, underneath the resists to eliminate substrate reflections that can affect resolution and repeatability for small, near resolution limited, feature sizes.'' |
|||
#*''Datasheets are provided for reference on use of the materials.'' |
|||
#'''[[#ContrastEnhancement |Contrast Enhancement Materials (CEM)]]''' |
|||
#*[[#Photolithography_Recipes |The Photoresist Recipes]] section contains recipes using these materials. |
|||
#*''Used for resolution enhancement. Not for use in contact aligners, typically used on I-Line Steppers.'' |
|||
#*''Datasheets provided with usage info.'' |
|||
#'''[[#AdhesionPromoters |Adhesion Promoters]]''' |
|||
#*''These are used to improve wetting of photoresists to your substrate.'' |
|||
#*''Datasheets are provided on use of these materials.'' |
|||
#'''[[#SpinOnDielectrics |Low-K Spin-on Dielectrics]]''' |
|||
#*[[Lithography Recipes#SpinOnDielectrics|Spin-On Dielectrics]] |
|||
#**''Datasheets for BCB, Photo-BCB, and SOG (spin-on-glass) for reference on use.'' |
|||
#*[[#Low-K_Spin-On_Dielectric_Recipes |Low-K Spin-On Dielectric Recipes]] |
|||
#**''Recipes for usage of some spin-on dielectrics.'' |
|||
#'''[[#Developers |Developers and Removers]]''' |
|||
#*''Datasheets provided for reference.'' |
|||
#*''Remover and Photoresist Strippers are used to dissolve PR during lift-off or after etching.'' |
|||
|} |
|||
==General Photolithography Techniques== |
|||
*{{fl|AXP4000pb-Datasheet.pdf|AZP4000 (AZ4110, AZ4210, AZ4330)}} |
|||
*{{fl|OCG825-Positive-Resist-Datasheet.pdf|OCG825}} |
|||
*{{fl|SPR220-Positive-Resist-Datasheet.pdf|SPR220 (SPR220-3, SPR220-7)}} |
|||
*{{fl|SPR950-Positive-Resist-Datasheet.pdf|SPR950-0.8}} |
|||
*{{fl|SPR955-Positive-Resist-Datasheet.pdf|SPR955CM (SPR955CM-0.9, SPR955CM-1.8)}} |
|||
*{{fl|UV6-Positive-Resist-Datasheet.pdf|UV6-0.7}} |
|||
*{{fl|UV210-Positive-Resist-Datasheet.pdf|UV210-0.3}} |
|||
====[[Photolithography - Improving Adhesion Photoresist Adhesion|'''HMDS Process for Improving Adhesion''']]==== |
|||
;Negative Photoresists |
|||
*''Use these procedures if you are finding poor adhesion PR lifting-off), or for chemicals (like BHF) that attack the PR adhesion interface strongly.'' |
|||
*{{fl|AZ5214-Negative-Resist-Datasheet.pdf|AZ5214}} |
|||
*{{fl|AZnLOF5510-Negative-Resist-Datasheet.pdf|AZnLOF5510}} |
|||
*{{fl|AZnLOF2020-Negative-Resist-Datasheet.pdf|AZnLOF2000 (AZnLOF2020, AZnLOF2035, AZnLOF2070)}} |
|||
*{{fl|UVN2300-Negative-Resist-Datasheet.pdf|UVN2300-0.5}} |
|||
*{{fl|xxxx.pdf|SU-8-2015 and 2075}} |
|||
====[[Photolithography - Manual Edge-Bead Removal Techniques|'''Edge-Bead Removal Techniques''']]==== |
|||
;Underlayers |
|||
*''These techniques are required for loading full-wafers into etchers that use top-side clamps, to prevent photoresist from sticking to the clamp (and potentially destroying your wafer).'' |
|||
*{{fl|PMGI-Underlayer-Datasheet.pdf|PMGI (PMGI SF-11, PMGI SF-13, PMGI SF-15)}} |
|||
*''For contact lithography, this improves the proximity of the mask plate and sample, improving resolution. For some projection systems, such as the [[Maskless Aligner (Heidelberg MLA150)|Maskless Aligner]], EBR can help with autofocus issues.'' |
|||
*{{fl|LOL2000-Underlayer-Datasheet.pdf|Shipley LOL2000}} |
|||
*''For small samples (<1cm) some techniques here to reduce edge-bead via shims.'' |
|||
====[https://www.microchemicals.com/technical_information/reflow_photoresist.pdf '''Photoresist reflow (MicroChem)''']==== |
|||
;E-beam resists |
|||
*''To create slanted sidewalls or curved surfaces.'' |
|||
*{{fl|PMMA-E-Beam-Resist-Datasheet.pdf|PMMA (PMMA, P(MMA-MAA) copolymer)}} |
|||
[[Lithography Calibration - Analyzing a Focus-Exposure Matrix|'''Lithography Calibration - Analyzing a Focus-Exposure Matrix''']] |
|||
*{{fl|maN2403-E-Beam-Resist-Datasheet.pdf|maN 2403}} |
|||
* ''On tools with autofocus (steppers & direct-write litho tools), you calibrate your litho by shooting a "Focus Exposure Matrix/Array", or "FEM/FEA", which exposes a grid with Dose varying in one axis, and Focus varying in the other.'' |
|||
;Nanoimprinting |
|||
* ''Explains how to locate the'' ''"Process Window" for your lithography.'' |
|||
==Photolithography Recipes== |
|||
*{{fl|NX1020-Nanoimprinting-Datasheet.pdf|NX1020}} |
|||
*{{fl|MRI-7020-Nanoimprinting-Datasheet.pdf|MRI-7020}} |
|||
| |
|||
;Contrast Enhancement Materials |
|||
==== Process Maturity Ranking ==== |
|||
*{{fl|CEM365iS-Contrast-Enhancement-Datasheet.pdf|CEM365iS}} |
|||
* <code>'''R6'''</code> - most mature process with regular calibrations recorded on SPC charts. |
|||
;Anti-Reflection Coatings |
|||
* … |
|||
* <code>'''R1'''</code> - least mature - "possible" but you'll have to figure it out. |
|||
* '''<code>A</code>''': Material is available for use, but no recipes are provided. |
|||
''The Key/Legend for this table's <code>R1...R6</code> values is at the <u>[[Lithography Recipes#Process Ranking Table|bottom of the table]]</u>.'' |
|||
*{{fl|XHRiC-Anti-Reflective-Coating.pdf|XHRiC}} |
|||
*{{fl|AR2-Anti-Reflective-Coating.pdf|AR2}} |
|||
*{{fl|DS-K101-Anti-Reflective-Coating.pdf|DS-K101 (developable BARC)}} |
|||
''Click the tool title to go to recipes for that tool.'' |
|||
;Adhesion Promoters |
|||
''Click the photoresist title to get the datasheet, also found in [[Lithography Recipes#Chemicals Stocked .2B Datasheets|Stocked Chemicals + Datasheets]].'' |
|||
*{{fl|xxxx.pdf|HMDS}} |
|||
{| class="wikitable" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center;" border="1" |
|||
*{{fl|xxxx.pdf|AP3000 BCB Adhesion Promoter}} |
|||
|- |
|||
*{{fl|xxxx.pdf|Omnicoat, SU-8 Adhesion Promoter}} |
|||
! colspan="8" height="45" |<div style="font-size: 150%;">Photolithography Recipes</div> |
|||
|- |
|||
;BCB and SOG |
|||
| bgcolor="#EAECF0" |<!-- INTENTIONALLY BLANK --> |
|||
! colspan="2" align="center" |'''[[Contact Alignment Recipes|<big>Contact Aligner Recipes</big>]]''' |
|||
! colspan="3" align="center" |'''[[Stepper Recipes|<big>Stepper Recipes</big>]]''' |
|||
! colspan="2" align="center" |[[Direct-Write Lithography Recipes|Direct-Write Litho. Recipes]] |
|||
|- |
|||
! width="150" bgcolor="#D0E7FF" align="center" |'''Positive Resists''' |
|||
{{LithRecipe Table}} |
|||
|Picomaster XF200 (Raith) |
|||
|- bgcolor="EEFFFF"<!-- This is the Row color: lightblue --> |
|||
| bgcolor="#D0E7FF" align="center" |[[:File:AXP4000pb-Datasheet.pdf|AZ4110]] |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)|R1}} |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|A |
|||
|A |
|||
| |
|||
|{{rl|MLA_Recipes|Positive Resist (MLA 150)}} |
|||
|{{rl|Direct-Write Lithography Recipes|Positive Resist (Raith PicoMaster)}} |
|||
|-<!-- This is a White row color --> |
|||
*{{fl|BCB-cyclotene-3000-revA.pdf|BCB, Cyclotene 3022-46(Not Photosensitive)}} |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/f/fc/AXP4000pb-Datasheet.pdf AZ4210] |
|||
*{{fl|BCB-cyclotene-4000-revA.pdf|PhotoBCB, Cyclotene 4022-40(Negative Polarity)}} |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)|R1}} |
|||
*{{fl|xxxx.pdf|Spin-on-Glass, Honeywell 512B (Not Photosensitive)}} |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
*{{fl|xxxx.pdf|Honeywell 512B Apps Data}} |
|||
|A |
|||
|A |
|||
| |
|||
|A |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/f/fc/AXP4000pb-Datasheet.pdf AZ4330RS] |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|A |
|||
|A |
|||
| |
|||
|{{rl|MLA_Recipes|Positive Resist (MLA 150)}} |
|||
|{{rl|Direct-Write Lithography Recipes|Positive Resist (Raith PicoMaster)|R1}} |
|||
|- |
|||
;Developers |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/a/a2/Az_p4620_photoresist_data_package.pdf AZ4620] |
|||
|A |
|||
|A |
|||
|A |
|||
|A |
|||
| |
|||
|A |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/8/8b/OCG825-Positive-Resist-Datasheet.pdf OCG 825-35CS] |
|||
|A |
|||
|A |
|||
|A |
|||
|A |
|||
| |
|||
|A |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/2/29/SPR955-Positive-Resist-Datasheet.pdf SPR 955 CM-0.9] |
|||
|A |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|{{rl|Stepper Recipes|Positive Resist (GCA 6300)|R3}} |
|||
|{{rl|Stepper Recipes|Positive Resist (AutoStep 200)|R5}} |
|||
| |
|||
|{{rl|MLA Recipes|Positive Resist (MLA 150)|R3}} |
|||
|{{rl|Direct-Write Lithography Recipes|Positive Resist (Raith)|R3}} |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/2/29/SPR955-Positive-Resist-Datasheet.pdf SPR 955 CM-1.8] |
|||
|A |
|||
|A |
|||
|{{rl|Stepper Recipes|Positive Resist (GCA 6300)|R3}} |
|||
|{{rl|Stepper Recipes|Positive Resist (AutoStep 200)|R3}} |
|||
| |
|||
|{{rl|MLA Recipes|Positive Resist (MLA 150)|R3}} |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/3/3f/SPR220-Positive-Resist-Datasheet.pdf SPR 220-3.0] |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|{{rl|Stepper Recipes|Positive Resist (GCA 6300)|R3}} |
|||
|{{rl|Stepper Recipes|Positive Resist (AutoStep 200)|R3}} |
|||
| |
|||
|{{rl|MLA Recipes|Positive Resist (MLA 150)}} |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/3/3f/SPR220-Positive-Resist-Datasheet.pdf SPR 220-7.0] |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|{{rl|Stepper Recipes|Positive Resist (GCA 6300)|R3}} |
|||
|{{rl|Stepper Recipes|Positive Resist (AutoStep 200)|R3}} |
|||
| |
|||
|{{rl|MLA Recipes|Positive_Resist_.28MLA150.29}} |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/b/be/3600_D%2C_D2v_Spin_Speed_Curve.pdf THMR-IP3600 HP D] |
|||
| |
|||
| |
|||
|A |
|||
|A |
|||
| |
|||
|{{rl|MLA Recipes|Positive Resist (MLA 150)}} |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/3/38/UV6-Positive-Resist-Datasheet.pdf UV6-0.8] |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|{{rl|Stepper Recipes|Positive Resist (ASML DUV)|R5}} |
|||
| |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/f/ff/UV210-Positive-Resist-Datasheet.pdf UV210-0.3] |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|{{rl|Stepper Recipes|Positive Resist (ASML DUV)}} |
|||
| |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/0/07/UV26-Positive-Resist-Datasheet.pdf UV26-2.5] |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|A |
|||
| |
|||
|- bgcolor="EEFFFF" |
|||
! bgcolor="#D0E7FF" align="center" |'''Negative Resists''' |
|||
{{LithRecipe Table}} |
|||
|Picomaster XF200 (Raith) |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/b/b0/AZ5214-Negative-Resist-Datasheet.pdf AZ5214-EIR] |
|||
|{{rl|Contact_Alignment_Recipes|Negative Resist (MJB-3)}} |
|||
|{{rl|Contact_Alignment_Recipes|Negative Resist (MA-6)}} |
|||
|{{rl|Stepper Recipes|Negative Resist (GCA 6300)}} |
|||
|{{rl|Stepper Recipes|Negative Resist (AutoStep 200)}} |
|||
| |
|||
|{{rl|MLA Recipes|Negative Resist (MLA 150)}} |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/5/5e/AZnLOF2020-Negative-Resist-Datasheet.pdf AZnLOF 2020] |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|{{rl|Contact_Alignment_Recipes|Negative Resist (MA-6)}} |
|||
|{{rl|Stepper Recipes|Negative Resist (GCA 6300)|R3}} |
|||
|{{rl|Stepper Recipes|Negative Resist (AutoStep 200)|R3}} |
|||
| |
|||
|{{rl|MLA Recipes|Negative Resist (MLA 150)|R3}} |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/5/5e/AZnLOF2020-Negative-Resist-Datasheet.pdf AZnLOF 2035] |
|||
| bgcolor="#D0E7FF" |A |
|||
| {{rl|Contact_Alignment_Recipes|Negative Resist (MA-6)|R1}} |
|||
| bgcolor="EEFFFF" |A |
|||
| bgcolor="EEFFFF" |A |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" |A |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/5/5e/AZnLOF2020-Negative-Resist-Datasheet.pdf AZnLOF 2070] |
|||
|A |
|||
|A |
|||
|{{Rl|Stepper_Recipes|Negative_Resist_.28GCA_6300.29|R2}} |
|||
|A |
|||
| |
|||
|A |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/8/82/AZnLOF5510-Negative-Resist-Datasheet.pdf AZnLOF 5510] |
|||
|A |
|||
|A |
|||
|{{rl|Stepper Recipes|Negative Resist (GCA 6300)}} |
|||
|{{rl|Stepper Recipes|Negative Resist (AutoStep 200)}} |
|||
| |
|||
|A |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/c/c9/UVN-30_-_Negative-Resist-Datasheet_-_Apr_2004.pdf UVN30-0.8] |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|{{rl|Stepper Recipes|Negative Resist (ASML DUV)|R6}} |
|||
| |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/7/78/SU-8-2015-revA.pdf SU-8 2005,2010,2015] |
|||
|A |
|||
|{{rl|Contact_Alignment_Recipes|Negative Resist (MA-6)}} |
|||
|A |
|||
|A |
|||
| |
|||
|A |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/2/2c/SU-8-2075-revA.pdf SU-8 2075] |
|||
|A |
|||
|A |
|||
|A |
|||
|A |
|||
| |
|||
|{{rl|MLA Recipes|Negative Resist (MLA 150)}} |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |NR9-[//wiki.nanotech.ucsb.edu/w/images/8/8f/NR9-1000PY-revA.pdf 1000],[//wiki.nanotech.ucsb.edu/w/images/7/71/NR9-3000PY-revA.pdf 3000],[//wiki.nanotech.ucsb.edu/w/images/f/f9/NR9-6000PY-revA.pdf 6000]PY |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|{{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|A |
|||
|{{rl|Stepper Recipes|Negative Resist (AutoStep 200)|R3}} |
|||
| |
|||
|A |
|||
|- bgcolor="EEFFFF" |
|||
! bgcolor="#D0E7FF" align="center" |'''Anti-Reflection Coatings''' |
|||
{{LithRecipe Table}} |
|||
|Picomaster XF200 (Raith) |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/3/33/XHRiC-Anti-Reflective-Coating.pdf XHRiC-11] |
|||
| |
|||
| |
|||
|A |
|||
|A |
|||
| |
|||
|A |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/0/07/DUV42P-Anti-Reflective-Coating.pdf DUV42-P] |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|{{rl|Stepper Recipes|DUV-42P-6|R3}} |
|||
| |
|||
|- |
|||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/a/af/DS-K101-304-Anti-Reflective-Coating.pdf DS-K101-304] |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|{{rl|Stepper Recipes|DS-K101-304|R3}} |
|||
| |
|||
|- |
|||
! bgcolor="#D0E7FF" align="center" | |
|||
{{LithRecipe Table}} |
|||
|Picomaster XF200 (Raith) |
|||
|} |
|||
<!-- end Litho Recipes table --> |
|||
==='''Process Ranking Table'''=== |
|||
*{{fl|AZ400K-Developer-Datasheet.pdf|AZ400K (AZ400K, AZ400K1:4)}} |
|||
Processes in the table above are ranked by their "''Process Maturity Level''" as follows: |
|||
*{{fl|AZ300MIF-Developer-Datasheet.pdf|AZ300MIF}} |
|||
{| class="wikitable" |
|||
*{{fl|xxxxxx.pdf|DS2100 BCB Developer}} |
|||
!Process Level |
|||
*{{fl|xxxxxx.pdf|SU-8 Developer}} |
|||
! colspan="11" |Description of Process Level Ranking |
|||
*{{fl|xxxxxx.pdf|101A Developer (for DUV Flood Exposed PMGI)}} |
|||
|- |
|||
|A |
|||
| colspan="11" |Process '''A'''llowed and materials available but never done |
|||
|- |
|||
|R1 |
|||
| colspan="11" |Process has been ran at least once |
|||
|- |
|||
|R2 |
|||
| colspan="11" |Process has been ran and procedure is documented |
|||
|- |
|||
|R3 |
|||
| colspan="11" |Process has been ran, procedure is documented, and data is available |
|||
|- |
|||
|R4 |
|||
| colspan="11" |Process has a documented procedure with regular (≥4x per year) data '''no''' in-Situ control available |
|||
|- |
|||
|R5 |
|||
| colspan="11" |Process has a documented procedure with regular (≥4x per year) data '''and''' in-Situ control available |
|||
|- |
|||
|R6 |
|||
| colspan="11" |Process has a documented procedure and control charts/limits available. Controlled process. |
|||
|} |
|||
==Lift-Off Recipes== |
|||
*{{fl|Liftoff-Techniques.pdf|Lift-Off Description/Tutorial}} |
|||
;Photoresist Removers |
|||
**How it works, process limits and considerations for designing your process |
|||
*[[Lift-Off with I-Line Imaging Resist + LOL2000 Underlayer|I-Line Lift-Off: Bi-Layer Process with LOL2000 Underlayer]] |
|||
**''Single Expose/Develop process for simplicity'' |
|||
**''Up to ~130nm metal thickness & ≥500nm-1000nm gap between metal.'' |
|||
**''Can use any I-Line litho tool (GCA Stepper, Contact aligner, MLA)'' |
|||
*{{fl|Bi-LayerContactprocesswithPMGI.pdf|I-Line Lift-Off: Bi-Layer Process with PMGI Underlayer and Contact Aligner}} |
|||
**''Multiple processes for Metal thicknesses ~800nm to ~2.5µm'' |
|||
**''Uses multiple [[DUV Flood Expose|DUV Flood exposure]]/develop cycles to create undercut.'' |
|||
**''Can be transferred to other I-Line litho tools (Stepper, MLA etc.)'' |
|||
**<u>Updates to the process:</u> |
|||
***You now only need to do 5min [[DUV Flood Expose|DUV Flood exposure]] each cycle (lamp optics were replaced) |
|||
***Use the new "101B" developer, which replaced the "SAL/101A" developers but works the same. |
|||
**There are 2 [[DUV Flood Expose|DUV Flood exposure]] + 101B develop steps because the first exposure doesn't expose the entire PMGI - because PMGI absorbs DUV very strongly so the light doesn't penetrate to the bottom of thick PMGI on the first exposure. |
|||
*[[Lift-off DUV Lift-off: LOR Underlayers|DUV Lift-off: LOR Underlayers]] |
|||
**1µm (LOR-10A) & 2-3µm (LOR-30C) versions |
|||
**Written for ASML DUV Stepper, but should work for any Imaging (top) resist, incl. I-Line tools. |
|||
*[[Lift-Off with DUV Imaging + PMGI Underlayer|DUV Lift-Off: UV6 Imaging Resist + PMGI Underlayer]] |
|||
**''Single-expose/develop process'' |
|||
**''Up to ~65nm metal thickness & ~350nm gap between metal'' |
|||
**''Use thicker PMGI for thicker metals'' |
|||
==[[E-Beam Lithography System (JEOL JBX-6300FS)|E-Beam Lithography Recipes (JEOL JBX-6300FS)]]== |
|||
*{{fl|1165-Resist-Remover.pdf|1165}} |
|||
*{{fl|AZ300T-Resist-Remover.pdf|AZ300T}} |
|||
*{{fl|PRX-127-Resist-Remover.pdf|PRX-127}} |
|||
*{{fl|xxxx.pdf|Remover PG, SU-8 stripper}} |
|||
*Under Development. |
|||
|} |
|||
==[[Focused Ion-Beam Lithography (Raith Velion)|FIB Lithography Recipes (Raith Velion)]]== |
|||
== Recipes == |
|||
''To Be Added'' |
|||
==[[Automated Coat/Develop System (S-Cubed Flexi)|Automated Coat/Develop System Recipes (S-Cubed Flexi)]]== |
|||
{{Recipe Table Explanation}} |
|||
Recipes pre-loaded on the S-Cubed Flexi automated coat/bake/develop system. Only staff may write new recipes, contact the tool supervisor for more info. |
|||
===Available Variations=== |
|||
{| border="1" style="border: 1px solid #D0E7FF; background-color:#ffffff; text-align:center;" class="wikitable" |
|||
|- bgcolor="#D0E7FF" |
|||
*We have different recipes with varyious UV6 spin speeds - the same spin speed optionss as found on our manual Headway spinners. This allows for PR thickness control. See the linked UV6 datasheets below for thickness vs. rpm spin curves. '''Note''' that exact spin RPM may be slightly different between Headway spinners (on benches) vs. S-Cubed. |
|||
! height="45" colspan="9" | <div style="font-size: 150%;">Lithography Recipes</div> |
|||
*DSK is recommended to be spun at 1.5krpm (~40nm) for best anti-reflection properties. 5krpm (~20nm) recipes are also provided for historical/legacy processes. |
|||
|- bgcolor="#D0E7FF" |
|||
*DSK can be baked at either 220C to act as a Dry-etchable BARC (similar to DUV-42P), or at lower temps as a developable BARC (no dry etch required). |
|||
| <!-- INTENTIONALLY LEFT BLANK --> <br> |
|||
*"Chain" recipes (with DSK+UV6 spin/cured in succession) are only available for DSK Baked at 185C & 220C, and all UV6 Spin-speed variations. For the other DSK temps you can use the single-PR "Routes". |
|||
! bgcolor="#D0E7FF" align="center" colspan="2" | '''[[Contact Alignment Recipes|Contact Aligners]]''' |
|||
*Developer recipes are now available for 300 MiF Developer. '''Note''' that developer is flowing continuously during the develop, so <u>develop times are shorter by ~50%</u> compared to beaker developing. |
|||
! bgcolor="#D0E7FF" align="center" colspan="3" | '''[[Stepper Recipes|Steppers]]''' |
|||
**'''DO NOT EDIT developer recipes''', they can damage the tool when programmed incorrectly! |
|||
! width="30" bgcolor="#D0E7FF" align="center" | '''[[Flood Exposure Recipes|Flood Expose]]''' |
|||
*"Chain" recipes for 135*C Post-exposure Bake + Develop have been made. |
|||
! bgcolor="#D0E7FF" align="center" colspan="2" | '''[[E-Beam Lithography Recipes|E-Beam Lithography]]''' |
|||
*ASML cassettes can be placed directly on this tool for spin / exposure / develop process. Please make sure all cassettes are kept back at their respecting tools when done. |
|||
===Recipes Table (S-Cubed Flexi)=== |
|||
{| class="wikitable" |
|||
|+''Ask [[Tony Bosch|Staff]] if you need a new recipe.'' |
|||
|'''''<u>Coating Material</u>''''' |
|||
|'''''<u>Route/Chain</u>''''' |
|||
|'''''<u>Name</u>''': (User: "UCSB Users")'' |
|||
|'''''<u>Spin Speed (krpm)</u>''''' |
|||
|'''''<u>Bake Temp</u>''''' |
|||
|'''''<u>Notes</u>''''' |
|||
|- |
|- |
||
! colspan="6" | |
|||
! width="150" bgcolor="#D0E7FF" align="center" | '''Positive Resists''' |
|||
|- |
|||
{{LithRecipe Table}} |
|||
! colspan="6" |BEFORE LITHOGRAPHY (PR Coat and Bake) |
|||
|- |
|||
! colspan="6" | |
|||
|- |
|||
|'''''Hotplate Set''''' |
|||
|Route |
|||
| colspan="4" |To pre-set the DSK Hotplate temp (HP4). |
|||
Note: Only HP4 can be changed. |
|||
HP1-HP3 remains fixed at: HP1=135°C, HP2=170°C & HP3=170°C |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | AZ4110 |
|||
| |
|||
| bgcolor="EEFFFF" | {{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|HP4-SET-'''220C''' |
|||
| bgcolor="EEFFFF" | {{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
| |
|||
| bgcolor="EEFFFF" | A |
|||
|220°C |
|||
| bgcolor="EEFFFF" | A |
|||
|Will over shoot +-2°C when done. |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | AZ4210 |
|||
| |
|||
| {{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|HP4-SET-'''210C''' |
|||
| {{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
| |
| |
||
|210°C |
|||
| A |
|||
| |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | AZ4330RS |
|||
| |
|||
| bgcolor="EEFFFF" | {{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|HP4-SET-'''200C''' |
|||
| bgcolor="EEFFFF" | {{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
| |
|||
| bgcolor="EEFFFF" | A |
|||
|200°C |
|||
| bgcolor="EEFFFF" | A |
|||
| |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | OCG 825-35CS |
|||
| |
| |
||
|HP4-SET-'''185C''' |
|||
| A |
|||
| |
| |
||
|185°C |
|||
| A |
|||
| |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
|- |
|- |
||
| colspan="6" |'''<u>''Bottom Anti-Reflection Coating (BARC), DSK101 only:''</u>''' |
|||
| bgcolor="#D0E7FF" align="center" | SPR 950-0.8 |
|||
| bgcolor="EEFFFF" | A |
|||
| bgcolor="EEFFFF" | A |
|||
| bgcolor="EEFFFF" | A |
|||
| bgcolor="EEFFFF" | A |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
|- |
|- |
||
|'''''[https://wiki.nanotech.ucsb.edu/wiki/images/a/af/DS-K101-304-Anti-Reflective-Coating.pdf DS-K101]''''' |
|||
| bgcolor="#D0E7FF" align="center" | SPR 955 CM-0.9 |
|||
|Route |
|||
| A |
|||
| colspan="4" |''DSK101 Develop Rate depends on Bake temp - you can use this to control undercut.'' ''See: [[DS-K101-304 Bake Temp. versus Develop Rate|DSK Bake vs. Dev rate]]'' |
|||
| A |
|||
DSK101 spun at 1.5K is equivalent to DUV-42P. See: [[Stepper Recipes#Anti-Reflective Coatings]] |
|||
| {{rl|Stepper Recipes|Positive Resist (GCA 6300)}} |
|||
| {{rl|Stepper Recipes|Positive Resist (AutoStep 200)}} |
|||
'''185°C bake''' allows the DSK to dissolve during develop, and allows for undercut (may lift-off small features) |
|||
| <br> |
|||
| <br> |
|||
'''220°C bake''' allows DSK to be used as dry-etchable BARC (equiv. to DUV42P), requiring O2 etch to remove. Better for small features. See [[Stepper Recipes#Anti-Reflective Coatings|here for relevant processing info from DUV42P]]. |
|||
| <br> |
|||
| <br> |
|||
'''Note: All PR coat recipes have EBR backside clean steps included in the recipe.''' |
|||
|- |
|- |
||
|''<u>1.5krpm</u> recipes'' |
|||
| bgcolor="#D0E7FF" align="center" | SPR 955 CM-1.8 |
|||
| |
|||
| bgcolor="EEFFFF" | A |
|||
|COAT-DSK101-304[1.5K]-'''185C''' |
|||
| bgcolor="EEFFFF" | A |
|||
|1.5krpm |
|||
| bgcolor="EEFFFF" | {{rl|Stepper Recipes|Positive Resist (GCA 6300)}} |
|||
|185°C |
|||
| bgcolor="EEFFFF" | {{rl|Stepper Recipes|Positive Resist (AutoStep 200)}} |
|||
|Requires: HP4=185°C, |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | SPR 220-3.0 |
|||
| |
|||
| {{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|COAT-DSK101-304[1.5K]-'''200C''' |
|||
| {{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|1.5krpm |
|||
| {{rl|Stepper Recipes|Positive Resist (GCA 6300)}} |
|||
|200°C |
|||
| {{rl|Stepper Recipes|Positive Resist (AutoStep 200)}} |
|||
|Requires: HP4=200°C |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | SPR 220-7.0 |
|||
| |
|||
| bgcolor="EEFFFF" | {{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|COAT-DSK101-304[1.5K]-'''210C''' |
|||
| bgcolor="EEFFFF" | {{rl|Contact_Alignment_Recipes|Positive Resist (MA-6)}} |
|||
|1.5krpm |
|||
| bgcolor="EEFFFF" | {{rl|Stepper Recipes|Positive Resist (GCA 6300)}} |
|||
|210°C |
|||
| bgcolor="EEFFFF" | {{rl|Stepper Recipes|Positive Resist (AutoStep 200)}} |
|||
|Requires: HP4=210°C |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | THMR-IP3600 HP D |
|||
| |
|||
| <br> |
|||
|COAT-DSK101-304[1.5K]-'''220C''' |
|||
| <br> |
|||
|1.5krpm |
|||
| A |
|||
|220°C |
|||
| A |
|||
|Requires: HP4=220°C, |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
|- |
|- |
||
| colspan="6" | |
|||
| bgcolor="#D0E7FF" align="center" | UV6-0.7 |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | {{rl|Stepper Recipes|Positive Resist (ASML DUV)}} |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
|- |
|- |
||
|''<u>5krpm</u> recipes'' |
|||
| bgcolor="#D0E7FF" align="center" | UV210-0.3 |
|||
| |
|||
| bgcolor="EEFFFF" | |
|||
|COAT-DSK101-304[5K]-'''185C''' |
|||
| bgcolor="EEFFFF" | |
|||
|5.0krpm |
|||
| bgcolor="EEFFFF" | |
|||
|185°C |
|||
| bgcolor="EEFFFF" | |
|||
|Requires: HP4=185°C, |
|||
| bgcolor="EEFFFF" | {{rl|Stepper Recipes|Positive Resist (ASML DUV)}} |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
| bgcolor="EEFFFF" | |
|||
|- |
|- |
||
| |
|||
! bgcolor="#D0E7FF" align="center" | '''Negative Resists''' |
|||
| |
|||
{{LithRecipe Table}} |
|||
|COAT-DSK101-304[5K]-'''200C''' |
|||
|5.0krpm |
|||
|- bgcolor="EEFFFF" |
|||
|200°C |
|||
| bgcolor="#D0E7FF" align="center" | AZ5214-EIR |
|||
|Requires: HP4=200°C |
|||
| {{rl|Contact_Alignment_Recipes|Negative Resist (MJB-3)}} |
|||
| {{rl|Contact_Alignment_Recipes|Negative Resist (MA-6)}} |
|||
| {{rl|Stepper Recipes|Negative Resist (GCA 6300)}} |
|||
| {{rl|Stepper Recipes|Negative Resist (AutoStep 200)}} |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | AZnLOF 2020 |
|||
| |
|||
| {{rl|Contact_Alignment_Recipes|Positive Resist (MJB-3)}} |
|||
|COAT-DSK101-304[5K]-'''210C''' |
|||
| {{rl|Contact_Alignment_Recipes|Negative Resist (MA-6)}} |
|||
|5.0krpm |
|||
| {{rl|Stepper Recipes|Negative Resist (GCA 6300)}} |
|||
|210°C |
|||
| {{rl|Stepper Recipes|Negative Resist (AutoStep 200)}} |
|||
|Requires: HP4=210°C |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | AZnLOF 2035 |
|||
| A |
|||
| A |
|||
| A |
|||
| A |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | AZnLOF 2070 |
|||
| |
| |
||
|COAT-DSK101-304[5K]-'''220C''' |
|||
| A |
|||
|5.0krpm |
|||
| A |
|||
|220°C |
|||
| A |
|||
|Requires: HP4=220°C, |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | AZnLOF 5510 |
|||
| A |
|||
| A |
|||
| {{rl|Stepper Recipes|Negative Resist (GCA 6300)}} |
|||
| {{rl|Stepper Recipes|Negative Resist (AutoStep 200)}} |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| colspan="6" |'''<u>''Imaging resist (UV6) only:''</u>''' |
|||
| bgcolor="#D0E7FF" align="center" | UVN2300-0.5 |
|||
| |
|- |
||
|'''''[https://wiki.nanofab.ucsb.edu/w/images/3/38/UV6-Positive-Resist-Datasheet.pdf UV6-0.8]''''' |
|||
| |
|||
|Route |
|||
| |
|||
|COAT-UV6['''2K''']-135C |
|||
| |
|||
|2.0krpm |
|||
| {{rl|Stepper Recipes|Negative Resist (ASML DUV)}} |
|||
|135°C |
|||
| |
|||
|Requires: HP1=135°C |
|||
| |
|||
| |
|- |
||
|''varying spin speed'' |
|||
|- bgcolor="EEFFFF" |
|||
| |
|||
| bgcolor="#D0E7FF" align="center" | SU-8 2015 |
|||
|COAT-UV6['''2.5K''']-135C |
|||
| A |
|||
|2.5krpm |
|||
| A |
|||
|135°C |
|||
| A |
|||
|Requires: HP1=135°C |
|||
| A |
|||
| |
|- |
||
| |
| |
||
| |
| |
||
|COAT-UV6['''3K''']-135C |
|||
| |
|||
|3.0krpm |
|||
|135°C |
|||
|Requires: HP1=135°C |
|||
|- |
|||
| |
|||
| |
|||
|COAT-UV6['''3.5K''']-135C |
|||
|3.5krpm |
|||
|135°C |
|||
|Requires: HP1=135°C |
|||
|- |
|||
| |
|||
| |
|||
|COAT-UV6['''4K''']-135C |
|||
|4.0krpm |
|||
|135°C |
|||
|Requires: HP1=135°C |
|||
|- |
|||
| |
|||
| |
|||
|COAT-UV6['''5K''']-135C |
|||
|5.0krpm |
|||
|135°C |
|||
|Requires: HP1=135°C |
|||
|- |
|||
| |
|||
| |
|||
|COAT-UV6['''6K''']-135C |
|||
|6.0krpm |
|||
|135°C |
|||
|Requires: HP1=135°C |
|||
|- |
|||
| colspan="6" |'''''<u>UV6 Coat with Developable BARC underlayer:</u>''''' |
|||
|- |
|- |
||
|'''''DS-K101 @ 185°C''''' |
|||
! bgcolor="#D0E7FF" align="center" | '''Underlayers''' |
|||
'''''+ UV6''''' |
|||
{{LithRecipe Table}} |
|||
|Chain |
|||
|COAT-DSK101['''1.5K''']-185C-UV6['''2K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 2.0krpm |
|||
|DSK: 185°C |
|||
UV6: 135°C |
|||
|Requires: |
|||
– HP4=185°C |
|||
– HP1=135°C |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | PMGI SF-11 |
|||
Plan for ~10-15 min per wafer. |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
| <br> |
|||
|- |
|- |
||
|''<u>1.5krpm DSK</u> recipes with'' |
|||
| bgcolor="#D0E7FF" align="center" | PMGI SF-15 |
|||
''UV6- various spin speeds'' |
|||
| |
|||
| |
| |
||
|COAT-DSK101['''1.5K''']-185C-UV6['''2.5K''']-135C |
|||
| |
|||
|DSK: 1.5krpm |
|||
| |
|||
UV6: 2.5krpm |
|||
| |
|||
|same as above |
|||
| |
|||
|same as above |
|||
| |
|||
| |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | LOL 2000 |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | XHRIC-11 (i-lineBARC) |
|||
| |
| |
||
|COAT-DSK101['''1.5K''']-185C-UV6['''3K''']-135C |
|||
| |
|||
|DSK: 1.5krpm |
|||
| |
|||
UV6: 3.0krpm |
|||
| |
|||
|same as above |
|||
| |
|||
|same as above |
|||
| |
|||
| |
|||
| |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | AR-2 (DUV BARC) |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| |
|||
| bgcolor="#D0E7FF" align="center" | DS-K101 (DUV BARC) |
|||
| |
| |
||
|COAT-DSK101['''1.5K''']-185C-UV6['''3.5K''']-135C |
|||
| |
|||
|DSK: 1.5krpm |
|||
| |
|||
UV6: 3.5krpm |
|||
| |
|||
|same as above |
|||
| |
|||
|same as above |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| |
|||
! bgcolor="#D0E7FF" align="center" | '''Overlayers''' |
|||
| |
|||
{{LithRecipe Table}} |
|||
|COAT-DSK101['''1.5K''']-185C-UV6['''4K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 4.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''1.5K''']-185C-UV6['''5K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 5.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''1.5K''']-185C-UV6['''6K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 6.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| colspan="6" | |
|||
|- |
|||
|''<u>5krpm DSK</u> recipes with'' |
|||
''UV6- various spin speeds'' |
|||
| |
|||
|COAT-DSK101['''5K''']-185C-UV6['''2K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 2.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-185C-UV6['''2.5K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 2.5krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-185C-UV6['''3K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 3.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-185C-UV6['''3.5K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 3.5krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-185C-UV6['''4K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 4.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-185C-UV6['''5K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 5.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-185C-UV6['''6K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 6.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| colspan="6" |'''<u>''UV6 Coat with Dry-Etchable BARC underlayer:''</u>''' |
|||
|- |
|||
|'''''DS-K101 @ 220°C''''' |
|||
'''''+ UV6''''' |
|||
|Chain |
|||
|COAT-DSK101['''1.5K''']-220C-UV6['''2K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 2.0krpm |
|||
|DSK: 220°C |
|||
UV6: 135°C |
|||
|Requires: |
|||
– HP4=220°C |
|||
– HP1=135°C |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | CEM-365 IS |
|||
Plan for ~10-15 min per wafer. |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
|''<u>1.5krpm DSK</u> recipes with'' |
|||
! bgcolor="#D0E7FF" align="center" | '''E-Beam Resists''' |
|||
''UV6- various spin speeds'' |
|||
{{LithRecipe Table}} |
|||
| |
|||
|COAT-DSK101['''1.5K''']-220C-UV6['''2.5K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 2.5krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''1.5K''']-220C-UV6['''3K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 3.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''1.5K''']-220C-UV6['''3.5K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 3.5krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''1.5K''']-220C-UV6['''4K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 4.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''1.5K''']-220C-UV6['''5K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 5.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''1.5K''']-220C-UV6['''6K''']-135C |
|||
|DSK: 1.5krpm |
|||
UV6: 6.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| colspan="6" | |
|||
|- |
|||
|''<u>5krpm DSK</u> recipes with'' |
|||
''UV6- various spin speeds'' |
|||
| |
|||
|COAT-DSK101['''5K''']-220C-UV6['''2K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 2.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-220C-UV6['''2.5K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 2.5krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-220C-UV6['''3K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 3.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-220C-UV6['''3.5K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 3.5krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-220C-UV6['''4K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 4.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-220C-UV6['''5K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 5.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|COAT-DSK101['''5K''']-220C-UV6['''6K''']-135C |
|||
|DSK: 5krpm |
|||
UV6: 6.0krpm |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
! colspan="6" | |
|||
|- |
|||
! colspan="6" |AFTER LITHOGRAPHY (PEB and Developing) |
|||
|- |
|||
! colspan="6" | |
|||
|- |
|||
|'''''PEB Wafer Bake''''' |
|||
|Route |
|||
| colspan="4" |To bake wafer with UV6 after exposure (PEB) for 90sec and cool for 15sec |
|||
|- |
|||
| |
|||
| |
|||
|BAKE-135C-90S |
|||
| |
|||
|135°C |
|||
|Requires: HP1=135°C |
|||
|- |
|||
| colspan="6" | |
|||
|- |
|||
|'''''PEB and Developing''''' |
|||
|Route |
|||
| colspan="4" |To post-exposure bake wafer after exposure (std. PEB for UV6) 90sec, cool 15sec, develop using AZ300MIF and water rinse 60sec |
|||
'''WARNING''': DO NOT USE ANY OTHER DEVELOPER RECIPES OTHER THAN THE ONES LISTED HERE (or equipment damage may occur!) |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | ?????? |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
|''Varying developer time'' |
|||
| bgcolor="#D0E7FF" align="center" | ??????? |
|||
| |
| |
||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''10S''' |
|||
| |
|||
|Developer chuck: 300rpm |
|||
| |
|||
|135°C |
|||
| |
|||
|Requires: HP1=135°C |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| |
|||
! bgcolor="#D0E7FF" align="center" | '''Nanoimprint Resists''' |
|||
| |
|||
{{LithRecipe Table}} |
|||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''15S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''20S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''25S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''30S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''35S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''40S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-'''45S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| colspan="6" | |
|||
|- |
|||
|'''Developing''' |
|||
|Route |
|||
| colspan="4" |To only develop wafer using AZ300MIF and water rinse 60sec (No PEB) |
|||
'''WARNING''': DO NOT USE ANY OTHER DEVELOPER RECIPES OTHER THAN THE ONES LISTED HERE (or equipment damage may occur!) |
|||
|- bgcolor="EEFFFF" |
|||
| bgcolor="#D0E7FF" align="center" | MR-I 7020 |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
|''Varying developer time'' |
|||
| bgcolor="#D0E7FF" align="center" | Nanonex NX-1020 |
|||
| |
| |
||
|DEV[MIF300]-SPIN[300RPM]-'''10S''' |
|||
| |
|||
|Developer chuck: 300rpm |
|||
| |
|||
|135°C |
|||
| |
|||
|Requires: HP1=135°C |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|- |
|- |
||
| |
|||
! bgcolor="#D0E7FF" align="center" | |
|||
| |
|||
{{LithRecipe Table}} |
|||
|DEV[MIF300]-SPIN[300RPM]-'''15S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|DEV[MIF300]-SPIN[300RPM]-'''20S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|DEV[MIF300]-SPIN[300RPM]-'''25S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|DEV[MIF300]-SPIN[300RPM]-'''30S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|DEV[MIF300]-SPIN[300RPM]-'''35S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|DEV[MIF300]-SPIN[300RPM]-'''40S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
|DEV[MIF300]-SPIN[300RPM]-'''45S''' |
|||
|same as above |
|||
|same as above |
|||
|same as above |
|||
|- |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
| |
|||
|} |
|||
==[[Holographic Lith/PL Setup (Custom)|Holography Recipes]]== |
|||
''The Holography recipes here use the BARC layer XHRiC-11 & the high-res. I-Line photoresist THMR-IP3600HP-D.'' |
|||
*{{fl|Holography_Process_for_1D-lines_and_2D-dots_%28ARC-11_%26_THMR-IP3600HP-D%29-updated-4-8-2021.pdf|Standard Holography Process - on SiO2 on Si}} |
|||
*{{fl|Holography-Process-Variation-revA.pdf|Holography Process Variations - Set-up Angle - Etching into SiO2 and Si}} |
|||
*{{fl|05-SiO2_Nano-structure_Etch.pdf|Etch SiO2 Nano-structure - Changing Side-wall Angle - Etching into Si with a different line-width}} |
|||
*{{fl|30-Redicing_Nanowire_Diameter_by_Thermal_Oxidation_and_Vapored_HF_Etch.pdf|Reduce SiO2 Nanowire Diameter - Thermal Oxidation - Vapor HF Etching}} |
|||
==Low-K Spin-On Dielectric Recipes== |
|||
*{{fl|Lithography-BCB-photo-lowk-dielectric-spinon-4024-40-revA.docx|Photo BCB (4024-40)}} |
|||
*{{fl|BCB-cyclotene-3000-revA.pdf|Standard BCB (3022-46)}} |
|||
*{{fl|512B-Application-Data-Bake-revA.pdf|SOG (T512B)}} |
|||
==Chemicals Stocked + Datasheets== |
|||
''The following is a list of the lithography chemicals we stock in the lab, with links to the datasheets for each. The datasheets will often have important processing info such as spin-speed vs. thickness curves, typical process parameters, bake temps/times etc.'' |
|||
Item numbers in (parentheses) indicate the specific stocked formulation, when the datasheet shows multiple formulations. |
|||
{| |
|||
|- valign="top" |
|||
| width="400" | |
|||
;<div id="PositivePR"><big>Positive Photoresists</big></div> |
|||
'''''i-line and broadband''''' |
|||
*{{fl|AXP4000pb-Datasheet.pdf|AZP4000 (AZ4110, AZ4210, AZ4330)}} |
|||
*{{Fl|Az_p4620_photoresist_data_package.pdf|AZ P4620}} |
|||
*{{fl|OCG825-Positive-Resist-Datasheet.pdf|OCG825}} |
|||
*{{fl|SPR220-Positive-Resist-Datasheet.pdf|SPR220 (SPR220-3, SPR220-7)}} |
|||
*{{fl|SPR955-Positive-Resist-Datasheet.pdf|SPR955CM (SPR955CM-0.9, SPR955CM-1.8)}} |
|||
*THMR-3600HP (Thin I-Line & Holography) |
|||
**{{fl|THMR_iP_3500_iP3600.pdf|Evaluation Results: THMR-3600HP}} |
|||
**{{fl|3600_D,_D2v_Spin_Speed_Curve.pdf|Spin Curves for THMR-3600HP}} |
|||
**{{fl|THMR-iP3600_HP_D_20140801_(B)_GHS_US.pdf|Safety Datasheet for THMR-3600HP}} |
|||
'''''DUV-248nm''''' |
|||
*{{fl|UV210-Positive-Resist-Datasheet.pdf|UV210-0.3}} |
|||
*{{fl|UV6-Positive-Resist-Datasheet.pdf|UV6-0.8}} |
|||
*{{fl|UV26-Positive-Resist-Datasheet.pdf|UV26-2.5}} |
|||
;<div id="NegativePR"><big>Negative Photoresists</big></div> |
|||
'''''i-line and broadband''''' |
|||
*{{fl|AZ5214-Negative-Resist-Datasheet.pdf|AZ5214}} |
|||
*{{fl|AZnLOF5510-Negative-Resist-Datasheet.pdf|AZnLOF5510}} |
|||
*{{fl|AZnLOF2020-Negative-Resist-Datasheet.pdf|AZnLOF2000 (AZnLOF2020, AZnLOF2035, AZnLOF2070)}} |
|||
*{{fl|NR9-1000PY-revA.pdf|Futurrex NR9-1000PY(use AZ300MIF dev)}} |
|||
*{{fl|NR9-3000PY-revA.pdf|Futurrex NR9-3000PY(use AZ300MIF dev)}} |
|||
*{{fl|NR9-6000PY-revA.pdf|Futurrex NR9-6000PY(use AZ300MIF dev)}} |
|||
*{{fl|SU-8-2015-revA.pdf|SU-8-2005,2010, 2015}} |
|||
*{{fl|SU-8-2075-revA.pdf|SU-8-2075}} |
|||
'''''DUV-248nm''''' |
|||
*{{fl|UVN-30_-_Negative-Resist-Datasheet_-_Apr_2004.pdf|UVN-30-0.8}} |
|||
;<div id="Underlayers"><big>Underlayers</big></div> |
|||
See [[Lithography Recipes#Lift-Off Recipes|this page for lift-off recipes.]] |
|||
*{{fl|PMGI-Underlayer-Datasheet.pdf|PMGI (PMGI SF3,5,8,11,15)}} |
|||
*LOR ([[Media:Lor pds.pdf|LOR 10A & LOR 30C]]) |
|||
*{{fl|LOL2000-Underlayer-Datasheet.pdf|Shipley LOL2000}} |
|||
;<div id="EBLPR"><big>E-beam resists</big></div> |
|||
*{{fl|PMMA-E-Beam-Resist-Datasheet.pdf|PMMA (PMMA, P(MMA-MAA) copolymer)}} |
|||
*{{fl|maN2403-E-Beam-Resist-Datasheet.pdf|maN 2403}} |
|||
;<div id="NanoImprinting"><big>Nanoimprinting</big></div> |
|||
*{{fl|NX1020-Nanoimprinting-Datasheet.pdf|NX1020}} |
|||
*{{fl|MRI-7020-Nanoimprinting-Datasheet.pdf|MRI-7020}} |
|||
*{{fl|Mr-UVCur21.pdf|MR-UVCur21}} |
|||
*{{fl|OrmoStamp-NIL-Lithography-UV-Soft-RevA.pdf|Ormostamp}} |
|||
| |
|||
;<div id="ContrastEnhancement"><big>Contrast Enhancement Materials</big></div> |
|||
*{{fl|CEM365iS-Contrast-Enhancement-Datasheet.pdf|CEM365iS}} |
|||
;<div id="AntiReflectionCoatings"><big>Anti-Reflection Coatings</big></div> |
|||
*{{fl|XHRiC-Anti-Reflective-Coating.pdf|XHRiC-11 (i-line)}} |
|||
*{{fl|DUV42P-Anti-Reflective-Coating.pdf|DUV42P-6 (DUV) (For AR2 replacement)}} |
|||
*{{fl|DS-K101-304-Anti-Reflective-Coating.pdf|DS-K101-304 (DUV developable BARC)}} |
|||
;<div id="AdhesionPromoters"><big>Adhesion Promoters</big></div> |
|||
*HMDS |
|||
*AP3000 BCB Adhesion Promoter |
|||
*{{fl|OMNICOAT-revA.pdf|Omnicoat, SU-8 Adhesion Promoter}} |
|||
*{{fl|OrmoPrime-NIL-Adhesion-RevA.pdf|Ormoprime08-Ormostsmp Adhesion Promoter}} |
|||
;<div id="SpinOnDielectrics"><big>Spin-On Dielectrics</big></div> |
|||
''Low-K Spin-On Dielectrics such as Benzocyclobutane and Spin-on Glass'' |
|||
*{{fl|BCB-cyclotene-3000-revA.pdf|BCB, Cyclotene 3022-46(Not Photosensitive)}} |
|||
*{{fl|BCB-cyclotene-4000-revA.pdf|PhotoBCB, Cyclotene 4024-40(Negative Polarity)}} |
|||
*{{fl|BCB-adhesion.pdf|BCB Adhesion Notes from Vendor}} |
|||
*{{fl|BCB-rework.pdf|BCB rework Notes from Vendor}} |
|||
;<div id="Developers"><big>Developers</big></div> |
|||
*{{fl|AZ400K-Developer-Datasheet.pdf|AZ400K (AZ400K, AZ400K1:4)}} |
|||
*{{fl|AZ300MIF-Developer-Datasheet.pdf|AZ300MIF}} |
|||
*DS2100 BCB Developer |
|||
*SU-8 Developer |
|||
*101A Developer (for DUV Flood Exposed PMGI) |
|||
;<div id="PRRemovers"><big>Photoresist Removers</big></div> |
|||
*[http://www.microchemicals.com/products/remover_stripper/nmp.html AZ NMP] |
|||
**''This replaces {{fl|1165-Resist-Remover.pdf|1165}}'' |
|||
*{{fl|AZ300T-Resist-Remover.pdf|AZ300T}} |
|||
*{{fl|RemoverPG-revA.pdf|Remover PG, SU-8 stripper}} |
|||
*AZ EBR ("Edge Bead Remover", PGMEA) |
|||
|} |
|} |
||
[[Category: Processing]] |
|||
[[category: Lithography]] |
|||
[[category: Recipes]] |
|||
Latest revision as of 22:43, 22 October 2026
| Table of Contents |
|---|
Photolithography Processes
|
Photolithography Chemicals/Materials
|
General Photolithography Techniques
HMDS Process for Improving Adhesion
- Use these procedures if you are finding poor adhesion PR lifting-off), or for chemicals (like BHF) that attack the PR adhesion interface strongly.
Edge-Bead Removal Techniques
- These techniques are required for loading full-wafers into etchers that use top-side clamps, to prevent photoresist from sticking to the clamp (and potentially destroying your wafer).
- For contact lithography, this improves the proximity of the mask plate and sample, improving resolution. For some projection systems, such as the Maskless Aligner, EBR can help with autofocus issues.
- For small samples (<1cm) some techniques here to reduce edge-bead via shims.
Photoresist reflow (MicroChem)
- To create slanted sidewalls or curved surfaces.
Lithography Calibration - Analyzing a Focus-Exposure Matrix
- On tools with autofocus (steppers & direct-write litho tools), you calibrate your litho by shooting a "Focus Exposure Matrix/Array", or "FEM/FEA", which exposes a grid with Dose varying in one axis, and Focus varying in the other.
- Explains how to locate the "Process Window" for your lithography.
Photolithography Recipes
Process Maturity Ranking
R6- most mature process with regular calibrations recorded on SPC charts.- …
R1- least mature - "possible" but you'll have to figure it out.A: Material is available for use, but no recipes are provided.
The Key/Legend for this table's R1...R6 values is at the bottom of the table.
Click the tool title to go to recipes for that tool.
Click the photoresist title to get the datasheet, also found in Stocked Chemicals + Datasheets.
Process Ranking Table
Processes in the table above are ranked by their "Process Maturity Level" as follows:
| Process Level | Description of Process Level Ranking | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| A | Process Allowed and materials available but never done | ||||||||||
| R1 | Process has been ran at least once | ||||||||||
| R2 | Process has been ran and procedure is documented | ||||||||||
| R3 | Process has been ran, procedure is documented, and data is available | ||||||||||
| R4 | Process has a documented procedure with regular (≥4x per year) data no in-Situ control available | ||||||||||
| R5 | Process has a documented procedure with regular (≥4x per year) data and in-Situ control available | ||||||||||
| R6 | Process has a documented procedure and control charts/limits available. Controlled process. | ||||||||||
Lift-Off Recipes
- Lift-Off Description/Tutorial
- How it works, process limits and considerations for designing your process
- I-Line Lift-Off: Bi-Layer Process with LOL2000 Underlayer
- Single Expose/Develop process for simplicity
- Up to ~130nm metal thickness & ≥500nm-1000nm gap between metal.
- Can use any I-Line litho tool (GCA Stepper, Contact aligner, MLA)
- I-Line Lift-Off: Bi-Layer Process with PMGI Underlayer and Contact Aligner
- Multiple processes for Metal thicknesses ~800nm to ~2.5µm
- Uses multiple DUV Flood exposure/develop cycles to create undercut.
- Can be transferred to other I-Line litho tools (Stepper, MLA etc.)
- Updates to the process:
- You now only need to do 5min DUV Flood exposure each cycle (lamp optics were replaced)
- Use the new "101B" developer, which replaced the "SAL/101A" developers but works the same.
- There are 2 DUV Flood exposure + 101B develop steps because the first exposure doesn't expose the entire PMGI - because PMGI absorbs DUV very strongly so the light doesn't penetrate to the bottom of thick PMGI on the first exposure.
- DUV Lift-off: LOR Underlayers
- 1µm (LOR-10A) & 2-3µm (LOR-30C) versions
- Written for ASML DUV Stepper, but should work for any Imaging (top) resist, incl. I-Line tools.
- DUV Lift-Off: UV6 Imaging Resist + PMGI Underlayer
- Single-expose/develop process
- Up to ~65nm metal thickness & ~350nm gap between metal
- Use thicker PMGI for thicker metals
E-Beam Lithography Recipes (JEOL JBX-6300FS)
- Under Development.
FIB Lithography Recipes (Raith Velion)
To Be Added
Automated Coat/Develop System Recipes (S-Cubed Flexi)
Recipes pre-loaded on the S-Cubed Flexi automated coat/bake/develop system. Only staff may write new recipes, contact the tool supervisor for more info.
Available Variations
- We have different recipes with varyious UV6 spin speeds - the same spin speed optionss as found on our manual Headway spinners. This allows for PR thickness control. See the linked UV6 datasheets below for thickness vs. rpm spin curves. Note that exact spin RPM may be slightly different between Headway spinners (on benches) vs. S-Cubed.
- DSK is recommended to be spun at 1.5krpm (~40nm) for best anti-reflection properties. 5krpm (~20nm) recipes are also provided for historical/legacy processes.
- DSK can be baked at either 220C to act as a Dry-etchable BARC (similar to DUV-42P), or at lower temps as a developable BARC (no dry etch required).
- "Chain" recipes (with DSK+UV6 spin/cured in succession) are only available for DSK Baked at 185C & 220C, and all UV6 Spin-speed variations. For the other DSK temps you can use the single-PR "Routes".
- Developer recipes are now available for 300 MiF Developer. Note that developer is flowing continuously during the develop, so develop times are shorter by ~50% compared to beaker developing.
- DO NOT EDIT developer recipes, they can damage the tool when programmed incorrectly!
- "Chain" recipes for 135*C Post-exposure Bake + Develop have been made.
- ASML cassettes can be placed directly on this tool for spin / exposure / develop process. Please make sure all cassettes are kept back at their respecting tools when done.
Recipes Table (S-Cubed Flexi)
| Coating Material | Route/Chain | Name: (User: "UCSB Users") | Spin Speed (krpm) | Bake Temp | Notes |
| BEFORE LITHOGRAPHY (PR Coat and Bake) | |||||
|---|---|---|---|---|---|
| Hotplate Set | Route | To pre-set the DSK Hotplate temp (HP4).
Note: Only HP4 can be changed. HP1-HP3 remains fixed at: HP1=135°C, HP2=170°C & HP3=170°C | |||
| HP4-SET-220C | 220°C | Will over shoot +-2°C when done. | |||
| HP4-SET-210C | 210°C | ||||
| HP4-SET-200C | 200°C | ||||
| HP4-SET-185C | 185°C | ||||
| Bottom Anti-Reflection Coating (BARC), DSK101 only: | |||||
| DS-K101 | Route | DSK101 Develop Rate depends on Bake temp - you can use this to control undercut. See: DSK Bake vs. Dev rate
DSK101 spun at 1.5K is equivalent to DUV-42P. See: Stepper Recipes#Anti-Reflective Coatings 185°C bake allows the DSK to dissolve during develop, and allows for undercut (may lift-off small features) 220°C bake allows DSK to be used as dry-etchable BARC (equiv. to DUV42P), requiring O2 etch to remove. Better for small features. See here for relevant processing info from DUV42P. Note: All PR coat recipes have EBR backside clean steps included in the recipe. | |||
| 1.5krpm recipes | COAT-DSK101-304[1.5K]-185C | 1.5krpm | 185°C | Requires: HP4=185°C, | |
| COAT-DSK101-304[1.5K]-200C | 1.5krpm | 200°C | Requires: HP4=200°C | ||
| COAT-DSK101-304[1.5K]-210C | 1.5krpm | 210°C | Requires: HP4=210°C | ||
| COAT-DSK101-304[1.5K]-220C | 1.5krpm | 220°C | Requires: HP4=220°C, | ||
| 5krpm recipes | COAT-DSK101-304[5K]-185C | 5.0krpm | 185°C | Requires: HP4=185°C, | |
| COAT-DSK101-304[5K]-200C | 5.0krpm | 200°C | Requires: HP4=200°C | ||
| COAT-DSK101-304[5K]-210C | 5.0krpm | 210°C | Requires: HP4=210°C | ||
| COAT-DSK101-304[5K]-220C | 5.0krpm | 220°C | Requires: HP4=220°C, | ||
| Imaging resist (UV6) only: | |||||
| UV6-0.8 | Route | COAT-UV6[2K]-135C | 2.0krpm | 135°C | Requires: HP1=135°C |
| varying spin speed | COAT-UV6[2.5K]-135C | 2.5krpm | 135°C | Requires: HP1=135°C | |
| COAT-UV6[3K]-135C | 3.0krpm | 135°C | Requires: HP1=135°C | ||
| COAT-UV6[3.5K]-135C | 3.5krpm | 135°C | Requires: HP1=135°C | ||
| COAT-UV6[4K]-135C | 4.0krpm | 135°C | Requires: HP1=135°C | ||
| COAT-UV6[5K]-135C | 5.0krpm | 135°C | Requires: HP1=135°C | ||
| COAT-UV6[6K]-135C | 6.0krpm | 135°C | Requires: HP1=135°C | ||
| UV6 Coat with Developable BARC underlayer: | |||||
| DS-K101 @ 185°C
+ UV6 |
Chain | COAT-DSK101[1.5K]-185C-UV6[2K]-135C | DSK: 1.5krpm
UV6: 2.0krpm |
DSK: 185°C
UV6: 135°C |
Requires:
– HP4=185°C – HP1=135°C Plan for ~10-15 min per wafer. |
| 1.5krpm DSK recipes with
UV6- various spin speeds |
COAT-DSK101[1.5K]-185C-UV6[2.5K]-135C | DSK: 1.5krpm
UV6: 2.5krpm |
same as above | same as above | |
| COAT-DSK101[1.5K]-185C-UV6[3K]-135C | DSK: 1.5krpm
UV6: 3.0krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-185C-UV6[3.5K]-135C | DSK: 1.5krpm
UV6: 3.5krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-185C-UV6[4K]-135C | DSK: 1.5krpm
UV6: 4.0krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-185C-UV6[5K]-135C | DSK: 1.5krpm
UV6: 5.0krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-185C-UV6[6K]-135C | DSK: 1.5krpm
UV6: 6.0krpm |
same as above | same as above | ||
| 5krpm DSK recipes with
UV6- various spin speeds |
COAT-DSK101[5K]-185C-UV6[2K]-135C | DSK: 5krpm
UV6: 2.0krpm |
same as above | same as above | |
| COAT-DSK101[5K]-185C-UV6[2.5K]-135C | DSK: 5krpm
UV6: 2.5krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-185C-UV6[3K]-135C | DSK: 5krpm
UV6: 3.0krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-185C-UV6[3.5K]-135C | DSK: 5krpm
UV6: 3.5krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-185C-UV6[4K]-135C | DSK: 5krpm
UV6: 4.0krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-185C-UV6[5K]-135C | DSK: 5krpm
UV6: 5.0krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-185C-UV6[6K]-135C | DSK: 5krpm
UV6: 6.0krpm |
same as above | same as above | ||
| UV6 Coat with Dry-Etchable BARC underlayer: | |||||
| DS-K101 @ 220°C
+ UV6 |
Chain | COAT-DSK101[1.5K]-220C-UV6[2K]-135C | DSK: 1.5krpm
UV6: 2.0krpm |
DSK: 220°C
UV6: 135°C |
Requires:
– HP4=220°C – HP1=135°C Plan for ~10-15 min per wafer. |
| 1.5krpm DSK recipes with
UV6- various spin speeds |
COAT-DSK101[1.5K]-220C-UV6[2.5K]-135C | DSK: 1.5krpm
UV6: 2.5krpm |
same as above | same as above | |
| COAT-DSK101[1.5K]-220C-UV6[3K]-135C | DSK: 1.5krpm
UV6: 3.0krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-220C-UV6[3.5K]-135C | DSK: 1.5krpm
UV6: 3.5krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-220C-UV6[4K]-135C | DSK: 1.5krpm
UV6: 4.0krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-220C-UV6[5K]-135C | DSK: 1.5krpm
UV6: 5.0krpm |
same as above | same as above | ||
| COAT-DSK101[1.5K]-220C-UV6[6K]-135C | DSK: 1.5krpm
UV6: 6.0krpm |
same as above | same as above | ||
| 5krpm DSK recipes with
UV6- various spin speeds |
COAT-DSK101[5K]-220C-UV6[2K]-135C | DSK: 5krpm
UV6: 2.0krpm |
same as above | same as above | |
| COAT-DSK101[5K]-220C-UV6[2.5K]-135C | DSK: 5krpm
UV6: 2.5krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-220C-UV6[3K]-135C | DSK: 5krpm
UV6: 3.0krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-220C-UV6[3.5K]-135C | DSK: 5krpm
UV6: 3.5krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-220C-UV6[4K]-135C | DSK: 5krpm
UV6: 4.0krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-220C-UV6[5K]-135C | DSK: 5krpm
UV6: 5.0krpm |
same as above | same as above | ||
| COAT-DSK101[5K]-220C-UV6[6K]-135C | DSK: 5krpm
UV6: 6.0krpm |
same as above | same as above | ||
| AFTER LITHOGRAPHY (PEB and Developing) | |||||
| PEB Wafer Bake | Route | To bake wafer with UV6 after exposure (PEB) for 90sec and cool for 15sec | |||
| BAKE-135C-90S | 135°C | Requires: HP1=135°C | |||
| PEB and Developing | Route | To post-exposure bake wafer after exposure (std. PEB for UV6) 90sec, cool 15sec, develop using AZ300MIF and water rinse 60sec
WARNING: DO NOT USE ANY OTHER DEVELOPER RECIPES OTHER THAN THE ONES LISTED HERE (or equipment damage may occur!) | |||
| Varying developer time | BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-10S | Developer chuck: 300rpm | 135°C | Requires: HP1=135°C | |
| BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-15S | same as above | same as above | same as above | ||
| BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-20S | same as above | same as above | same as above | ||
| BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-25S | same as above | same as above | same as above | ||
| BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-30S | same as above | same as above | same as above | ||
| BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-35S | same as above | same as above | same as above | ||
| BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-40S | same as above | same as above | same as above | ||
| BAKE-135C-90S-DEV[MIF300]-SPIN[300RPM]-45S | same as above | same as above | same as above | ||
| Developing | Route | To only develop wafer using AZ300MIF and water rinse 60sec (No PEB)
WARNING: DO NOT USE ANY OTHER DEVELOPER RECIPES OTHER THAN THE ONES LISTED HERE (or equipment damage may occur!) | |||
| Varying developer time | DEV[MIF300]-SPIN[300RPM]-10S | Developer chuck: 300rpm | 135°C | Requires: HP1=135°C | |
| DEV[MIF300]-SPIN[300RPM]-15S | same as above | same as above | same as above | ||
| DEV[MIF300]-SPIN[300RPM]-20S | same as above | same as above | same as above | ||
| DEV[MIF300]-SPIN[300RPM]-25S | same as above | same as above | same as above | ||
| DEV[MIF300]-SPIN[300RPM]-30S | same as above | same as above | same as above | ||
| DEV[MIF300]-SPIN[300RPM]-35S | same as above | same as above | same as above | ||
| DEV[MIF300]-SPIN[300RPM]-40S | same as above | same as above | same as above | ||
| DEV[MIF300]-SPIN[300RPM]-45S | same as above | same as above | same as above | ||
Holography Recipes
The Holography recipes here use the BARC layer XHRiC-11 & the high-res. I-Line photoresist THMR-IP3600HP-D.
- Standard Holography Process - on SiO2 on Si
- Holography Process Variations - Set-up Angle - Etching into SiO2 and Si
- Etch SiO2 Nano-structure - Changing Side-wall Angle - Etching into Si with a different line-width
- Reduce SiO2 Nanowire Diameter - Thermal Oxidation - Vapor HF Etching
Low-K Spin-On Dielectric Recipes
Chemicals Stocked + Datasheets
The following is a list of the lithography chemicals we stock in the lab, with links to the datasheets for each. The datasheets will often have important processing info such as spin-speed vs. thickness curves, typical process parameters, bake temps/times etc.
Item numbers in (parentheses) indicate the specific stocked formulation, when the datasheet shows multiple formulations.